Accuracy testing of a new stereotactic CT-guided brain biopsy device in the dog

被引:32
作者
Moissonnier, P [1 ]
Bordeau, W [1 ]
Delisle, F [1 ]
Devauchelle, P [1 ]
机构
[1] Ecole Natl Vet Alfort, F-94700 Maisons Alfort, France
关键词
D O I
10.1053/rvsc.1999.0370
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The accuracy testing of a new CT-guided stereotactic device was undertaken via an experimental study. Ten dog cadaver heads were submitted to a CT examination and a brain target was determined (coordinates: xl, yl, zl). 0.3 mi of a radiopaque paste was injected in this chosen point. The head was analysed twice by CT and the centre of the injection (x2, y2, z2) was measured on slices. Tridimensional statistical analysis of the variance of the two points was carried out. The procedure always led to the injection of the radiopaque paste and created an intracerebral mass which gave a measurable image on cr examination. The distance between the original target and the centre of the injection was 2.9 +/- 1.08 mm. It is concluded that this new device allows the opportunity to perform stereotactic CT-guided brain biopsy for all unknown brain lesions exceeding 6 mm in diameter in the dog.
引用
收藏
页码:243 / 247
页数:5
相关论文
共 19 条
[1]   COMPUTED IMAGING STEREOTAXY - EXPERIENCE AND PERSPECTIVE RELATED TO 500 PROCEDURES APPLIED TO BRAIN MASSES [J].
APUZZO, MLJ ;
CHANDRASOMA, PT ;
COHEN, D ;
ZEE, CS ;
ZELMAN, V .
NEUROSURGERY, 1987, 20 (06) :930-937
[2]   ULTRASOUND-GUIDED STEREOTAXIC BIOPSY USING A NEW APPARATUS [J].
BERGER, MS .
JOURNAL OF NEUROSURGERY, 1986, 65 (04) :550-554
[3]  
Bradford R, 1987, Acta Neurochir Suppl (Wien), V39, P25
[5]  
BROWN RA, 1981, AM J NEURORADIOL, V2, P181
[6]   STEREOTACTIC BIOPSY IN THE DIAGNOSIS OF BRAIN MASSES - COMPARISON OF RESULTS OF BIOPSY AND RESECTED SURGICAL SPECIMEN [J].
CHANDRASOMA, PT ;
SMITH, MM ;
APUZZO, MLJ .
NEUROSURGERY, 1989, 24 (02) :160-165
[7]  
COFFEY RJ, 1987, APPL NEUROPHYSIOL, V50, P81
[8]  
HARARI J, 1993, PROG VET NEUROL, V4, P41
[9]  
HARIZ M I, 1990, British Journal of Neurosurgery, V4, P397, DOI 10.3109/02688699008992762
[10]   STEREOTAXIC BIOPSY - A METHOD FOR STUDY OF DISCRETE BRAIN REGIONS OF ANIMALS AND MAN [J].
HEATH, RG ;
JOHN, S ;
FOSS, O .
ARCHIVES OF NEUROLOGY, 1961, 4 (03) :291-&