Investigating the inaccuracies of indirect method of blood pressure measurement associated to the cuff constructing fabrics

被引:1
作者
Naqvi, Shenela [1 ]
Husain, Muhammad Dawood [1 ]
Lewis, Philip [2 ]
机构
[1] NED Univ Engn & Technol, Text Engn Dept, Karachi 75270, Sindh, Pakistan
[2] Stepping Hill Hosp, Blood Pressure & Heart Res Ctr, Stockport SK2 7JE, Lancs, England
关键词
Blood pressure measurement; BP monitoring; Cuff constructing fabric; Bare metal cylinder; Pressure sensing; HYPERTENSION; SIZE;
D O I
10.22581/muet1982.2202.12
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Numerous cuffs have been manufactured using different materials in which textile fabric is the main constituent. There are no specific guidelines for the selection of the cuff fabric and its design. In this investigation, five different types of cuffs (made of woven and non-woven fabrics) were selected, and their pressure distribution was noted while mimicking the process of blood pressure measurement. It was achieved by measuring pressure at the interface of the selected cuffs and a bare metal cylinder. I-scan, an interface pressure measurement system was employed to record the interface pressure profiles under the cuffs against 140 mmHg to 40 mmHg pressure inside the cuffs. The data obtained from the pressure sensing system was processed in MATLAB to examine pressure distribution at 96 points in detail. The results show that the cuffs registered non-uniform pressure distribution at the interface which is also non-identical among the selected cuffs. The pressure at the interface is found to be lower than the pressure inside the cuff, which shows that pressure attenuates as transfers to the surface of the object underneath it. The range of the pressure difference varies from 10 mmHg to 15 mmHg. This study indicates that the pressure distribution under a cuff depends on the constructing fabric and its properties. It is concluded that there is a need to select appropriate fabric with optimized properties for desired pressure distribution which may lead to accurate estimation of blood pressure.
引用
收藏
页码:127 / 134
页数:8
相关论文
共 14 条
[1]   SYSTEMATIC DIFFERENCE BETWEEN BLOOD-PRESSURE READINGS CAUSED BY CUFF TYPE [J].
BOVET, P ;
HUNGERBUHLER, P ;
QUILINDO, J ;
GRETTVE, ML ;
WAEBER, B ;
BURNAND, B .
HYPERTENSION, 1994, 24 (06) :786-792
[2]   Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients [J].
Bur, A ;
Hirschl, MM ;
Herkner, H ;
Oschatz, E ;
Kofler, J ;
Woisetschläger, C ;
Laggner, AN .
CRITICAL CARE MEDICINE, 2000, 28 (02) :371-376
[3]  
Campbell NRC, 1999, CAN MED ASSOC J, V161, P277
[4]  
CLANCY F, 1978, Journal of Clinical Engineering, V3, P49
[5]  
Corbman B.P., 1983, Textiles: Fiber to Fabric
[6]   Monitoring arterial pressure [J].
Dalrymple, Phil .
ANAESTHESIA AND INTENSIVE CARE MEDICINE, 2006, 7 (03) :93-94
[7]  
Kumar S. Ashish, 2007, COATED TEXTILES PRIN
[8]  
Naqvi N.H., 1998, Blood pressure measurement: An illustrated history
[9]   Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British Hypertension Society [J].
O'Brien, E ;
Coats, A ;
Owens, P ;
Petrie, J ;
Padfield, PL ;
Littler, WA ;
de Swiet, M ;
Mee, F .
BMJ-BRITISH MEDICAL JOURNAL, 2000, 320 (7242) :1128-1134
[10]   Association of Blood Pressure and Cardiovascular Mortality in India: Mumbai Cohort Study [J].
Pednekar, Mangesh S. ;
Gupta, Rajeev ;
Gupta, Prakash C. .
AMERICAN JOURNAL OF HYPERTENSION, 2009, 22 (10) :1076-1084