Comparative performance analysis for abdominal phantom ROI detectability according to CT reconstruction algorithm: ADMIRE

被引:3
|
作者
Shin, Jun-Bong [1 ,2 ]
Yoon, Do-Kun [1 ,2 ]
Pak, Seongyong [3 ]
Kwon, Yang-Ho [3 ]
Suh, Tae Suk [1 ,2 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
[3] Siemens Healthineers Ltd, Seoul, South Korea
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2020年 / 21卷 / 01期
关键词
abdomen phantom; ADMIRE; FBP; MDCT; SAFIRE; MODELED ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; RADIATION-DOSE REDUCTION; COMPUTED-TOMOGRAPHY;
D O I
10.1002/acm2.12765
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We compared and analyzed the detectability performance pertaining to an abdominal phantom including a region of interest (ROI) according to a computed tomography (CT) reconstruction algorithm. Methods Three types of reconstruction algorithms (FBP, SAFIRE, and ADMIRE) were used to evaluate the detectability performance using the abdominal phantom (phantom size: 25 x 18 x 28 cm(3)). The vendor default settings for routine multi-detector computed tomography abdominal scans were used. As the quantitative evaluation method, the contrast-to-noise ratio (CNR), difference in coefficient of variation (COV) with the normalization based on the FBP data, and the noise power spectrum (NPS) were measured. Results The characteristic of the ADMIRE-3 reconstructed image was higher than those of the FBP and SAFIRE-3 reconstructed images. The CNR values of the SAFIRE and ADMIRE images were much higher than the corresponding values of the FBP images. The difference in COV values for the ADMIRE images was similar to 1.2 times lower than the corresponding values of the SAFIRE images. Conclusion The comparative analysis of the abdominal phantom low-contrast resolution differences for each CT exposure parameters showed that ADMIRE demonstrated better results than SAFIRE and FBP in terms of contrast, CNR, COV difference, and 1D NPS. This indicates that ADMIRE can provide a clearer observation even with the same number of contrast objects as compared to SAFIRE and FBP owing to its better contrast resolution in the central part of the contrast hole at low kV.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [1] CT Reconstruction Algorithm and Low Contrast Detectability of Phantom Study: A Systematic Review and Meta-Analysis
    Yusof, Nur Aimi Adibah
    Karim, Muhammad Khalis Abdul
    Asikin, Nursyazalina Mohd
    Paiman, Suriati
    Kechik, Mohd Mustafa Awang
    Rahman, Mohd Amiruddin Abdul
    Noor, Noramaliza Mohd
    CURRENT MEDICAL IMAGING, 2023, 19 (10) : 1105 - 1113
  • [2] Advanced Modeled Iterative Reconstruction (ADMIRE) Facilitates Radiation Dose Reduction in Abdominal CT
    Ellmann, Stephan
    Kammerer, Ferdinand
    Allmendinger, Thomas
    Hammon, Matthias
    Janka, Rolf
    Lell, Michael
    Uder, Michael
    Kramer, Manuel
    ACADEMIC RADIOLOGY, 2018, 25 (10) : 1277 - 1284
  • [3] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Rubert, Nicholas
    Southard, Richard
    Hamman, Susan M.
    Robison, Ryan
    PEDIATRIC RADIOLOGY, 2020, 50 (03) : 345 - 356
  • [4] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Nicholas Rubert
    Richard Southard
    Susan M. Hamman
    Ryan Robison
    Pediatric Radiology, 2020, 50 : 345 - 356
  • [5] Effects of various generations of iterative CT reconstruction algorithms on low-contrast detectability as a function of the effective abdominal diameter: A quantitative task-based phantom study
    Viry, Anais
    Aberle, Christoph
    Racine, Damien
    Knebel, Jean-Francois
    Schindera, Sebastian T.
    Schmidt, Sabine
    Becce, Fabio
    Verdun, Francis R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 48 : 111 - 118
  • [6] Hybrid Iterative Reconstruction Technique for Abdominal CT Protocols in Obese Patients: Assessment of Image Quality, Radiation Dose, and Low-Contrast Detectability in a Phantom
    Schindera, Sebastian T.
    Odedra, Devang
    Mercer, Diego
    Thipphavong, Seng
    Chou, Paul
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (02) : W146 - W152
  • [7] IMAGE QUALITY AND POTENTIAL DOSE REDUCTION USING ADVANCED MODELED ITERATIVE RECONSTRUCTION (ADMIRE) IN ABDOMINAL CT - A REVIEW
    Kataria, B.
    Althen, J. Nilsson
    Smedby, O.
    Persson, A.
    Sokjer, H.
    Sandborg, M.
    RADIATION PROTECTION DOSIMETRY, 2021, 195 (3-4) : 177 - 187
  • [8] Image quality assessment of an iterative reconstruction algorithm applied to abdominal CT imaging
    Funama, Yoshinori
    Taguchi, Katsuyuki
    Utsunomiya, Daisuke
    Oda, Seitaro
    Katahira, Kazuhiro
    Tokuyasu, Shinichi
    Yamashita, Yasuyuki
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (04): : 527 - 534
  • [9] Low contrast detectability performance of model observers based on CT phantom images: kVp influence
    Hernandez-Giron, I.
    Calzado, A.
    Geleijns, J.
    Joemai, R. M. S.
    Veldkamp, W. J. H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 798 - 807
  • [10] Iterative Reconstruction Algorithm for CT: Can Radiation Dose Be Decreased While Low-Contrast Detectability Is Preserved?
    Schindera, Sebastian T.
    Odedra, Devang
    Raza, Syed Arsalan
    Kim, Tae Kyoung
    Jang, Hyun-Jung
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    RADIOLOGY, 2013, 269 (02) : 510 - 517