摘要
Background It is well established that indirect measurements of blood pressure made with a standard 13 cm-wide cuff are erroneously high for large arms and low for small arms. To correct for this error, the American Heart Association recommends adjusting cuff width to 40% of the arm's circumference, Objective To test the validity of this method of correction. Design This study was a prospective, nonblinded, paired Student's t-test analysis. Methods Blood pressures in 50 subjects were measured directly by using a radial artery line and indirectly by the Korotkov method. For each subject multiple indirect measurements of blood pressure were made with the cuff width:arm circumference ratio varied from 30-55% in 5% increments. Error was defined as indirect blood pressure minus direct blood pressure. Results A ratio of 40% resulted in overestimation of blood pressure for most arms and with particularly high errors for small arms. The ratio producing zero mean error for the pooled study group was 46.4 +/- 0.7% (mean +/- SEM). Using this ratio of 46.4%, the error varied inversely with arm circumference (P < 0.02), resulting in overestimation of systolic blood pressure for small arms and underestimation of systolic blood pressure for large arms. This error is comparable in magnitude, but opposite in sign, to that which occurs with a standard 13 cm-wide cuff for all arms. The optimum ratio was found to be closely approximated by the relationship, cuff width = 9.34 log(10) arm circumference. Using this relationship, error in systolic blood pressure was insensitive to arm circumference (r = 0.04, P > 0.05) and near zero. Conclusion The optimum cuff width for the indirect measurement of blood pressure is not directly proportional to arm circumference, but is proportional to the logarithm of the arm's circumference. Blood Press Monit 5:153-158 (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:153 / 158
页数:6
相关论文
共 50 条
  • [1] OPTIMIZING CUFF WIDTH FOR THE NONINVASIVE MEASUREMENT OF BLOOD-PRESSURE
    MARKS, LA
    GROCH, AJ
    CIRCULATION, 1994, 90 (04) : 616 - 616
  • [2] Cuff width and accuracy of measurement of blood pressure
    Alpert, BS
    BLOOD PRESSURE MONITORING, 2000, 5 (03) : 151 - 152
  • [3] THE IMPORTANCE OF CUFF WIDTH IN MEASUREMENT OF BLOOD-PRESSURE IN PREGNANCY
    SCHOENFELD, A
    ZIV, I
    NERI, A
    OVADIA, J
    EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1982, 14 (02): : 89 - 100
  • [4] CUFF BLADDER WIDTH AND BLOOD-PRESSURE MEASUREMENT IN CHILDREN AND ADOLESCENTS
    GOMEZMARIN, O
    PRINEAS, RJ
    RASTAM, L
    JOURNAL OF HYPERTENSION, 1992, 10 (10) : 1235 - 1241
  • [5] EFFECT OF CUFF WIDTH ON ACCURACY OF INDIRECT MEASUREMENT OF BLOOD PRESSURE IN DOGS
    VALTONEN, MH
    ERIKSSON, LM
    RESEARCH IN VETERINARY SCIENCE, 1970, 11 (04) : 358 - &
  • [6] Evaluation of the oscillometric method for noninvasive blood pressure measurement during cuff deflation and cuff inflation with reference to intra-arterial blood pressure
    Celler, Branko G.
    Yong, Andy
    Rubenis, Imants
    Butlin, Mark
    Argha, Ahmadreza
    Rehan, Rajan
    Avolio, Alberto
    JOURNAL OF HYPERTENSION, 2024, 42 (07) : 1235 - 1247
  • [7] Cuff width influence on blood pressure measurement during the pregnant-puerperal cycle
    Oliveira, SMJV
    Arcuri, EAM
    Santos, JLF
    JOURNAL OF ADVANCED NURSING, 2002, 38 (02) : 180 - 189
  • [8] The cuff width influences the toe blood pressure value
    Påhlsson, HI
    Jörneskog, G
    Wahlberg, E
    VASA-JOURNAL OF VASCULAR DISEASES, 2004, 33 (04): : 215 - 218
  • [9] Blood pressure determination in children - The effect of the width of the cuff
    Day, R
    JOURNAL OF PEDIATRICS, 1939, 14 : 148 - 155
  • [10] The optimal cuff width for measuring toe blood pressure
    Pahlsson, Hans-Ivar
    Laskar, Carolina
    Stark, Karin
    Andersson, Anna
    Jogestrand, Tomas
    Wahlberg, Eric
    ANGIOLOGY, 2007, 58 (04) : 472 - 476