Eigencolor radiochromic film dosimetry

被引:4
作者
Bouchard, Hugo [1 ,2 ,3 ]
Billas, Ilias [4 ]
Subiel, Anna [4 ]
Duane, Simon [4 ]
机构
[1] Univ Montreal, Dept Phys, Complexe Sci,1375 Ave Therese Lavoie Roux, Montreal, PQ H2V 0B3, Canada
[2] Ctr Rech CHUM, 900 Rue St Denis, Montreal, PQ H2X 3H8, Canada
[3] Ctr Hosp Univ Montreal CHUM, Dept Radiooncol, 1051 Rue Sanguinet, Montreal, PQ H2X 3E4, Canada
[4] Natl Phys Lab, Chem Med & Environm Sci Dept, Hampton Rd, Teddington TW11 0LW, Middx, England
关键词
eigencolor; multichannel method; nonstandard beams; radiation dosimetry; radiochromic film; small fields; uncertainty analysis;
D O I
10.1002/mp.14742
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The goal of this work is to propose a new multichannel method correcting for systematic thickness disturbances and to evaluate its precision in relevant radiation dosimetry applications. Methods The eigencolor ratio technique is introduced and theoretically developed to provide a method correcting for thickness disturbances. The method is applied to EBT3 Gafchromic(TM) film irradiated with cobalt-60 and 6 MV photon beams and digitized with an Epson 10000XL photo scanner. Dose profiles and output factors of different field sizes are measured and analyzed. Variance analysis of the previous method of Bouchard et al. ["On the characterization and uncertainty analysis of radiochromic film dosimetry" Med Phys. 2009;36:1931-1946] is adapted to the new approach. Uncertainties are predicted for relevant applications. Results Results show that systematic disturbances attributed to thickness variations are efficiently corrected. The method is shown efficient to identify and correct for dark spots which cause systematic errors in single-channel distributions. Applications of the method in the context of relative dosimetry yields standard uncertainties ranging between 0.8% and 1.9%, depending on the region of interest (ROI) size and the film irradiation. Variance analysis predicts that uncertainty levels between 0.3% and 0.6% are achievable with repeated measurements. Uncertainties are found to vary with absorbed dose and ROI size. Conclusions The proposed multichannel method is efficient for accurate dosimetry, reaching uncertainty levels comparable to previous publications with EBT film. The method is also promising for applications beyond clinical QA, such as machine characterization and other advanced dosimetry applications.
引用
收藏
页码:2592 / 2603
页数:12
相关论文
共 35 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]   Gafgui: An Open-Source Project for Radiochromic Film Dosimetry [J].
Bouchard, H. ;
Lacroix, F. .
MEDICAL PHYSICS, 2010, 37 (06)
[3]  
Bouchard H., GAFGUI
[4]   On the characterization and uncertainty analysis of radiochromic film dosimetry [J].
Bouchard, Hugo ;
Lacroix, Frederic ;
Beaudoin, Gilles ;
Carrier, Jean-Francois ;
Kawrakow, Iwan .
MEDICAL PHYSICS, 2009, 36 (06) :1931-1946
[5]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[6]   Calibration of EBT2 film by the PDD method with scanner non-uniformity correction [J].
Chang, Liyun ;
Chui, Chen-Shou ;
Ding, Hueisch-Jy ;
Hwang, Ing-Ming ;
Ho, Sheng-Yow .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (18) :5875-5887
[7]   Investigation of three radiation detectors for accurate measurement of absorbed dose in nonstandard fields [J].
Chung, Eunah ;
Bouchard, Hugo ;
Seuntjens, Jan .
MEDICAL PHYSICS, 2010, 37 (06) :2404-2413
[8]   Technical Note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry [J].
Desroches, Joannie ;
Bouchard, Hugo ;
Lacroix, Frederic .
MEDICAL PHYSICS, 2010, 37 (04) :1565-1570
[9]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[10]   Reference dosimetry using radiochromic film [J].
Girard, Frederic ;
Bouchard, Hugo ;
Lacroix, Frederic .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (06) :339-353