Optimizing cuff width for noninvasive measurement of blood pressure

被引:52
|
作者
Marks, LA
Groch, A
机构
[1] St Christophers Hosp Children, Philadelphia, PA 19133 USA
[2] Temple Univ, Sch Med, Philadelphia, PA 19122 USA
[3] Univ Med & Dent New Jersey, Newark, NJ 07103 USA
关键词
blood pressure; blood pressure determination; hemodynamics; hypertension;
D O I
10.1097/00126097-200006000-00002
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background It is well established that indirect measurements of blood pressure made with a standard 13 cm-wide cuff are erroneously high for large arms and low for small arms. To correct for this error, the American Heart Association recommends adjusting cuff width to 40% of the arm's circumference, Objective To test the validity of this method of correction. Design This study was a prospective, nonblinded, paired Student's t-test analysis. Methods Blood pressures in 50 subjects were measured directly by using a radial artery line and indirectly by the Korotkov method. For each subject multiple indirect measurements of blood pressure were made with the cuff width:arm circumference ratio varied from 30-55% in 5% increments. Error was defined as indirect blood pressure minus direct blood pressure. Results A ratio of 40% resulted in overestimation of blood pressure for most arms and with particularly high errors for small arms. The ratio producing zero mean error for the pooled study group was 46.4 +/- 0.7% (mean +/- SEM). Using this ratio of 46.4%, the error varied inversely with arm circumference (P < 0.02), resulting in overestimation of systolic blood pressure for small arms and underestimation of systolic blood pressure for large arms. This error is comparable in magnitude, but opposite in sign, to that which occurs with a standard 13 cm-wide cuff for all arms. The optimum ratio was found to be closely approximated by the relationship, cuff width = 9.34 log(10) arm circumference. Using this relationship, error in systolic blood pressure was insensitive to arm circumference (r = 0.04, P > 0.05) and near zero. Conclusion The optimum cuff width for the indirect measurement of blood pressure is not directly proportional to arm circumference, but is proportional to the logarithm of the arm's circumference. Blood Press Monit 5:153-158 (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:153 / 158
页数:6
相关论文
共 50 条
  • [31] EFFECT OF CUFF WIDTH ON SYSTOLIC BLOOD-PRESSURE MEASUREMENTS IN NEONATES
    LUM, LG
    JONES, MD
    JOURNAL OF PEDIATRICS, 1977, 91 (06): : 963 - 966
  • [32] Intraoperative Blood Pressure Misclassification Due to Inaccuracy of Noninvasive Oscillometric Cuff-Measured Blood Pressure
    Kim, Danny Sungsoo
    Kwon, Sohee
    Bovik, Alan C.
    Markey, Mia K.
    Giovannucci, Edward L.
    Cannesson, Maxime
    HYPERTENSION, 2025, 82 (02) : e1 - e3
  • [33] Cuff-Less Methods for Blood Pressure Measurement
    Gan, Liangzhi
    Wang, Chen
    Liu, Haikuan
    Zhang, Lei
    2019 CHINESE AUTOMATION CONGRESS (CAC2019), 2019, : 4926 - 4930
  • [34] WHICH CUFF SIZE FOR BLOOD-PRESSURE MEASUREMENT
    VERDON, F
    SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 1992, 122 (40) : 1491 - 1496
  • [35] VALIDITY OF BLOOD PRESSURE MEASUREMENT WITH CUFF IN ARM AND FOREARM
    FORSBERG, SA
    DEGUZMAN, M
    BERLIND, S
    ACTA MEDICA SCANDINAVICA, 1970, 188 (05): : 389 - +
  • [37] Ankle systolic blood pressure measurement and cuff application
    Wertheim, D
    Trenary, K
    Melhuish, J
    Williams, R
    Harding, K
    LANCET, 1996, 347 (9018): : 1845 - 1846
  • [38] BLOOD-PRESSURE MEASUREMENT - GETTING THE RIGHT CUFF
    FEUSSNER, JR
    BLESSINGFEUSSNER, CL
    LINFORS, EW
    NORTH CAROLINA MEDICAL JOURNAL, 1983, 44 (04) : 241 - 241
  • [39] EFFECT OF CUFF LOCATION ON BLOOD-PRESSURE MEASUREMENT
    SHIMIZU, M
    BIORHEOLOGY, 1987, 24 (05) : 515 - 515
  • [40] INFLUENCE OF CUFF SIZE AND CUFF DESIGN ON INDIRECT BLOOD-PRESSURE MEASUREMENT
    HUIGE, MC
    VANVOLLENHOVEN, E
    WEEDA, HWH
    ARNTZENIUS, AC
    ACTA CARDIOLOGICA, 1978, 33 (02) : 88 - 90