Stress distribution and microgap formation in angulated zirconia abutments with a titanium base in narrow diameter implants: A 3D finite element analysis

被引:4
作者
Zhang, Yuqiang [1 ]
Yu, Ping [2 ]
Yu, Haiyang [1 ]
机构
[1] Sichuan Univ, Natl Clin Res Ctr Oral Dis, State Key Lab Oral Dis, West China Hosp Stomatol, Chengdu, Peoples R China
[2] Chengdu Second Peoples Hosp, Dept Stomatol, Chengdu, Peoples R China
基金
中国国家自然科学基金;
关键词
abutment design; angled abutment; finite element analysis; stress distribution; titanium; zirconia; DENTAL IMPLANTS; INTERNAL CONNECTION; GAP FORMATION; BONE STRESS; FOLLOW-UP; FRACTURE; MICROMOTION; ANTERIOR; RESTORATIONS; THICKNESSES;
D O I
10.1002/cnm.3610
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This in vitro study aimed to use failure stress and implant abutment interface (IAI) microgap size to find the compromised axial angle range of angulated zirconia abutments with a titanium base in narrow diameter implants in the esthetic region. A three-dimensional (3D) finite element model of maxillary central incisor implant prosthesis was reconstructed. Angulated zirconia abutments (0 degrees, 15 degrees, 30 degrees, and - 15 degrees) with a titanium base in narrow diameter implants (3.3 x 12 mm, Bone level, Roxolid SLActive, Straumann AG, Switzerland) were designed to simulate clinical scenarios of buccal inclination 0 degrees, 15 degrees, and 30 degrees, and palatal inclination 15 degrees of the implant long axis. Straight titanium abutment and pure titanium implant were used as two control groups. An oblique force at 30 degrees inclination to the long axis of the crown was applied 3 mm below the incisal edge on the palatal surface of the prosthesis. Under simulated dynamic chewing force, the stress distribution of the implant components and surrounding bone were investigated. The relative micromotion displacement between the implant and abutment models at the IAI area was recorded, and the influence of tightening torque on the IAI microgap was evaluated. The angulation of the zirconia abutment could affect the stress value and IAI microgap of implant restorations. When the zirconia abutment angle increased from -15 degrees to 30 degrees, the stress on the central screw, titanium base, and surrounding bone tissue gradually increased by 9%, 20%, and 23%, respectively. The stress levels of the 30 degrees zirconia abutment group showed the risk of exceeding the threshold. When the long axis of the implant was inclined in the palatal direction, the -15 degrees angle abutment reduced the stress by 3% and reduced the strain level of the implant system by 17% and the surrounding bone tissue by 26%. Under simulated dynamic chewing load, the displacement between the implants and the abutment occurred in each group of the implant system, and the amplitude of the micromotion fluctuated with the change in the load. The horizontal displacement caused a 0.075-1.459 mu m palatal microgap and 0.091-0.945 mu m distal microgap in the IAI. The microgap between the lip and palate was more evident, and the vertical displacement difference was manifested as the abutment sliding down the implant. In cases of upper implant restoration with difficulties such as small gaps and axial defects in the esthetic zone, the abutment angle is highly recommended to be in a slightly palatal-inclined direction or to not exceed 15 degrees when the implant is inclined to the labial side to avoid mechanical damage and leakage caused by the appearance of excessively large micromotion gaps.
引用
收藏
页数:16
相关论文
共 51 条
  • [1] Dental Implant Length and Diameter: A Proposed Classification Scheme
    Al-Johany, Sulieman S.
    Al Amri, Mohammad D.
    Alsaeed, Suliman
    Alalola, Bassam
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (03): : 252 - 260
  • [2] Bahuguna Rohit, 2013, Natl J Maxillofac Surg, V4, P46, DOI 10.4103/0975-5950.117882
  • [3] Clinical Outcomes of Implant Abutments in the Anterior Region: A Systematic Review
    Bidra, Avinash S.
    Rungruanganunt, Patchanee
    [J]. JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2013, 25 (03) : 159 - 176
  • [4] Fatigue induced changes in conical implant-abutment connections
    Blum, Kai
    Wiest, Wolfram
    Fella, Christian
    Balles, Andreas
    Dittmann, Jonas
    Rack, Alexander
    Maier, Dominik
    Thomann, Ralf
    Spies, Benedikt Christopher
    Kohal, Ralf Joachim
    Zabler, Simon
    Nelson, Katja
    [J]. DENTAL MATERIALS, 2015, 31 (11) : 1415 - 1426
  • [5] Influence of abutment material on the gingival color of implant-supported all-ceramic restorations: a prospective multicenter study
    Bressan, Eriberto
    Paniz, Gianluca
    Lops, Diego
    Corazza, Boris
    Romeo, Eugenio
    Favero, Gianantonio
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (06) : 631 - 637
  • [6] Titanium-zirconium alloy narrow-diameter implants (Straumann Roxolid®) for the rehabilitation of horizontally deficient edentulous ridges: prospective study on 18 consecutive patients
    Chiapasco, M.
    Casentini, P.
    Zaniboni, M.
    Corsi, E.
    Anello, T.
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (10) : 1136 - 1141
  • [7] Experimental evaluation of stress distribution with narrow diameter implants: A finite element analysis
    Cinel, Sezgi
    Celik, Ersan
    Sagirkaya, Elcin
    Sahin, Onur
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (03) : 417 - 425
  • [8] Influence of Tapered and External Hexagon Connections on Bone Stresses Around Tilted Dental Implants: Three-Dimensional Finite Element Method With Statistical Analysis
    de Faria Almeida, Daniel Augusto
    Pellizzer, Eduardo Piza
    Verri, Fellippo Ramos
    Santiago, Joel Ferreira, Jr.
    Perri de Carvalho, Paulo Segio
    [J]. JOURNAL OF PERIODONTOLOGY, 2014, 85 (02) : 261 - 269
  • [9] Implant-Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis
    Ding, Xi
    Zhu, Xing-Hao
    Liao, Sheng-Hui
    Zhang, Xiu-Hua
    Chen, Hong
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2009, 18 (05): : 393 - 402
  • [10] Application of finite element analysis in implant dentistry: A review of the literature
    Geng, JP
    Tan, KBC
    Liu, GR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (06) : 585 - 598