Preservation of residual hearing after cochlear implant surgery with slim modiolar electrode

被引:39
作者
Iso-Mustajarvi, Matti [1 ,2 ]
Sipari, Sini [1 ,2 ]
Lopponen, Heikki [1 ,2 ]
Dietz, Aarno [1 ]
机构
[1] Univ Eastern Finland, Yliopistonranta 1, Kuopio 70210, Finland
[2] Kuopio Univ Hosp, Ear Nose & Throat Dis, KNK Klin, Puijonlaaksontie 2, Kuopio 70210, Finland
关键词
Cochlear implant; Hearing preservation; Modiolar electrode; Electro-acoustic stimulation; ANGULAR INSERTION DEPTH; MATRIX SENTENCE TEST; SPEECH-PERCEPTION; EARLY OUTCOMES; LOCATION; IMPACT; ARRAY; STIMULATION; MULTICENTER;
D O I
10.1007/s00405-019-05708-x
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Purpose To evaluate the insertion results and hearing preservation of a novel slim modiolar electrode (SME) in patients with residual hearing. Methods We retrospectively collected the data from the medical files of 17 patients (18 ears) implanted with a SME. All patients had functional low frequency hearing (PTA ((0.125-0.5 kHz)) <= 80 dB HL). The insertion results were re-examined from the postoperative cone-beam computed tomography scans. Postoperative thresholds were obtained at the time of switch-on of the sound processors (mean 43 days) and at latest follow-up (mean 582 days). The speech recognition in noise was measured with the Finnish matrix sentence test preoperatively and at follow-up. Results The mean insertion depth angle (IDA) was 395 degrees. Neither scala dislocations nor tip fold over were detected. There were no total hearing losses. Functional low-frequency hearing was preserved in 15/18 (83%) ears at switch-on and in 14/17 (82%) ears at follow-up. According to HEARRING classification, 55% (10/18) had complete HP at switch-on and 41% (7/17) still at follow-up. Thirteen patients (14 ears) were initially fitted with electric-acoustic stimulation and seven patients (8 ears) continued to use it after follow-up. Conclusions The preliminary hearing preservation results with the SME were more favorable than reported for other perimodiolar electrodes. The results show that the array may also be feasible for electro-acoustic stimulation; it is beneficial in that it provides adequate cochlear coverage for pure electrical stimulation in the event of postoperative or progressive hearing loss.
引用
收藏
页码:367 / 375
页数:9
相关论文
共 47 条
[1]   Minimum Reporting Standards for Adult Cochlear Implantation [J].
Adunka, Oliver F. ;
Gantz, Bruce J. ;
Dunn, Camille ;
Gurgel, Richard K. ;
Buchman, Craig A. .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2018, 159 (02) :215-219
[2]   Quality control after insertion of the nucleus contour and contour advance electrode in adults [J].
Aschendorff, Antje ;
Kromeier, Jan ;
Klenzner, Thomas ;
Laszig, Roland .
EAR AND HEARING, 2007, 28 (02) :75S-79S
[3]   Clinical investigation of the Nucleus Slim Modiolar Electrode [J].
Aschendorff, Antje ;
Briggs, Robert ;
Brademann, Goetz ;
Helbig, Silke ;
Hornung, Joachim ;
Lenarz, Thomas ;
Marx, Mathieu ;
Ramos, Angel ;
Stoever, Timo ;
Escude, Bernard ;
James, Chris J. .
AUDIOLOGY AND NEURO-OTOLOGY, 2017, 22 (03) :169-179
[4]   Use of a mid-scala and a lateral wall electrode in children: insertion depth and hearing preservation [J].
Benghalem, Abdelhamid ;
Gazibegovic, Dzemal ;
Saadi, Fatima ;
Tazi-Chaoui, Zakia .
ACTA OTO-LARYNGOLOGICA, 2017, 137 (01) :1-7
[5]  
Briggs Robert J S, 2011, Cochlear Implants Int, V12, P129, DOI 10.1179/1754762811Y0000000007
[6]  
Buchman CA, 2014, OTOL NEUROTOL, V35, P1773, DOI 10.1097/MAO.0000000000000541
[7]   Implications of Minimizing Trauma During Conventional Cochlear Implantation [J].
Carlson, Matthew L. ;
Driscoll, Colin L. W. ;
Gifford, Rene H. ;
Service, Geoffrey J. ;
Tombers, Nicole M. ;
Hughes-Borst, Becky J. ;
Neff, Brian A. ;
Beatty, Charles W. .
OTOLOGY & NEUROTOLOGY, 2011, 32 (06) :962-968
[8]   A Retrospective Analysis of the Contribution of Reported Factors in Cochlear Implantation on Hearing Preservation Outcomes [J].
Causon, Andrew ;
Verschuur, Carl ;
Newman, Tracey A. .
OTOLOGY & NEUROTOLOGY, 2015, 36 (07) :1137-1145
[9]   Psychophysical measures in patients fitted with Contour™ and straight Nucleus electrode arrays [J].
Cohen, LT ;
Saunders, E ;
Knight, MR ;
Cowan, RSC .
HEARING RESEARCH, 2006, 212 (1-2) :160-175
[10]   An Evidence-Based Algorithm for Intraoperative Monitoring During Cochlear Implantation [J].
Cosetti, Maura K. ;
Troob, Scott H. ;
Latzman, Jonathan M. ;
Shapiro, William H. ;
Roland, John Thomas, Jr. ;
Waltzman, Susan B. .
OTOLOGY & NEUROTOLOGY, 2012, 33 (02) :169-176