共 3 条
Reinforcement of left ventricular assist device outflow grafts to prevent kinking
被引:9
作者:
Cohn, William E.
Gregoric, Igor D.
Frazier, O. H.
机构:
[1] St Lukes Episcopal Hosp, Texas Heart Inst, Ctr Cardiac Support, Houston, TX 77225 USA
[2] Baylor Coll Med, Michael E DeBakey Dept Surg, Div Cardiothorac Surg, Houston, TX 77030 USA
关键词:
D O I:
10.1016/j.athoracsur.2006.10.015
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Because of their small size, rotary left ventricular assist devices (LVADs) can be implanted through several alternative approaches, including transdiaphragmatic LVAD insertion through a left subcostal incision with anastomosis of the outflow graft to the retroperitoneal supraceliac aorta and a left thoracotomy approach with anastomosis to the descending thoracic aorta. More recently we have added a counter-incision to allow the outflow graft of a transdiaphragmatic LVAD to be tunneled through the right chest and anastomosed to the ascending aorta. However, constructing a tension-free, nonkinking lie of the outflow graft can be challenging. We have found that placing a 10-cm to 15-cm length of polytetrafluoroethylene graft coaxially over the LVAD outflow graft ensures a smooth, kink-free lie. Thus far 12 patients have undergone reinforcement of the LVAD outflow graft prior to graft-to-aorta anastomosis. In all cases, graft lie was facilitated and kinking was eliminated.
引用
收藏
页码:301 / 302
页数:2
相关论文