Dose evaluation of multi-slice CT for different parameters in chest examinations using TLD method

被引:8
|
作者
Chang, Yen-Lin [2 ]
Lee, Chi-Ming [2 ]
Hsiao, Wen-Tien [2 ]
Hsu, Fang-Yuh [3 ,4 ]
Kuo, Ming-Jay [4 ]
Chiu, Jih-Hung [1 ]
机构
[1] Yuanpei Univ, Dept Radiol Technol, Hsinchu 30015, Taiwan
[2] Taipei Med Univ Hosp, Dept Diagnost Radiol, Taipei, Taiwan
[3] Natl Tsing Hua Univ, Nucl Sci Technol & Dev Ctr, Hsinchu, Taiwan
[4] Natl Tsing Hua Univ, Dept Biomed Engn & Environm Sci, Hsinchu, Taiwan
关键词
MSCT; TLD; Effective dose; Organ dose;
D O I
10.1016/j.radmeas.2009.12.043
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In radiodiagnosis, optimization is defined as a method of reducing the patient-absorbed doses without affecting image quality. In this study, we aimed to determine the parameters of a multi-slice CT (MSCT) scanner (LightSpeed VCT 64: General Electric) that ensure the least organ dose with the best image quality (i.e., optimization) for chest examinations, which are the most frequently used radio diagnostic investigations. Organ doses of an anthropomorphic phantom were measured during MSCT chest scans by using GR-100 (LiF: Mg, Ti) thermo-luminescent dosimeters (TLDs) attached to various parts of the organs. The accumulated equivalent doses were calculated to evaluate the radiation risk to sensitive organs by using different scan conditions and measuring the absorbed doses, because different scan parameters (e.g., pitch, tube current, scan type, etc.) may affect the dose contribution to these organs. By means of the radiation and tissue weighting factors recommended in the International Commission on Radiological Protection (ICRP) 103 report, the effective dose was then estimated. The pitch value and tube current were the main parameters affecting the organ doses of the phantom: increase in the pitch value and reduction of the tube current decreased the absorbed doses without impacting the image resolution and diagnostic information. Consequently, radiologists can establish an optimization protocol in MSCT and acquire high-resolution images with adequate diagnostic information while using the lowest exposure dose for chest examinations. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:701 / 703
页数:3
相关论文
共 50 条
  • [1] Chest examination protocol with a reduced dose using a multi-slice spiral CT
    Rehbock, B
    Hieckel, HG
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2003, 175 (07): : 963 - 966
  • [2] Evaluation of Low-dose Multi-slice CT Scanning in Children Chest Examination
    Li, Guohua
    Tang, Weilin
    Sun, Shengjian
    PROCEEDINGS OF THE 2016 INTERNATIONAL CONFERENCE ON EDUCATION, MANAGEMENT AND COMPUTER SCIENCE (ICEMC 2016), 2016, 129 : 384 - 388
  • [3] The dose modulation in multi-slice CT of the chest: A new advancement for dose reduction
    Ujita, M
    Nishioka, M
    Kimura, M
    Fukuda, K
    Suess, C
    RADIOLOGY, 2001, 221 : 214 - 214
  • [4] Designing a phantom for dose evaluation in multi-slice CT
    Abboud, Samir
    Badal, Andreu
    Stern, Stanley H.
    Kyprianou, Iacovos S.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [5] Dose in multi-slice spiral CT
    Spreiter, Q
    Doubrava, C
    von der Haar, T
    Flohr, T
    Schaller, S
    PHYSICA MEDICA, 2000, 16 (03) : 151 - 154
  • [6] Multi-Slice CT Dose Evaluation Using Optically Stimulated Luminescence (OSL) System
    Ruan, C.
    Clouse, W.
    Yukihara, E.
    Gasparian, P.
    Ahmad, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [7] MULTI-SLICE CT OF THE CHEST: ROUTINE AND NEW APPLICATIONS
    Duyck, Ph.
    JBR-BTR, 2007, 90 (02): : 122 - 123
  • [8] Evaluating doses of multi-slice CT in brain examinations using various methods
    Lin, Hung-Chih
    Lai, Te-Jen
    Tseng, Hsien-Chun
    Lin, Cheng-Hsun
    Tseng, Yen-Ling
    Chen, Chien-Yi
    COMPUTER ASSISTED SURGERY, 2017, 22 : 54 - 60
  • [9] Performance evaluation of a multi-slice CT system
    McCollough, CH
    Zink, FE
    Welch, TJ
    RADIOLOGY, 1999, 210 (02) : 586 - 586
  • [10] Performance evaluation of a multi-slice CT system
    McCollough, CH
    Zink, FE
    MEDICAL PHYSICS, 1999, 26 (11) : 2223 - 2230