The effect of ion implantation on the wear of Co-Cr-Mo alloy

被引:11
作者
Werner, Z. [1 ]
Barlak, M.
Gradzka-Dahlke, M.
Diduszko, R.
Szymczyk, W.
Dabrowski, J.
Piekoszewski, J.
Borkowska, K.
机构
[1] Andrzej Soltan Inst Nucl Studies, PL-05400 Otwock, Poland
[2] Bialystok Tech Univ, PL-15351 Bialystok, Poland
[3] Inst Chem Phys, PAS, PL-01224 Warsaw, Poland
[4] Ind Inst Elect, PL-01241 Warsaw, Poland
[5] Inst Elect Mat Technol, PL-01919 Warsaw, Poland
[6] Inst Nucl Chem & Technol, PL-03145 Warsaw, Poland
关键词
Co-Cr-Mo alloys; ion implantation; wear;
D O I
10.1016/j.vacuum.2007.01.014
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The effect of C+N and O ion implantation upon tribological properties of Co-Cr-Mo alloy was studied for fluence range 3-7 x 10(17)cm(-2) and implantation temperature range 200-400 degrees C. Phase structure changes and implanted ion distributions were determined. A marked improvement of the wear rate was noticed. Higher implantation temperatures reduce the improvement. No traces of nitride or carbide phases have been found but chromium oxide phase has been detected after O implantation. (c) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1191 / 1194
页数:4
相关论文
共 9 条
  • [1] Plasma-surface modification of biomaterials
    Chu, PK
    Chen, JY
    Wang, LP
    Huang, N
    [J]. MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2002, 36 (5-6) : 143 - 206
  • [2] DABROWSKI JR, 2002, COMPOSITES, V2, P167
  • [3] Formation of Hcp martensite during the isothermal aging of an Fcc Co-27Cr-5Mo-0.05C orthopedic implant alloy
    García, ADJS
    Medrano, AM
    Rodríguez, AS
    [J]. METALLURGICAL AND MATERIALS TRANSACTIONS A-PHYSICAL METALLURGY AND MATERIALS SCIENCE, 1999, 30 (05): : 1177 - 1184
  • [4] Effects of grain size on development of athermal and strain induced ε martensite in Co-Cr-Mo implant alloy
    Huang, P
    López, HF
    [J]. MATERIALS SCIENCE AND TECHNOLOGY, 1999, 15 (02) : 157 - 164
  • [5] OKSIUTA Z, 2002, COMPOSITES, V2, P195
  • [6] Wear reduction effect on ultra-high-molecular-weight polyethylene by application of hard coatings and ion implantation on cobalt chromium alloy, as measured in a knee wear simulation machine
    Oñate, JI
    Comin, M
    Braceras, I
    Garcia, A
    Viviente, JL
    Brizuela, M
    Garagorri, N
    Peris, JL
    Alava, JI
    [J]. SURFACE & COATINGS TECHNOLOGY, 2001, 142 : 1056 - 1062
  • [7] Surface treatment of biomaterials by ion beam processes
    Sioshansi, P
    Tobin, EJ
    [J]. SURFACE & COATINGS TECHNOLOGY, 1996, 83 (1-3) : 175 - 182
  • [8] SIOSHANSI P, 1991, ORTHOPEDICS TODAY, V11, P1
  • [9] Carbide surface coating of Co-Cr-Mo implant alloys by a microwave plasma-assisted reaction
    Vandamme, NS
    Que, L
    Topoleski, LDT
    [J]. JOURNAL OF MATERIALS SCIENCE, 1999, 34 (14) : 3525 - 3531