Stress analysis of a polyethylene acetabular component in the extreme flexion position-a finite element analysis

被引:0
作者
Promsang, Trai [1 ]
Wilairatana, Vajara [1 ]
Tangpornprasert, Pairat [2 ]
Virulsri, Chanyaphan [2 ]
Itiravivong, Pibul [1 ]
机构
[1] Chulalongkorn Univ, Fac Med, Dept Orthoped, Bangkok 10330, Thailand
[2] Chulalongkorn Univ, Dept Mech Engn, Fac Engn, Bangkok 10330, Thailand
关键词
Finite element analysis; impingement; polyethylene; stress; total hip replacement; TOTAL HIP-ARTHROPLASTY; IMPINGEMENT; MOTION; RANGE; DISLOCATION; CRACKING;
D O I
10.5372/1905-7415.0706.253
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: Many factors cause impingement of the femoral neck and polyethylene liners. Impingements definitely increase the rate of polyethylene wear. This effect has been a major cause of revision hip surgery. Squatting and sitting cross-legged are specific sitting positions typically used by Asian people, including Thai. These types of positions may cause impingements and abnormal stress distributions, resulting in massive destruction of polyethylene liners. Objective: To analyze the effects of squatting and sitting in a cross-legged position in total hip replacement patients using finite element analysis. Methods: A three-dimensional finite element model was developed to study the effects of squatting and sitting in a cross-legged position. The study was divided into two parts. First, mean hip ranges of motion of sitting positions were analyzed. Second, hip ranges of motion at one standard deviation were also analyzed. Locations of the impingement area and stress distribution were demonstrated using finite element software. Results: Squatting and sitting cross-legged at one standard deviation obviously cause significant impingement at the superior part of the polyethylene liner. Maximum principal stresses are 103 MPa and 24.5 MPa in squatting and sitting cross-legged, respectively. There is no impingement when the mean hip ranges of motion are used. The distance between the neck and cup are 4.05 mm and 4.15 mm in squatting and sitting cross-legged, respectively. Conclusion: Squatting and sitting cross-legged can cause significant impingement in commonly used total hip replacements. Massive destruction of the prostheses can develop by the process of the impingement.
引用
收藏
页码:881 / 884
页数:4
相关论文
共 50 条
  • [31] Finite element analysis and the stress correspondence paradigm
    Prathap, G
    SADHANA-ACADEMY PROCEEDINGS IN ENGINEERING SCIENCES, 1996, 21 : 525 - 546
  • [32] Analysis and results of dissociation of the polyethylene liner in the Harris-Galante I acetabular component
    Saito, Shu
    Ryu, Junnosuke
    Seki, Masayuki
    Ishii, Takao
    Saigo, Kaichiro
    Mori, Sei
    JOURNAL OF ARTHROPLASTY, 2008, 23 (04) : 522 - 526
  • [33] Polyethylene sterilized without irradiation in a polished uncemented acetabular component - A clinical and radiographic analysis
    Crockarell, John R., Jr.
    JOURNAL OF ARTHROPLASTY, 2008, 23 (06) : 911 - 915
  • [34] Finite element analysis of infra-acetabular screw fixation for the treatment of acetabular posterior column fracture
    Wang, Tingguang
    Zhao, Bei
    Yan, Jun
    Shao, Bin
    Mu, Weidong
    INTERNATIONAL ORTHOPAEDICS, 2022, 46 (03) : 623 - 634
  • [35] Finite element analysis of infra-acetabular screw fixation for the treatment of acetabular posterior column fracture
    Tingguang Wang
    Bei Zhao
    Jun Yan
    Bin Shao
    Weidong Mu
    International Orthopaedics, 2022, 46 : 623 - 634
  • [36] Finite Element Analysis to Probe the Influence of Acetabular Shell Design, Liner Material, and Subject Parameters on Biomechanical Response in Periprosthetic Bone
    Chatterjee, Subhomoy
    Kobylinski, Sabine
    Basu, Bikramjit
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2018, 140 (10):
  • [37] Finite element analysis of acetabular fractures-development and validation with a synthetic pelvis
    Shim, Vickie
    Boehme, Jierg
    Vaitl, Peter
    Klima, Stefan
    Josten, Christoph
    Anderson, Iain
    JOURNAL OF BIOMECHANICS, 2010, 43 (08) : 1635 - 1639
  • [38] Stress/strain and curvature analysis of laser-scribed polyethylene terephthalate films with multiple grooves using finite element method
    Heo, Do Won
    Kang, Min Gi
    Lee, Ho
    Kim, Sung Yeol
    ADVANCES IN MECHANICAL ENGINEERING, 2018, 10 (08):
  • [39] The finite element method in the analysis of the stress and strain distribution in polyethylene elements of hip and knee joints endoprostheses
    Nabrdalik, Marcin
    Sobocinski, Michal
    XXIII POLISH-SLOVAK SCIENTIFIC CONFERENCE ON MACHINE MODELLING AND SIMULATIONS (MMS 2018), 2019, 254
  • [40] The Effect of Malrotation of Tibial Component of Total Knee Arthroplasty on Tibial Insert during High Flexion Using a Finite Element Analysis
    Osano, Kei
    Nagamine, Ryuji
    Todo, Mitsugu
    Kawasaki, Makoto
    SCIENTIFIC WORLD JOURNAL, 2014,