Iatrogenic catheter-related cardiac tamponade: A case report of fatal hydropericardium following subcutaneous implantation of a chemotherapeutic injection port
被引:0
作者:
Shields, LBE
论文数: 0引用数: 0
h-index: 0
机构:Urban Govt Ctr, Off Chief Med Examiner, Louisville, KY 40204 USA
Shields, LBE
Hunsaker, DM
论文数: 0引用数: 0
h-index: 0
机构:Urban Govt Ctr, Off Chief Med Examiner, Louisville, KY 40204 USA
Hunsaker, DM
Hunsaker, JC
论文数: 0引用数: 0
h-index: 0
机构:Urban Govt Ctr, Off Chief Med Examiner, Louisville, KY 40204 USA
Hunsaker, JC
机构:
[1] Urban Govt Ctr, Off Chief Med Examiner, Louisville, KY 40204 USA
[2] Univ Louisville, Dept Pathol & Lab Med, Sch Med, Louisville, KY 40292 USA
[3] Off Associate Chief Med Examiner, Frankfort, KY USA
[4] Univ Kentucky, Coll Med, Dept Pathol & Lab Med, Lexington, KY USA
forensic science;
forensic pathology;
central venous catheter;
adverse effect;
fatal hydropericardium;
cardiac tamponade;
and iatrogenesis;
D O I:
暂无
中图分类号:
DF [法律];
D9 [法律];
R [医药、卫生];
学科分类号:
0301 ;
10 ;
摘要:
The need to obtain dependable access to the vascular system constitutes a significant component in the treatment and management of critically ill patients. Intravenous chemotherapy administered to cancer patients over an extended period of time often results in loss of peripheral vascular access due to vein sclerosis, "exhaustion" or tissue necrosis. Medical investigators have designed and steadily upgraded a variety of devices constructed to improve venous access for long-term utilization. As with the introduction of any foreign object into the body, each of these devices has complications which may be life threatening and occasionally fatal. We present an unusual case of iatrogenic acute hydropericardium and cardiac tamponade caused by the percutaneous infusion of chemotherapeutic fluid via a right subclavian central venous implant system (Porta-Catho(R)). Failure to implant and monitor the device with a radiograph following placement according to manufacturer's guidelines and accepted standards of medical practice were causally related to an unusual complication, namely, perforation of the right cardiac ventricle by the catheter tip, resulting in sudden and unexpected cardiac death.