Artifact reduction in true-FISP imaging of the coronary arteries by adjusting imaging frequency

被引:62
作者
Deshpande, VS
Shea, SM
Li, DB
机构
[1] Northwestern Univ, Dept Radiol, Chicago, IL 60611 USA
[2] Northwestern Univ, Dept Biomed Engn, Evanston, IL USA
关键词
magnetic resonance imaging (MRI); fast imaging; steady-state free precession (SSFP); frequency adjustment; MR angiography; coronary arteries;
D O I
10.1002/mrm.10442
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The presence of resonance frequency offsets often causes artifacts in images acquired with true fast imaging with steady-state precession (true-FISP). One source of resonance offsets is a suboptimal setting of the synthesizer frequency. The goal of this work was to demonstrate that shifting the synthesizer frequency could minimize the off-resonance related image artifacts in true-FISP. A simple scouting method was developed to estimate the optimal synthesizer frequency for the volume of interest (VOI). To improve fat suppression, a similar scouting method was also developed to determine the optimal frequency offset for the fat saturation pulse. Coronary artery imaging was performed in healthy subjects using a 3D true-FISP sequence to validate the effectiveness of the frequency corrections. Substantial reduction in image artifacts and improvement in fat suppression were observed by using the water and fat frequencies estimated by the scouting scans. Frequency shifting is a useful and practical method for improving coronary artery imaging using true-FISP. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:803 / 809
页数:7
相关论文
共 8 条
  • [1] Cine MRI using steady state free precession in the radial long axis orientation is a fast accurate method for obtaining volumetric data of the left ventricle
    Bloomer, TN
    Plein, S
    Radjenovic, A
    Higgins, DM
    Jones, TR
    Ridgway, JP
    Sivananthan, MU
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (06) : 685 - 692
  • [2] Cine MR angiography of the heart with segmented true fast imaging with steady-state precession
    Carr, JC
    Simonetti, O
    Bundy, J
    Li, DB
    Pereles, S
    Finn, JP
    [J]. RADIOLOGY, 2001, 219 (03) : 828 - 834
  • [3] Reduction of transient signal oscillations in True-FISP using a linear flip angle series magnetization preparation
    Deshpande, VS
    Chung, YC
    Zhang, Q
    Shea, SM
    Li, DB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) : 151 - 157
  • [4] 3D magnetization-prepared true-FISP: A new technique for imaging coronary arteries
    Deshpande, VS
    Shea, SM
    Laub, G
    Simonetti, OP
    Finn, JP
    Li, DB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) : 494 - 502
  • [5] A method to improve the B-0 homogeneity of the heart in vivo
    Jaffer, FA
    Wen, H
    Balaban, RS
    Wolff, SD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (03) : 375 - 383
  • [6] NISHIMURA DG, 2000, P 8 ANN M ISMRM DENV, P301
  • [7] In vivo measurement of T*2 and field inhomogeneity maps in the human heart at 1.5 T
    Reeder, SB
    Faranesh, AZ
    Boxerman, JL
    McVeigh, ER
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (06) : 988 - 998
  • [8] MOTION-INSENSITIVE, STEADY-STATE FREE PRECESSION IMAGING
    ZUR, Y
    WOOD, ML
    NEURINGER, LJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (03) : 444 - 459