Validation of a near-infrared probe for detection of thin intracranial white matter structures

被引:36
作者
Giller, CA
Liu, RL
Gurnani, P
Victor, S
Yazdani, U
German, DC
机构
[1] Univ Texas, SW Med Ctr, Dept Neurosurg, Dallas, TX 75390 USA
[2] Univ Texas, SW Med Ctr, Dept Psychiat, Dallas, TX 75390 USA
[3] Univ Texas, Dept Bioengn, Arlington, TX 76019 USA
关键词
near-infrared spectroscopy; deep brain stimulation; stereotactic surgery; rat;
D O I
10.3171/jns.2003.98.6.1299
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The authors have developed an intracranial near-infrared (NIR) probe that analyzes the scattering of light emitted from its tip to measure the optical properties of cerebral tissue. Despite its success in distinguishing gray matter from white matter in humans during stereotactic surgery, the limits of this instrument's resolution remain unclear. In this study, the authors determined the spatial resolution of this new probe by using a rodent model supplemented with phantom measurements and computer simulation. Methods. A phantom consisting of Intralipid and gelatin was constructed to resemble a layer of white matter overlying a layer of gray matter. Near-infrared measurements were obtained as the probe was inserted through the gray-white matter transition. A computer simulation of NIR measurements through a gray-white matter transition was also performed using Monte Carlo techniques. The NIR probe was then used to study 19 tracks from the cortical surface through the corpus callosum in an in vivo rodent preparation. The animals were killed and histological sections through the tracks were obtained. Data from the phantom models and computer simulations showed that the NIR probe samples a volume of tissue extending 1 to 1.5 mm in front of the probe tip (this distance is termed the "lookthrough" distance). Measurements obtained from an NIR probe passing through a thin layer of white matter consisted of an initial segment of increasing values, a maximum (peak) value, and a trailing segment of decreasing values. The length of the initial segment is the lookthrough distance, the position of the peak indicates the location of the superficial white matter boundary, and the length of the trailing segment is the thickness of the layer. These considerations were confirmed in experiments with rodents. All tracks passed through the corpus callosum, which was demonstrated as a broad peak on each NIR graph. The position of the dorsal boundary of the corpus callosum and its width (based on histological measurements) correlated well with the peak of the NIR curve and its trailing segment, respectively. The initial segments correlated well with estimates of the lookthrough distance. Five of the tracks transected the smaller anterior commissure (diameter 0.2 mm), producing a narrow NIR peak at the correct depth. Conclusions. Data in this study confirm that the NIR probe can reliably detect and measure the thickness of layers of white matter as thin as 0.2 mm. Such resolution should be adequate to detect larger structures of interest encountered during stereotactic surgery in humans.
引用
收藏
页码:1299 / 1306
页数:8
相关论文
共 25 条
  • [1] Pallidotomy for Parkinson disease: a review of contemporary literature
    Alkhani, A
    Lozano, AM
    [J]. JOURNAL OF NEUROSURGERY, 2001, 94 (01) : 43 - 49
  • [2] [Anonymous], P SPIE IS
  • [3] [Anonymous], RAT BRAIN STEREOTAXI
  • [4] Long-term follow-up of unilateral pallidotomy in advanced Parkinson's disease
    Fine, J
    Duff, J
    Chen, R
    Hutchison, W
    Lozano, AM
    Lang, AE
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (23) : 1708 - 1714
  • [5] Use of an intracranial near-infrared probe for localization during stereotactic surgery for movement disorders - Technical note
    Giller, CA
    Johns, M
    Liu, HL
    [J]. JOURNAL OF NEUROSURGERY, 2000, 93 (03) : 498 - 505
  • [6] Variability in lesion location after microelectrode-guided pallidotomy for Parkinson's disease: anatomical, physiological, and technical factors that determine lesion distribution
    Gross, RE
    Lombardi, WJ
    Hutchison, WD
    Narula, S
    Saint-Cyr, JA
    Dostrovsky, JO
    Tasker, RR
    Lang, AE
    Lozano, AM
    [J]. JOURNAL OF NEUROSURGERY, 1999, 90 (03) : 468 - 477
  • [7] A brief history of pallidotomy
    Guridi, J
    Lozano, AM
    [J]. NEUROSURGERY, 1997, 41 (05) : 1169 - 1180
  • [8] Complications of deep brain stimulation surgery
    Hariz, MI
    [J]. MOVEMENT DISORDERS, 2002, 17 : S162 - S166
  • [9] Do microelectrode techniques increase accuracy or decrease risks in pallidotomy and deep brain stimulation? A critical review of the literature
    Hariz, MI
    Fodstad, H
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1999, 72 (2-4) : 157 - 169
  • [10] Neurophysiological identification of the subthalamic nucleus in surgery for Parkinson's disease
    Hutchison, WD
    Allan, RJ
    Opitz, H
    Levy, R
    Dostrovsky, JO
    Lang, AE
    Lozano, AM
    [J]. ANNALS OF NEUROLOGY, 1998, 44 (04) : 622 - 628