Improved control over implant anchorage under the use of the femoral neck system for fixation of femoral neck fractures: a technical note

被引:22
作者
Cha, Yonghan [1 ]
Song, Ji-Ung [2 ]
Yoo, Jun-Il [3 ]
Park, Ki Hoon [4 ]
Kim, Jung-Taek [4 ]
Park, Chan Ho [5 ]
Choy, Won-Sik [1 ]
机构
[1] Eulji Univ Hosp, Dept Orthopaed Surg, Daejeon, South Korea
[2] Chamjoen Hosp, Dept Orthopaed Surg, Gwangju, South Korea
[3] Gyeongsang Natl Univ Hosp, Dept Orthopaed Surg, Jinju, South Korea
[4] Ajou Univ, Sch Med, Dept Orthoped Surg, Ajou Med Ctr, 164 World Cup Ro, Suwon 16499, Gyeonggi Do, South Korea
[5] Yeungnam Univ, Med Ctr, Dept Orthopaed Surg, Daegu, South Korea
关键词
Femoral neck fracture; Treatment; Fixation; Femoral neck system; Hip; INTRACAPSULAR FRACTURES; INTERNAL-FIXATION; MANAGEMENT; SCREW; COMPLICATIONS;
D O I
10.1186/s12891-021-04497-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background The depth of bolt in Femoral neck system (FNS, DePuy Synthes, Oberdorf, Switzerland) is difficult to finely control as the length of the bolt is in units of 5 mm. Thus, this study introduces a method to control the depth of FNS bolt in analogue scale in patients with femoral neck fracture. Methods By the technique of control of reaming and retraction of bolt, the tip of implant could be positioned close to subchondral bone without harming it. The position of implant tip in four cases in which the introduced technique was applied was compared to that of eight cases where the standard technique was performed. Results The average tip-apex distance measured in the cases that underwent surgery using the suggested technique in this study was statistically significantly shorter than that measured in the cases that underwent surgery under manufacturer guidelines. Conclusion Even though the bolt of FNS is manufactured in the unit of 5 mm, the technique proposed in this study helps surgeons to adjust the depth of bolt for the fixation of femoral neck fracture using FNS.
引用
收藏
页数:8
相关论文
共 17 条
[1]   THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[2]   Internal fixation compared with arthroplasty for displaced fractures of the femoral neck - A meta-analysis [J].
Bhandari, M ;
Devereaux, PJ ;
Swiontkowski, MF ;
Tornetta, P ;
Obremskey, W ;
Koval, KJ ;
Nork, S ;
Sprague, S ;
Schemitsch, EH ;
Guyatt, GH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (09) :1673-1681
[3]   Optimal Internal Fixation for Femoral Neck Fractures: Multiple Screws or Sliding Hip Screws? [J].
Bhandari, Mohit ;
Tornetta, Paul, III ;
Hanson, Beate ;
Swiontkowski, Marc F. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (06) :403-407
[4]   Biomechanical Evaluation of Internal Fixation of Pauwels Type III Femoral Neck Fractures: A Systematic Review of Various Fixation Methods [J].
Cha, Yong-Han ;
Yoo, Jun-Il ;
Hwang, Seok-Young ;
Kim, Kap-Jung ;
Kim, Ha-Yong ;
Choy, Won-Sik ;
Hwang, Sun-Chul .
CLINICS IN ORTHOPEDIC SURGERY, 2019, 11 (01) :1-14
[5]   Early mechanical complications following fixation of proximal femur fractures: From prevention to treatment [J].
Ehlinger, Matthieu ;
Favreau, Henri ;
Eichler, David ;
Adam, Philippe ;
Bonnomet, Francois .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2020, 106 (01) :S79-S87
[6]   Internal Screw Fixation Compared with Bipolar Hemiarthroplasty for Treatment of Displaced Femoral Neck Fractures in Elderly Patients [J].
Gjertsen, J. -E. ;
Vinje, T. ;
Engesaeter, L. B. ;
Lie, S. A. ;
Havelin, L. I. ;
Furnes, O. ;
Fevang, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (03) :619-628
[7]   The complications of displaced intracapsular fractures of the hip [J].
Gurusamy, K ;
Parker, MJ ;
Rowlands, TK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (05) :632-634
[8]  
Hossain S, 2001, J BONE JOINT SURG BR, V83B, P618, DOI 10.1302/0301-620X.83B4.0830618
[9]   EARLY LOSS OF FIXATION OF FEMORAL-NECK FRACTURES - COMPARISON OF 3 DEVICES IN 244 CASES [J].
HUSBY, T ;
ALHO, A ;
NORDSLETTEN, L ;
BUGGE, W .
ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (01) :69-72
[10]   CORTICAL SCREW SUPPORT IN FEMORAL-NECK FRACTURES - A RADIOGRAPHIC ANALYSIS OF 87 FRACTURES WITH A NEW MENSURATION TECHNIQUE [J].
LINDEQUIST, S .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (03) :289-293