Effects of scanning and reconstruction parameters on image quality in electron-beam CT angiography: Coronary Artery Phantom Study

被引:10
作者
Lu, B
Dai, RP
Bai, H
He, S
Jiang, SL
Zhuang, N
Budoff, MJ
机构
[1] Harbor UCLA Med Ctr, Dept Med, Div Cardiol, Torrance, CA 90502 USA
[2] Peking Union Med Coll, FuWai Cardiovasc Inst & Hosp, Dept Radiol, Beijing, Peoples R China
[3] Chinese Acad Med Sci, Beijing 100037, Peoples R China
[4] St Johns Cardiovasc Res Ctr, Torrance, CA 90502 USA
关键词
computed tomography (CT); electron beam; image processing; vascular studies; coronary vessels; CT;
D O I
10.1016/S1076-6332(00)80174-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. This study compared the image quality obtained with different scanning and reconstruction parameters for electron-beam computed tomographic (CT) angiography and sought optimal methods for visualizing the coronary artery lumen. Materials and Methods. Electron-beam CT angiography with contrast material enhancement was used to image 35 branches of fresh postmortem swine coronary arteries. Different collimation widths, fields of view (FOVs), reconstruction kernels, and algorithms were employed to reconstruct the acquired raw data into CT angiographic images. Image quality was compared and analyzed. Results. The contrast-to-noise ratios (C/Ns) for 1.5-, 2-, and 3-mm section thickness were 28.4 +/- 15.2, 31.9 +/- 9.3, and 33.8 +/- 14.5, respectively (P < .05). The lengths of visualized coronary artery lumina were significantly longer for 1.5-mm scanning (71.6 mm +/- 4.3) than for 2-mm (58.3 mm +/- 5.5) and 3-mm scanning (59.0 mm +/- 8.0) (P < .01). The C/Ns for 12.7-, 18.0-, and 26.0-cm FOV reconstruction were 32.8 +/- 9.9, 28.9 +/- 8.2, and 27.1 +/- 8.2, respectively (not significant), and the visualized luminal lengths were 76.1 mm +/- 12.5, 71.7 mm +/- 14.6, and 65.4 mm +/- 13.1, respectively (not significant). The highest C/N (48.2 +/- 13.3) was achieved with smooth kernels and a cone-beam algorithm, and the lowest (14.7 +/- 3.4) with very sharp kernels and a normal algorithm. Cone-beam algorithm images had significantly higher C/Ns than did normal algorithm images (P < .001), and they demonstrated longer coronary artery lumina (P < .01). Conclusion. Collimation width, FOV, reconstruction kernels, and algorithms are important in the processing of high-quality electron-beam coronary angiograms. A 1.5-mm collimation width, 12.7-cm FOV, cone-beam reconstruction algorithm, and very sharp kernels should help in obtaining the best image quality and depicting the longest segments of coronary artery lumen.
引用
收藏
页码:927 / 933
页数:7
相关论文
共 23 条
  • [1] Comparison of vessel diameters in electron beam tomography and quantitative coronary angiography
    Achenbach, S
    Moshage, W
    Ropers, D
    Bachmann, K
    [J]. INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 1998, 14 (01): : 1 - 7
  • [2] Curved multiplanar reconstructions for the evaluation of contrast-enhanced electron beam CT of the coronary arteries
    Achenbach, S
    Moshage, W
    Ropers, D
    Bachmann, K
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (04) : 895 - 899
  • [3] BRINK JA, 1992, RADIOLOGY, V185, P126
  • [4] TECHNICAL OPTIMIZATION OF SPIRAL CT FOR DEPICTION OF RENAL-ARTERY STENOSIS - IN-VITRO ANALYSIS
    BRINK, JA
    LIM, JT
    WANG, G
    HEIKEN, JP
    DEYOE, LA
    VANNIER, MW
    [J]. RADIOLOGY, 1995, 194 (01) : 157 - 163
  • [5] BRINK JA, 1993, RADIOLOGY, V189, P189
  • [6] Intravenous three-dimensional coronary angiography using contrast enhanced electron beam computed tomography
    Budoff, MJ
    Oudiz, RJ
    Zalace, CP
    Bakhsheshi, H
    Goldberg, SL
    French, WJ
    Rami, TG
    Brundage, BH
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1999, 83 (06) : 840 - 845
  • [7] Evaluation of electron beam CT coronary angiography in healthy subjects
    Chernoff, DM
    Ritchie, CJ
    Higgins, CB
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 169 (01) : 93 - 99
  • [8] A phantom study: Evaluation of renal artery stenosis using helical CT and 3D reconstructions
    Davros, WJ
    Obuchowski, NA
    Berman, PM
    Zeman, RK
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (01) : 156 - 161
  • [9] Blurring of vessels in spiral CT angiography: Effects of collimation width, pitch, viewing plane, and windowing in maximum intensity projection
    Diederichs, CG
    Keating, DP
    Glatting, G
    Oestmann, JW
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (06) : 965 - 974
  • [10] CT ANGIOGRAPHY - APPLICATION TO THE EVALUATION OF CAROTID-ARTERY STENOSIS
    DILLON, EH
    VANLEEUWEN, MS
    FERNANDEZ, MA
    EIKELBOOM, BC
    MALI, WPTM
    [J]. RADIOLOGY, 1993, 189 (01) : 211 - 219