Effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure due to tight orbit syndrome

被引:15
作者
Lee, Y. K. [1 ]
Lee, J. Y. [1 ]
Moon, J. I. [1 ]
Park, M. H. [1 ]
机构
[1] Catholic Univ Korea, Coll Med, Yeouido St Marys Hosp, Dept Ophthalmol, Seoul 150713, South Korea
关键词
Glaucoma; Goldmann applanation tonometer; ICare tonometer; Prostaglandin analogs; Tight orbit syndrome; GOLDMANN APPLANATION TONOMETER; TOPICAL BIMATOPROST THERAPY; OCULAR-RESPONSE-ANALYZER; GROWTH-FACTOR-BETA; GLAUCOMA PATIENTS; BIOMECHANICAL PROPERTIES; CLINICAL-EVALUATION; PHOSPHOLIPASE A(2); LID SULCUS; REPRODUCIBILITY;
D O I
10.1007/s10384-014-0343-3
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To evaluate the effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure (IOP) due to tight orbit syndrome and to identify factors affecting the development of tight orbit syndrome in glaucoma patients. We investigated 84 eyes in 84 glaucoma patients, of which 14 eyes were classified in the tight orbit syndrome group and 70 eyes in the control group. IOP was measured using the ICare tonometer and the Goldmann applanation tonometer (GAT). The demographic data, medical histories, ocular histories, and detailed ocular drug histories of the two groups were compared to identify factors contributing to the development of tight orbit syndrome. In the tight orbit syndrome group, the ICare tonometer significantly underestimated the IOP by approximately 8.6 mmHg compared with the GAT. In the control group, the IOP readings of the GAT and the ICare tonometer did not differ significantly. Bland-Altman analysis showed that the mean difference between measurements taken using the GAT and those taken using the ICare tonometer was 2.5 +/- 6.3 mmHg. The difference between the GAT and ICare tonometer measurements was greater in the tight orbit syndrome group (8.6 +/- A 5.3 mmHg) than in the control group (1.3 +/- A 2.7 mmHg). Multivariate regression analysis revealed that only the use of prostaglandin analogs (PGAs) was associated with the development of tight orbit syndrome. The ICare tonometer is a suitable alternative device for use in patients with tight orbit syndrome in whom the IOP may be overestimated with the GAT. The prolonged use of PGAs is significantly associated with the development of tight orbit syndrome.
引用
收藏
页码:496 / 502
页数:7
相关论文
共 35 条
[1]   Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists [J].
Abraham, L. M. ;
Epasinghe, N. C. R. ;
Selva, D. ;
Casson, R. .
EYE, 2008, 22 (04) :503-506
[2]   Incidence of deepening of the upper eyelid sulcus after switching from latanoprost to bimatoprost [J].
Aihara, Makoto ;
Shirato, Shiroaki ;
Sakata, Rei .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2011, 55 (06) :600-604
[3]   Side Effects Associated with Prostaglandin Analog Therapy [J].
Alm, Albert ;
Grierson, Ian ;
Shields, M. Bruce .
SURVEY OF OPHTHALMOLOGY, 2008, 53 :S93-S105
[4]   TRANSFORMING GROWTH-FACTOR-BETA IN DISEASE - THE DARK SIDE OF TISSUE-REPAIR [J].
BORDER, WA ;
RUOSLAHTI, E .
JOURNAL OF CLINICAL INVESTIGATION, 1992, 90 (01) :1-7
[5]  
BORDER WA, 1994, NEW ENGL J MED, V331, P1286
[6]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[7]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209
[8]   DYNAMIC TONOMETRY [J].
DEKKING, HM ;
COSTER, HD .
OPHTHALMOLOGICA, 1967, 154 (01) :59-&
[9]   Prostaglandin F2α facilitates collagen synthesis in cardiac fibroblasts via an F-prostanoid receptor/protein kinase C/Rho kinase pathway independent of transforming growth factor β1 [J].
Ding, Wen-yuan ;
Ti, Yun ;
Wang, Jia ;
Wang, Zhi-hao ;
Xie, Guo-lu ;
Shang, Yuan-yuan ;
Tang, Meng-xiong ;
Zhang, Yun ;
Zhang, Wei ;
Zhong, Ming .
INTERNATIONAL JOURNAL OF BIOCHEMISTRY & CELL BIOLOGY, 2012, 44 (06) :1031-1039
[10]   Bimatoprost and travoprost: A review of recent studies of two new glaucoma drugs [J].
Eisenberg, DL ;
Toris, CB ;
Camras, CB .
SURVEY OF OPHTHALMOLOGY, 2002, 47 :S105-S115