Polytetrafluoroethylene coating fragments during neuroendovascular therapy: An analysis of two damaged microguidewires

被引:7
作者
Dahl, Rasmus Holmboe [1 ,2 ]
Larsen, Rene Wugt [3 ]
Thormann, Esben [3 ]
Benndorf, Goetz [1 ,4 ]
机构
[1] Rigshosp, Dept Radiol, Univ Hosp, DK-2100 Copenhagen, Denmark
[2] Naestved Slagelse Hosp, Dept Radiol, Slagelse, Denmark
[3] Tech Univ Denmark, Dept Chem, Kongens, Denmark
[4] Baylor Coll Med, Dept Radiol, Houston, TX 77030 USA
关键词
Attenuated-total-reflection Fourier transform infrared spectroscopy; hydrophilic polymers; microguidewire coating; polymer coating embolism; polytetrafluoroethylene; silicone oil; COMPLICATIONS; EMBOLIZATION;
D O I
10.1177/15910199211015127
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Cerebral polymer coating embolism from intravascular devices represents a potentially serious complication to endovascular therapy (EVT). We report two cases of neuroendovascular treatment where filamentous polymer fragments were noted possibly due to damage of the surface coating during manipulation and backloading of microguidewires. As the exact origin of the debris was initially not known, microguidewires and fragments were examined with light microscopy, stereomicroscopy, scanning electron microscopy and attenuated-total-reflection Fourier transform infrared spectroscopy. Fragments consisted of polytetrafluoroethylene and silicone oil stemming from the proximal shaft of a standard microguidewire. To our knowledge, this is the first report of polytetrafluoroethylene coating fragments created during EVT. Future studies should assess the mechanism of polymer coating delamination and its potential consequences during EVT including inadvertent fragment migration into the cerebral circulation.
引用
收藏
页码:16 / 21
页数:6
相关论文
共 15 条
[1]   Hydrophilic polymer embolic complication during diagnostic cerebral angiography presenting with delayed intracranial hemorrhage: case report and literature review [J].
Ansari, Sameer A. ;
Anderson, Richard R. ;
Caron, Michael J. ;
Shaibani, Ali ;
Hurley, Michael C. ;
Jahromi, Babak S. ;
Potts, Matthew B. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (01) :80-83
[2]  
Barnwell SL, 1997, AM J NEURORADIOL, V18, P1886
[3]   Polymer coating embolism from intravascular medical devices - a clinical literature review [J].
Chopra, Amitabh M. ;
Mehta, Monik ;
Bismuth, Jean ;
Shapiro, Maksim ;
Fishbein, Michael C. ;
Bridges, Alina G. ;
Vinters, Harry V. .
CARDIOVASCULAR PATHOLOGY, 2017, 30 :45-54
[4]   HISTOLOGICAL RESPONSE TO INJECTED POLYTEF AND BIOPLASTIQUE IN THE SHEEP BRAIN [J].
DEWAN, PA ;
OWEN, AJ ;
BYARD, RW .
BRITISH JOURNAL OF UROLOGY, 1995, 75 (05) :666-669
[5]   Complications of endovascular polymers associated with vascular introducer sheaths and metallic coils in 3 patients, with literature review [J].
Fealey, Michael E. ;
Edwards, William D. ;
Giannini, Caterina ;
Piepgras, David G. ;
Cloft, Harry ;
Rihal, Charanjit S. .
AMERICAN JOURNAL OF SURGICAL PATHOLOGY, 2008, 32 (09) :1310-1316
[6]   Distal Embolization of Hydrophilic-Coating Material From Coronary Guidewires After Percutaneous Coronary Interventions [J].
Grundeken, Maik J. ;
Li, Xiaofei ;
Kurpershoek, C. Eline ;
Kramer, Miranda C. ;
Vink, Aryan ;
Piek, Jan J. ;
Tijssen, Jan G. P. ;
Koch, Karel T. ;
Wykrzykowska, Joanna J. ;
de Winter, Robbert J. ;
van der Wal, Allard C. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2015, 8 (02)
[7]  
HO KJ, 1979, J THORAC CARDIOV SUR, V77, P875
[8]   Histopathological assessment of fatal ipsilateral intraparenchymal hemorrhages after the treatment of supraclinoid aneurysms with the Pipeline Embolization Device [J].
Hu, Yin C. ;
Deshmukh, Vivek R. ;
Albuquerque, Felipe C. ;
Fiorella, David ;
Nixon, Randal R. ;
Heck, Donald V. ;
Barnwell, Stanley L. ;
McDougall, Cameron G. .
JOURNAL OF NEUROSURGERY, 2014, 120 (02) :365-374
[9]  
Kallmes DF, 1997, AM J NEURORADIOL, V18, P1243
[10]  
Kitamura Takayuki, 2020, NMC Case Rep J, V7, P5, DOI 10.2176/nmccrj.cr.2018-0319