Accuracy and consistency of intensity-based deformable image registration in 4DCT for tumor motion estimation in liver radiotherapy planning

被引:6
|
作者
Tascon-Vidarte, Jose D. [1 ]
Stick, Line Bjerregaard [2 ]
Josipovic, Mirjana [2 ]
Risum, Signe [2 ]
Jomier, Julien [3 ]
Erleben, Kenny [1 ]
Vogelius, Ivan Richter [2 ]
Darkner, Sune [1 ]
机构
[1] Univ Copenhagen, Dept Comp Sci, Copenhagen, Denmark
[2] Dept Oncol, Copenhagen, Denmark
[3] Kitware SAS, Villeurbanne, France
来源
PLOS ONE | 2022年 / 17卷 / 07期
关键词
STEREOTACTIC BODY RADIOTHERAPY; RADIATION-THERAPY; TARGET; CT; MANAGEMENT; REDUCTION; FRAMEWORK; MODEL;
D O I
10.1371/journal.pone.0271064
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
We investigate the accuracy of intensity-based deformable image registration (DIR) for tumor localization in liver stereotactic body radiotherapy (SBRT). We included 4DCT scans to capture the breathing motion of eight patients receiving SBRT for liver metastases within a retrospective clinical study. Each patient had three fiducial markers implanted. The liver and the tumor were delineated in the mid-ventilation phase, and their positions in the other phases were estimated with deformable image registration. We tested referenced and sequential registrations strategies. The fiducial markers were the gold standard to evaluate registration accuracy. The registration errors related to measured versus estimated fiducial markers showed a mean value less than 1.6mm. The positions of some fiducial markers appeared not stable on the 4DCT throughout the respiratory phases. Markers' center of mass tends to be a more reliable measurement. Distance errors of tumor location based on registration versus markers center of mass were less than 2mm. There were no statistically significant differences between the reference and the sequential registration, i.e., consistency and errors were comparable to resolution errors. We demonstrated that intensitybased DIR is accurate up to resolution level for locating the tumor in the liver during breathing motion.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Validation of a Deformable Image Registration Method to Assess Lung Ventilation From 4DCT
    Laplante, C.
    Bedwani, S.
    Carrier, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E562 - E562
  • [22] Ventilation Series Similarity: A Study for Ventilation Calculation Using Deformable Image Registration and 4DCT to Avoid Motion Artifacts
    Zhang, Geoffrey G.
    Latifi, Kujtim
    Feygelman, Vladimir
    Chou, Kuei-Ting
    Huang, Tzung-Chi
    Dilling, Thomas J.
    Perez, Bradford A.
    Moros, Eduardo G.
    CONTRAST MEDIA & MOLECULAR IMAGING, 2017,
  • [23] Scale-Adaptive Convolutional Neural Network for Deformable Image Registration of Lung 4DCT
    Sang, Y.
    Ruan, D.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [24] GPU-based Deformable Image Registration for Internal Target Volume Generation using 4DCT Imaging
    Samant, S.
    Lynch, B.
    Zlotecki, R.
    Xia, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S524 - S525
  • [25] Comparison of Demons and Optical Flow Deformable Image Registration Algorithms on Clinical 4DCT Images
    Durghalli, A.
    Song, W.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [26] A multivariable study of deformable image registration evaluation metrics in 4DCT of thoracic cancer patients
    Juan-Cruz, Celia
    Fast, Martin F.
    Sonke, Jan-Jakob
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03):
  • [27] Lobar-Level Lung Ventilation Analysis Using 4DCT and Deformable Image Registration
    Du, K.
    Patton, T.
    Reinhardt, J.
    Christensen, G.
    Zhao, B.
    Gerard, S.
    Pan, Y.
    Bayouth, J.
    MEDICAL PHYSICS, 2015, 42 (06) : 3284 - 3285
  • [28] Deformable lung 4DCT image registration via landmark-driven cycle network
    Matkovic, Luke
    Lei, Yang
    Fu, Yabo
    Wang, Tonghe
    Kesarwala, Aparna H.
    Axente, Marian
    Roper, Justin
    Higgins, Kristin
    Bradley, Jeffrey D.
    Liu, Tian
    Yang, Xiaofeng
    MEDICAL PHYSICS, 2024, 51 (03) : 1974 - 1984
  • [29] Relative accuracy of spin-image-based registration of partial capitate bones in 4DCT of the wrist
    Breighner, Ryan E.
    Holmes, David R., III
    Leng, Shuai
    An, Kai-Nan
    McCollough, Cynthia H.
    Zhao, Kristin D.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING-IMAGING AND VISUALIZATION, 2016, 4 (06): : 360 - 367
  • [30] Use of Maximum Intensity Projections (MIPs) for Target Outlining in 4DCT Radiotherapy Planning
    Muirhead, Rebecca
    McNee, Stuart G.
    Featherstone, Carrie
    Moore, Karen
    Muscat, Sarah
    JOURNAL OF THORACIC ONCOLOGY, 2008, 3 (12) : 1433 - 1438