Comparative performance analysis for abdominal phantom ROI detectability according to CT reconstruction algorithm: ADMIRE

被引:3
作者
Shin, Jun-Bong [1 ,2 ]
Yoon, Do-Kun [1 ,2 ]
Pak, Seongyong [3 ]
Kwon, Yang-Ho [3 ]
Suh, Tae Suk [1 ,2 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 01期
关键词
abdomen phantom; ADMIRE; FBP; MDCT; SAFIRE; MODELED ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY;
D O I
10.1002/acm2.12765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We compared and analyzed the detectability performance pertaining to an abdominal phantom including a region of interest (ROI) according to a computed tomography (CT) reconstruction algorithm. Methods Three types of reconstruction algorithms (FBP, SAFIRE, and ADMIRE) were used to evaluate the detectability performance using the abdominal phantom (phantom size: 25 x 18 x 28 cm(3)). The vendor default settings for routine multi-detector computed tomography abdominal scans were used. As the quantitative evaluation method, the contrast-to-noise ratio (CNR), difference in coefficient of variation (COV) with the normalization based on the FBP data, and the noise power spectrum (NPS) were measured. Results The characteristic of the ADMIRE-3 reconstructed image was higher than those of the FBP and SAFIRE-3 reconstructed images. The CNR values of the SAFIRE and ADMIRE images were much higher than the corresponding values of the FBP images. The difference in COV values for the ADMIRE images was similar to 1.2 times lower than the corresponding values of the SAFIRE images. Conclusion The comparative analysis of the abdominal phantom low-contrast resolution differences for each CT exposure parameters showed that ADMIRE demonstrated better results than SAFIRE and FBP in terms of contrast, CNR, COV difference, and 1D NPS. This indicates that ADMIRE can provide a clearer observation even with the same number of contrast objects as compared to SAFIRE and FBP owing to its better contrast resolution in the central part of the contrast hole at low kV.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [21] Usefulness of a Metal Artifact Reduction Algorithm for Orthopedic Implants in Abdominal CT: Phantom and Clinical Study Results
    Jeong, Seonji
    Kim, Se Hyung
    Hwang, Eui Jin
    Shin, Cheong-Il
    Han, Joon Koo
    Choi, Byung Ihn
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (02) : 307 - 317
  • [22] Visual impression and texture analysis of advanced modeled iterative reconstruction (ADMIRE): improved assessment of image quality in CT for better estimation of dose reduction potential
    Alikhani, Babak
    Raatschen, Hans-Juergen
    Wacker, Frank
    Werncke, Thomas
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (03)
  • [23] Impact of CT reconstruction algorithm on auto-segmentation performance
    Miller, Claudia
    Mittelstaedt, Daniel
    Black, Noel
    Klahr, Paul
    Nejad-Davarani, Siamak
    Schulz, Heinrich
    Goshen, Liran
    Han, Xiaoxia
    Ghanem, Ahmed, I
    Morris, Eric D.
    Glide-Hurst, Carri
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (09): : 95 - 103
  • [24] Speedup performance analysis of parallel Katsevich algorithm for 3D CT image reconstruction
    Ni, Jun
    Deng, Junjun
    He, Tao
    Yu, Hengyong
    Wang, Ge
    INTERNATIONAL JOURNAL OF COMPUTATIONAL SCIENCE AND ENGINEERING, 2011, 6 (03) : 151 - 159
  • [25] Comparative analysis of the radiation shield effect in an abdominal CT scan
    Kim, Seon-Chil
    Kim, Young-Jae
    Lee, Joon-Seok
    Dong, Kyung-Rae
    Chung, Woon-Kwan
    Lim, Chang-Seon
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (06) : 929 - 935
  • [26] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Jeong Hee Yoon
    Jeong Min Lee
    Bo Yun Hur
    Jeehyun Baek
    Hackjoon Shim
    Joon Koo Han
    Byung Ihn Choi
    Abdominal Imaging, 2015, 40 : 1843 - 1852
  • [27] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Yoon, Jeong Hee
    Lee, Jeong Min
    Hur, Bo Yun
    Baek, Jeehyun
    Shim, Hackjoon
    Han, Joon Koo
    Choi, Byung Ihn
    ABDOMINAL IMAGING, 2015, 40 (06): : 1843 - 1852
  • [28] Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study
    Greffier, Joel
    Hamard, Aymeric
    Pereira, Fabricio
    Barrau, Corinne
    Pasquier, Hugo
    Beregi, Jean Paul
    Frandon, Julien
    EUROPEAN RADIOLOGY, 2020, 30 (07) : 3951 - 3959
  • [29] Performance of Iterative Reconstruction and Automated Tube Voltage Selection on the Image Quality and Radiation Dose in Abdominal CT Scans
    Desai, Gaurav S.
    Orrego, Jorge M. Fuentes
    Kambadakone, Avinash R.
    Sahani, Dushyant V.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (06) : 897 - 903
  • [30] A tensor PRISM algorithm for multi-energy CT reconstruction and comparative studies
    Li, Liang
    Chen, Zhiqiang
    Wang, Ge
    Chu, Jiyang
    Gao, Hao
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (02) : 147 - 163