Do flat detector cardiac X-ray systems convey advantages over image-intensifier-based systems? Study comparing X-ray dose and image quality

被引:26
作者
Davies, Andrew G.
Cowen, Arnold R.
Kengyelics, Stephen M.
Moore, Janet
Sivananthan, Mohan U.
机构
[1] Univ Leeds, Acad Unit Med Phys, Leeds LS2 9JT, W Yorkshire, England
[2] Gen Infirm, Yorkshire Heart Ctr, Leeds LS1 3EX, W Yorkshire, England
关键词
X-ray dose; image quality; angiography; image-intensifier system; flat-panel detector;
D O I
10.1007/s00330-006-0458-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The recent introduction of "flat-panel detector" (FD)-based cardiac catheterisation laboratories should offer improvements in image quality and/or dose efficiency over X-ray systems of conventional design. We compared three X-ray systems, one image-intensifier (II)-based system (system A), and two FD-based designs (systems B and C), assessing their image quality and dose efficiency. Phantom measurements were performed to assess dose rates in fluoroscopy and cine acquisition. Phantom dose rates were broadly similar for all systems, with all systems classified as offering "low" dose rates in fluoroscopy on standard phantoms. Patient X-ray dose rate and subjective image quality was assessed for 90 patients. Dose area product (DAP) rates were similar for all systems, except system C, which had a lower DAP rate in fluoroscopy. In terms of subjective image quality, the order of preference was (best to worst): system C, system A, system B. This study indicates that the use of an FD detector does not infer an automatic improvement in image quality or dose efficiency over II based designs. Specification and configuration of all of the components in the X-ray system contribute to the dose levels used and image quality achieved.
引用
收藏
页码:1787 / 1794
页数:8
相关论文
共 19 条
[1]   Design and performance of a high-quality cardiac flat detector [J].
Busse, F ;
Rütten, W ;
Sandkamp, B ;
Alving, PL ;
Bastiaens, RJM ;
Ducourant, T .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :819-827
[2]   Does digital acquisition reduce patients' skin dose in cardiac interventional procedures? An experimental study [J].
Chida, K ;
Saito, H ;
Zuguchi, M ;
Shirotori, K ;
Kumagai, S ;
Nakayama, H ;
Matsubara, K ;
Kohzuki, M .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (04) :1111-1114
[3]   Threshold contrast detail detectability measurement of the fluoroscopic image quality of a dynamic solid-state digital x-ray image detector [J].
Davies, AG ;
Cowen, AR ;
Kengyelics, SM ;
Bury, RF ;
Bruijns, TJ .
MEDICAL PHYSICS, 2001, 28 (01) :11-15
[4]  
DOWLING A, 2003, EQUIPMENT STANDARDS
[5]   Fluoroscopy: Recording of fluoroscopic images and automatic exposure control [J].
Geise, RA .
RADIOGRAPHICS, 2001, 21 (01) :227-236
[6]   Performance of a flat panel cardiac detector [J].
Granfors, PR ;
Albagli, D ;
Tkaczyk, JE ;
Aufrichtig, R ;
Netel, H ;
Brunst, G ;
Boudry, J ;
Luo, DS .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :77-86
[7]   Reference values for diagnostic radiology: Application and impact [J].
Gray, JE ;
Archer, BR ;
Butler, PF ;
Hobbs, BB ;
Mettler, FA ;
Pizzutiello, RJ ;
Schueler, BA ;
Strauss, KJ ;
Suleiman, OH ;
Yaffe, MJ .
RADIOLOGY, 2005, 235 (02) :354-358
[8]   Flat-panel detectors in the cardiac catheterization laboratory: Revolution or evolution - What are the issues? [J].
Holmes, DR ;
Laskey, WK ;
Wondrow, MA ;
Cusma, JT .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (03) :324-330
[9]  
Institute of Physics and Engineering in Medicine, 2002, MED DENT GUID NOT GO
[10]   Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures [J].
Kuon, E ;
Glaser, C ;
Dahm, JB .
BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (906) :406-413