The Scope Of Three-Dimensional Digital Visualization Systems In Vitreoretinal Surgery

被引:43
作者
Agranat, Joshua S. [1 ]
Miller, John B. [1 ]
Douglas, Vivian Paraskevi [1 ]
Douglas, Konstantinos A. A. [1 ]
Marmalidou, Anna [1 ]
Cunningham, Matthew A. [2 ]
Houston, Samuel K., III [2 ]
机构
[1] Harvard Med Sch, Dept Ophthalmol, Retina Serv, Massachusetts Eye & Ear, Boston, MA 02115 USA
[2] Florida Retina Inst, 95 Columbia St, Orlando, FL 32806 USA
来源
CLINICAL OPHTHALMOLOGY | 2019年 / 13卷
关键词
heads-up display; 3-D Visualization; NGENUITY; retina; surgery;
D O I
10.2147/OPTH.S213834
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Introduction: Significant advances in three-dimensional (3D) imaging technology have allowed for the incorporation of 3D digital displays into medical and surgical devices. Despite initial adoption of the NGENUITY (R) 3D Visualization System in vitreoretinal surgery, there are limited publications regarding its use. The generally accepted main benefits include improved ergonomics, enhanced surgical team communication and education, reduced retinal phototoxicity, increased depth of field, and display image manipulation. Despite these potential benefits, many retina specialists have questioned its universal applicability to a wide variety of vitreoretinal surgeries. Objective: To report on the variety of indications and surgical efficacy of the NGENUITY (R) 3D Visualization System in vitreoretinal surgery via a review of surgical experience at two vitreoretinal practices in both the academic and community settings. Methods: A retrospective review was conducted of consecutive surgical cases performed on the NGENUITY (R) 3D Visualization System at Massachusetts Eye and Ear Infirmary and Florida Retina Institute from June 1st, 2017 to November 1st, 2018. Age, presenting diagnosis, surgical procedure, and intraoperative details were recorded. Results: 272 vitreoretinal surgeries on the Alcon NGENUITY (R) 3D Visualization System were identified between June 1st, 2017 and November 1st, 2018 at the participating institutions. A detailed breakdown of the indications for surgery and related procedures is reported. During all 272 cases on the 3D digital system, there were no complications attributed to the visualization system. Conclusion: This series illustrates the diversity of vitreoretinal surgeries that can be performed on this system without compromising surgical viewing or increasing surgical complications. The Alcon NGENUITY (R) 3D Visualization System possesses favorable ergonomics, illumination levels, depth of field, display filters, and trainee experience.
引用
收藏
页码:2093 / 2096
页数:4
相关论文
共 11 条
  • [1] Adam MK, 2017, RETINA-J RET VIT DIS, V37, P1746, DOI 10.1097/IAE.0000000000001420
  • [2] COMPARISON OF 2D AND 3D VIDEO DISPLAYS FOR TEACHING VITREORETINAL SURGERY
    Chhaya, Nisarg
    Helmy, Omar
    Piri, Niloofar
    Palacio, Agustina
    Schaal, Shlomit
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2018, 38 (08): : 1556 - 1561
  • [3] Three-Dimensional Head-Mounted Display System for Ophthalmic Surgical Procedures
    Dutra-Medeiros, Marco
    Nascimento, Joao
    Henriques, Jose
    Barrao, Sandra
    Fernandes-Fonseca, Ana
    Aguiar-Silva, Nuno
    Moura-Coelho, Nuno
    Agoas, Victor
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2017, 37 (07): : 1411 - 1414
  • [4] HEADS-UP SURGERY FOR VITREORETINAL PROCEDURES An Experimental and Clinical Study
    Eckardt, Claus
    Paulo, Erica B.
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2016, 36 (01): : 137 - 147
  • [5] THE INTEGRATIVE SURGICAL THEATER Combining Intraoperative Optical Coherence Tomography and 3D Digital Visualization for Vitreoretinal Surgery in the DISCOVER Study
    Ehlers, Justis P.
    Uchida, Atsuro
    Srivastava, Sunil K.
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2018, 38 : S88 - S96
  • [6] Ittyerah T P, 1983, Indian J Ophthalmol, V31 Suppl, P1057
  • [7] Comparison of clinical outcomes between "heads-up" 3D viewing system and conventional microscope in macular hole surgeries: A pilot study
    Kumar, Atul
    Hasan, Nasiq
    Kakkar, Prateek
    Mutha, Vineet
    Karthikeya, R.
    Sundar, Dheepak
    Ravani, Raghav
    [J]. INDIAN JOURNAL OF OPHTHALMOLOGY, 2018, 66 (12) : 1816 - 1819
  • [8] Heads-up 3D Microscopy: An Ergonomic and Educational Approach to Microsurgery
    Mendez, Bernardino M.
    Chiodo, Michael V.
    Vandevender, Darl
    Patel, Parit A.
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY-GLOBAL OPEN, 2016, 4 (05)
  • [9] Argus II Retinal Prosthesis Implantation Using Three-Dimensional Visualization System
    Rachitskaya, Aleksandra
    Lane, Leslie
    Ehlers, Justis
    Debenedictis, Meghan
    Yuan, Alex
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2019, 39 : 199 - 200
  • [10] Riemann CD, 2010, AM AC OPHTH ANN M OC