Quantitative measurements of proton spin-lattice (T1) and spin-spin (T2) relaxation times in the mouse brain at 7.0 T

被引:43
作者
Guilfoyle, DN
Dyakin, VV
O'Shea, J
Pell, GS
Helpern, JA
机构
[1] Nathan S Kline Inst Psychiat Res, Div Med Phys, Orangeburg, NY 10962 USA
[2] Nathan S Kline Inst Psychiat Res, Ctr Adv Brain Imaging, Orangeburg, NY 10962 USA
[3] NYU, Sch Med, Dept Radiol, New York, NY USA
[4] NYU, Sch Med, Dept Psychiat, New York, NY USA
[5] NYU, Sch Med, Dept Physiol & Neurosci, New York, NY USA
关键词
T-1; T-2; k-space; snapshot FLASH; hippocampus; corpus callosum; retrosplenial granular (RSG) cortex;
D O I
10.1002/mrm.10371
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The goal of this work is to provide regional T-1 and T-2 values at a field strength of 7 T for the normal mouse brain at 6 weeks and 1 year old. A novel segmented snapshot FLASH sequence was used to measure T-1 in the hippocampus, corpus callosum, and the retrosplenial granular (RSG) cortex; T-2 measurements were made in the same regions using a single spin echo sequence repeated at six separate echo times. Both T-1 and T-2 measurements were validated with phantom measurements. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:576 / 580
页数:5
相关论文
共 21 条
[1]   A necessary role for cell shrinkage in apoptosis [J].
Bortner, CD ;
Cidlowski, JA .
BIOCHEMICAL PHARMACOLOGY, 1998, 56 (12) :1549-1559
[2]  
Calamante F, 1999, MAGN RESON MED, V41, P479, DOI 10.1002/(SICI)1522-2594(199903)41:3<479::AID-MRM9>3.0.CO
[3]  
2-2
[4]   Rapid T1 mapping using multislice echo planar imaging [J].
Clare, S ;
Jezzard, P .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) :630-634
[5]   ERRORS IN T2 ESTIMATION USING MULTISLICE MULTIPLE-ECHO IMAGING [J].
CRAWLEY, AP ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :34-47
[6]  
Dunn JF, 1999, MUSCLE NERVE, V22, P1367, DOI 10.1002/(SICI)1097-4598(199910)22:10<1367::AID-MUS5>3.0.CO
[7]  
2-H
[8]  
Franklin K B J, 2008, MOUSE BRAIN STEREOTA
[9]   FLASH IMAGING - RAPID NMR IMAGING USING LOW FLIP-ANGLE PULSES [J].
HAASE, A ;
FRAHM, J ;
MATTHAEI, D ;
HANICKE, W ;
MERBOLDT, KD .
JOURNAL OF MAGNETIC RESONANCE, 1986, 67 (02) :258-266
[10]   CHANGES OF RELAXATION-TIMES (T-1, T-2) AND APPARENT DIFFUSION-COEFFICIENT AFTER PERMANENT MIDDLE CEREBRAL-ARTERY OCCLUSION IN THE RAT - TEMPORAL EVOLUTION, REGIONAL EXTENT, AND COMPARISON WITH HISTOLOGY [J].
HOEHNBERLAGE, M ;
EIS, M ;
BACK, T ;
KOHNO, K ;
YAMASHITA, K .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (06) :824-834