Biparietal bidirectional bipolar radiofrequency in hybrid cardiac ablation: an in vitro evaluation

被引:2
作者
Matteucci, Francesco [1 ]
Maesen, Bart [1 ]
De Asmundis, Carlo [2 ]
Bidar, Elham [1 ]
Micali, Linda [1 ]
Parise, Gianmarco [1 ]
Maessen, Jos G. [1 ]
La Meir, Mark [1 ,2 ]
Gelsomino, Sandro [1 ,2 ]
机构
[1] Maastricht Univ Hosp, Cardiothorac Dept, Maastricht, Netherlands
[2] Brussels Univ Hosp, Cardiothorac Dept, Brussels, Belgium
关键词
Hybrid AF ablation; Bipolar ablation; In vitro simulation; Catheter ablation; Image processing; ATRIAL-FIBRILLATION; BEATING HEART; CATHETER ABLATION; PULMONARY VEINS; ENERGY; PERFORMANCE; MONOPOLAR; SAFETY;
D O I
10.1093/icvts/ivab047
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: The aim of this study was to evaluate the lesion size and depth of radiofrequency (RF) ablation in a simultaneous biparietal bidirectional bipolar (SBB) approach, compared to a simultaneous and staged unipolar and uniparietal bipolar setup [simultaneous uniparietal bipolar (SiUB) and staged uniparietal bipolar (StUB), respectively]. METHODS: Fresh left atrial porcine tissue was mounted into the ABLA-BOX simulator. Different ablation approaches were tested: (i) SBB: a concept consisting of SBB endo-epicardial ablation, (ii) SiUB: simultaneous epicardial uniparietal bipolar and endocardial unipolar ablation and (iii) StUB: staged epicardial uniparietal bipolar and endocardial unipolar ablation. In the StUB setup, a 1-h interval between the epi-endo ablation was respected. RESULTS: Transmural lesions were present in 90% of the bipolar biparietal ablations, yet no full transmurality was observed in the simultaneous nor in the staged unipolar with uniparietal bipolar ablation group. In SBB, the area and volume of the ablation lesions were smaller (523.33 mm(2)/mm and 52.33 mm(3)/mm, respectively) than in SiUB (588.17 mm(2)/mm and 58.81 mm(3)/mm, respectively) and StUB (583.76 mm(2)/mm and 58.37 mm(3)/mm, P=0.044). Also, in SBB, the overall, epicardial and endocardial maximum diameters of the lesions (1.59, 1.57 and 1.52 mm; respectively) were smaller than in SiUB (2.38, 2.26 and 2.33 mm; respectively) and in StUB (2.36, 2.28 and 2.14 mm; respectively, all P < 0.001). CONCLUSIONS: Although bipolar biparietal bidirectional RF ablation results in smaller lesions than uniparietal bipolar and unipolar ablation, their capacity to penetrate the tissue is much higher. Moreover, in uniparietal RF applications, the energy spreads in the superficial layers of the tissue but fails to penetrate. Therefore, the degree of transmurality is much higher when using such a 'truly bipolar' ablation approach.
引用
收藏
页码:34 / 42
页数:9
相关论文
共 29 条
  • [1] Persistent atrial fibrillation: A systematic review and meta-analysis of invasive strategies
    Berger, Wouter R.
    Meulendijks, Eva R.
    Limpens, Jacqueline
    van den Berg, Nicoline W. E.
    Neefs, Jolien
    Driessen, Antoine H. G.
    Krul, Sebastien P. J.
    van Boven, Wim Jan P.
    de Groot, Joris R.
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2019, 278 : 137 - 143
  • [2] No-Touch Radiofrequency Ablation: A Comparison of Switching Bipolar and Switching Monopolar Ablation in Ex Vivo Bovine Liver
    Chang, Won
    Lee, Jeong Min
    Lee, Sang Min
    Han, Joon Koo
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (02) : 279 - 288
  • [3] Duncan DI, 2015, AESTHET DERMATOL, V2, P1, DOI 10.1159/000362747
  • [4] Successful performance of cox-maze procedure on beating heart using bipolar radiofrequency ablation: A feasibility study in animals
    Gaynor, SL
    Ishii, Y
    Diodato, MD
    Prasad, SM
    Barnett, KM
    Damiano, NR
    Byrd, GD
    Wickline, SA
    Schuessler, RB
    Damiano, RJ
    [J]. ANNALS OF THORACIC SURGERY, 2004, 78 (05) : 1671 - 1677
  • [5] Haines D E., 2011, Catheter ablation of cardiac arrhythmias, V2nd, P2
  • [6] Late outcomes after the Cox maze IV procedure for atrial fibrillation
    Henn, Matthew C.
    Lancaster, Timothy S.
    Miller, Jacob R.
    Sinn, Laurie A.
    Schuessler, Richard B.
    Moon, Marc R.
    Melby, Spencer J.
    Maniar, Hersh S.
    Damiano, Ralph J., Jr.
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2015, 150 (05) : 1168 - +
  • [7] Issa ZF, 2012, CLIN ARRHYTHMOLOGY E, P144
  • [8] A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research
    Koo, Terry K.
    Li, Mae Y.
    [J]. JOURNAL OF CHIROPRACTIC MEDICINE, 2016, 15 (02) : 155 - 163
  • [9] Bipolar catheter ablation in ventricular myocardium
    Liu, Xiaolin
    Chen, Qiushi
    Fan, Baohan
    Zhu, Wenwu
    Zhao, Hongyan
    Zhu, Yeqian
    Zhao, Pengcheng
    Zhang, Fengxiang
    Kojodjojo, Pipin
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2020, 43 (01): : 54 - 61
  • [10] Pearls and Pitfalls in Catheter Ablation of Persistent Atrial Fibrillation
    Lo, Li-Wei
    Lin, Yenn-Jiang
    Chang, Shih-Lin
    Hu, Yu-Feng
    Chung, Fa-Po
    Chen, Shih-Ann
    [J]. CIRCULATION JOURNAL, 2016, 80 (02) : 306 - 313