In vitro performance of a centrifugal, a mixed flow, and an axial flow blood pump

被引:12
作者
Araki, K
Anai, H
Oshikawa, M
Nakamura, K
Onitsuka, T
机构
[1] Miyazaki Med Coll Hosp, Kiyotake, Miyazaki 88916, Japan
[2] Oita Med Univ, Dept Cardiovasc Surg, Oita, Japan
[3] Miyazaki Med Coll, Dept Surg 2, Miyazaki 88916, Japan
关键词
centrifugal pump; mixed flow pump; axial flow pump; hemolysis; efficiency;
D O I
10.1046/j.1525-1594.1998.06142.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We specially devised 3 types of turbo pumps, a centrifugal pump (CFP), a mixed flow pump (MFP), and an axial flow pump (AFP), and analyzed their in vitro performance. The common structural design elements were an impeller diameter of 20 mm and sealless magnet couple driving. In vitro tests were carried out using heparinized fresh bovine blood. The hemolysis was comprehensively evaluated at 7-16 points by changing the flow rate and pressure head (mapping of hemolytic property), The maximum efficiency (motor output to pump output) was 44.9% at 7,000 rpm. 3.17 L/min, 191 mm Hg in the CFP; 66.3% at 7,000 rpm, 6.9 L/min, 136 mm Hg in the MFP; and 20.6% at 9,000 rpm, 5.54 L/min, 74 mm Hg in the AFP, respectively. The minimum normalized index of hemolysis (NIH) (g/100 L) was 0.038 at 5,000 rpm, 4.60 L/min, 38 mm Hg in the CFP; 0.010 at 7,000 rpm, 8.22 L/min, 100 mm Hg in the MFP; and 0.033 at 7,000 rpm, 2.84 L/min, 48 mm Hg in the AFP, respectively. The best efficiency and NIH were achieved in the MFP.
引用
收藏
页码:366 / 370
页数:5
相关论文
共 7 条
[1]  
ANAI H, 1995, ASAIO J, V41, pM771, DOI 10.1097/00002480-199507000-00117
[2]   CONTRIBUTION TO IN VITRO TESTING OF PUMPS FOR EXTRACORPOREAL CIRCULATION [J].
KOLLER, T ;
HAWRYLENKO, A .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1967, 54 (01) :22-+
[3]   DEVELOPMENT OF THE BAYLOR NASA AXIAL-FLOW VENTRICULAR ASSIST DEVICE - IN-VITRO PERFORMANCE AND SYSTEMATIC HEMOLYSIS TEST-RESULTS [J].
MIZUGUCHI, K ;
DAMM, GA ;
BOZEMAN, RJ ;
AKKERMAN, JW ;
ABER, GS ;
SVEJKOVSKY, PA ;
BACAK, JW ;
ORIME, Y ;
TAKATANI, S ;
NOSE, Y ;
NOON, GP ;
DEBAKEY, ME .
ARTIFICIAL ORGANS, 1994, 18 (01) :32-43
[4]  
PARK CH, 1995, ASAIO J, V41, pM345, DOI 10.1097/00002480-199507000-00028
[5]  
PARNIS SM, 1995, ASAIO J, V41, pM333, DOI 10.1097/00002480-199507000-00025
[6]   Improvement in antithrombogenicity in a centrifugal pump with self wash-out structure for long-term use [J].
WAKISAKA, Y ;
TAENAKA, Y ;
TATSUMI, E ;
ARAKI, K ;
MASUZAWA, T ;
NAKATANI, T ;
BABA, Y ;
ANAI, H ;
EYA, K ;
TODA, K ;
TAKANO, H .
ASAIO JOURNAL, 1995, 41 (03) :M350-M355
[7]  
YAMAZAKI K, 1995, ASAIO J, V41, pM327, DOI 10.1097/00002480-199507000-00024