LDIAED: A lightweight deep learning algorithm implementable on automated external defibrillators

被引:3
|
作者
Nasimi, Fahimeh [1 ]
Yazdchi, Mohammadreza [1 ]
机构
[1] Univ Isfahan, Fac Engn, Dept Biomed Engn, Esfahan, Iran
来源
PLOS ONE | 2022年 / 17卷 / 02期
关键词
VENTRICULAR-FIBRILLATION; CARDIAC-ARREST;
D O I
10.1371/journal.pone.0264405
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Differentiating between shockable and non-shockable Electrocardiogram (ECG) signals would increase the success of resuscitation by the Automated External Defibrillators (AED). In this study, a Deep Neural Network (DNN) algorithm is used to distinguish 1.4-second segment shockable signals from non-shockable signals promptly. The proposed technique is frequency-independent and is trained with signals from diverse patients extracted from MIT-BIH, MIT-BIH Malignant Ventricular Ectopy Database (VFDB), and a database for ventricular tachyarrhythmia signals from Creighton University (CUDB) resulting, in an accuracy of 99.1%. Finally, the raspberry pi minicomputer is used to load the optimized version of the model on it. Testing the implemented model on the processor by unseen ECG signals resulted in an average latency of 0.845 seconds meeting the IEC 60601-2-4 requirements. According to the evaluated results, the proposed technique could be used by AED's.
引用
收藏
页数:12
相关论文
共 50 条
  • [11] Automated external defibrillators in the workplace
    Descatha, Alexis
    Baer, Michel
    BRITISH MEDICAL JOURNAL, 2008, 337 (7674):
  • [12] Endorsing automated external defibrillators
    不详
    CANADIAN FAMILY PHYSICIAN, 2000, 46 : 746 - 746
  • [13] Automated external defibrillators in school?
    Cornelis, C.
    Calle, P.
    Mpotos, N.
    Monsieurs, K.
    ACTA CLINICA BELGICA, 2022, 77 : 14 - 14
  • [14] Automated external defibrillators in schools?
    Cornelis, Charlotte
    Calle, Paul
    Mpotos, Nicolas
    Monsieurs, Koenraad
    ACTA CARDIOLOGICA, 2015, 70 (03) : 249 - 254
  • [15] Automated external defibrillators: To what extent does the algorithm delay CPR?
    Rea, TD
    Shah, S
    Kudenchuk, PJ
    Copass, MK
    Cobb, LA
    ANNALS OF EMERGENCY MEDICINE, 2005, 46 (02) : 132 - 141
  • [16] AUTOMATED EXTERNAL DEFIBRILLATORS Defibrillators are only as good as the provider
    Gregory, Anna R.
    BRITISH MEDICAL JOURNAL, 2008, 337
  • [17] Machine Learning Techniques for the Detection of Shockable Rhythms in Automated External Defibrillators
    Figuera, Carlos
    Irusta, Unai
    Morgado, Eduardo
    Aramendi, Elisabete
    Ayala, Unai
    Wik, Lars
    Kramer-Johansen, Jo
    Eftestol, Trygve
    Alonso-Atienza, Felipe
    PLOS ONE, 2016, 11 (07):
  • [18] Public use of automated external defibrillators
    Caffrey, SL
    Willoughby, PJ
    Pepe, PE
    Becker, LB
    NEW ENGLAND JOURNAL OF MEDICINE, 2002, 347 (16): : 1242 - 1247
  • [19] Automated external defibrillators: recalls & alerts
    不详
    EXPERT REVIEW OF MEDICAL DEVICES, 2006, 3 (05) : 534 - 535
  • [20] Automated external defibrillators. Reply
    Page, RL
    Kowal, RC
    McKenas, DK
    NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (10): : 773 - 773