Comparative performance analysis for abdominal phantom ROI detectability according to CT reconstruction algorithm: ADMIRE

被引:3
作者
Shin, Jun-Bong [1 ,2 ]
Yoon, Do-Kun [1 ,2 ]
Pak, Seongyong [3 ]
Kwon, Yang-Ho [3 ]
Suh, Tae Suk [1 ,2 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 01期
关键词
abdomen phantom; ADMIRE; FBP; MDCT; SAFIRE; MODELED ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY;
D O I
10.1002/acm2.12765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We compared and analyzed the detectability performance pertaining to an abdominal phantom including a region of interest (ROI) according to a computed tomography (CT) reconstruction algorithm. Methods Three types of reconstruction algorithms (FBP, SAFIRE, and ADMIRE) were used to evaluate the detectability performance using the abdominal phantom (phantom size: 25 x 18 x 28 cm(3)). The vendor default settings for routine multi-detector computed tomography abdominal scans were used. As the quantitative evaluation method, the contrast-to-noise ratio (CNR), difference in coefficient of variation (COV) with the normalization based on the FBP data, and the noise power spectrum (NPS) were measured. Results The characteristic of the ADMIRE-3 reconstructed image was higher than those of the FBP and SAFIRE-3 reconstructed images. The CNR values of the SAFIRE and ADMIRE images were much higher than the corresponding values of the FBP images. The difference in COV values for the ADMIRE images was similar to 1.2 times lower than the corresponding values of the SAFIRE images. Conclusion The comparative analysis of the abdominal phantom low-contrast resolution differences for each CT exposure parameters showed that ADMIRE demonstrated better results than SAFIRE and FBP in terms of contrast, CNR, COV difference, and 1D NPS. This indicates that ADMIRE can provide a clearer observation even with the same number of contrast objects as compared to SAFIRE and FBP owing to its better contrast resolution in the central part of the contrast hole at low kV.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [31] Spatial resolution, noise properties, and detectability index of a deep learning reconstruction algorithm for dual-energy CT of the abdomen
    Thor, Daniel
    Titternes, Rebecca
    Poludniowski, Gavin
    MEDICAL PHYSICS, 2023, 50 (05) : 2775 - 2786
  • [32] Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm
    Solomon, Justin
    Marin, Daniele
    Choudhury, Kingshuk Roy
    Patel, Bhavik
    Samei, Ehsan
    RADIOLOGY, 2017, 284 (03) : 777 - 787
  • [33] Effects of acquisition method and reconstruction algorithm for CT number measurement on standard-dose CT and reduced-dose CT: a QIBA phantom study
    Ohno, Yoshiharu
    Fujisawa, Yasuko
    Fujii, Kenji
    Sugihara, Naoki
    Kishida, Yuji
    Seki, Shinichiro
    Yoshikawa, Takeshi
    JAPANESE JOURNAL OF RADIOLOGY, 2019, 37 (05) : 399 - 411
  • [34] 80-kVp CT Using Iterative Reconstruction in Image Space Algorithm for the Detection of Hypervascular Hepatocellular Carcinoma: Phantom and Initial Clinical Experience
    Hur, Saebeom
    Lee, Jeong Min
    Kim, Soo Jin
    Park, Ji Hoon
    Han, Joon Koo
    Choi, Byung Ihn
    KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (02) : 152 - 164
  • [35] Efficient Evaluation of Low-contrast Detectability of Deep-CNN-based CT Reconstruction Using Channelized Hotelling Observer on the ACR Accreditation Phantom
    Fan, Mingdong
    Zhou, Zhongxing
    Vrieze, Thomas
    Wang, Jia
    McCollough, Cynthia
    Yu, Lifeng
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [36] Evaluation of an integrated 3D-printed phantom for coronary CT angiography using iterative reconstruction algorithm
    Abdullah, Kamarul A.
    McEntee, Mark F.
    Reed, Warren
    Kench, Peter L.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2020, 67 (03) : 170 - 176
  • [37] Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study
    Dabli, Djamel
    Frandon, Julien
    Belaouni, Asmaa
    Akessoul, Philippe
    Addala, Takieddine
    Berny, Laure
    Beregi, Jean-Paul
    Greffer, Joel
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2022, 103 (01) : 31 - 40
  • [38] Iterative CT reconstruction in abdominal low-dose CT used for hybrid SPECT-CT applications: effect on image quality, image noise, detectability, and reader's confidence
    Grosser, Oliver S.
    Ruf, Juri
    Kupitz, Dennis
    Czuczwara, Damian
    Loewenthal, David
    Thormann, Markus
    Furth, Christian
    Ricke, Jens
    Denecke, Timm
    Pech, Maciej
    Kreissl, Michael C.
    Amthauer, Holger
    ACTA RADIOLOGICA OPEN, 2019, 8 (06):
  • [39] Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT
    Zhong, Jingyu
    Shen, Hailin
    Chen, Yong
    Xia, Yihan
    Shi, Xiaomeng
    Lu, Wei
    Li, Jianying
    Xing, Yue
    Hu, Yangfan
    Ge, Xiang
    Ding, Defang
    Jiang, Zhenming
    Yao, Weiwu
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (04) : 1390 - 1407
  • [40] Effect of tube voltage and tube current in abdominal ct scans with principal components analysis and cluster analysis: Phantom study
    Acikgoz, Gunes
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2025, 16 (03): : 197 - 202