Spectroscopic analysis of x-ray bursts from nichrome and conichrome X-pinch plasmas

被引:8
作者
Chandler, KM
Shlyaptseva, AS
Ouart, ND
Hansen, SB
Mitchell, MD
Pikuz, SA
Shelkovenko, TA
Hammer, DA
Kantsyrev, VL
Fedin, DA
机构
[1] Cornell Univ, Plasma Studies Lab, Ithaca, NY 14853 USA
[2] Univ Nevada, Dept Phys, Reno, NV 89557 USA
[3] Lawrence Livermore Natl Lab, Livermore, CA 94551 USA
[4] PN Lebedev Phys Inst, Moscow 119991, Russia
关键词
D O I
10.1063/1.1785272
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Radiative properties of plasmas from X pinches with nichrome and conichrome wires have been studied using the presented diagnostic arrangement. The experimental results have been produced at the Cornell XP facility with a peak current of 450 kA and a full width at half maximum pulse duration of 100 ns. The spatially resolved, time-integrated x-ray line spectra from the region of the X-pinch cross point have been recorded using different crystal spectrometers. In particular, K-shell and L-shell x-ray spectra of Ni and Cr have been recorded through different filters in the same pulses for X pinches from two different wire alloys. A nonlocal thermodynamic equilibrium collisional-radiative atomic kinetic model of Ni has been developed to identify the useful diagnostic spectroscopic features and to model experimental spectra. The results of the modeling and radiative properties of different materials from nichrome and conichrome alloys are compared and discussed. (C) 2004 American Institute of Physics.
引用
收藏
页码:3702 / 3704
页数:3
相关论文
共 8 条
[1]   Indirect X-ray line-formation processes in iron L-shell ions [J].
Gu, MF .
ASTROPHYSICAL JOURNAL, 2003, 582 (02) :1241-1250
[2]   Applications of advanced theoretical x-ray L-shell spectroscopy to various plasma and collision experiments [J].
Hansen, SB ;
Shlyaptseva, AS ;
Kantsyrev, VL ;
Fedin, DA ;
Ouart, ND ;
Safronova, UI ;
Fournier, KB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03) :1943-1946
[3]  
HANSEN SB, 2003, THESIS U NEVADA RENO
[4]  
Regemorter H V, 1962, ASTROPHYS J, V136, P906
[5]   Point-projection x-ray radiography using an X pinch as the radiation source [J].
Shelkovenko, TA ;
Sinars, DB ;
Pikuz, SA ;
Chandler, KM ;
Hammer, DA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01) :667-670
[6]   Radiographic and spectroscopic studies of X-pinch plasma implosion dynamics and x-ray burst emission characteristics [J].
Shelkovenko, TA ;
Sinars, DB ;
Pikuz, SA ;
Hammer, DA .
PHYSICS OF PLASMAS, 2001, 8 (04) :1305-1318
[7]   Advanced spectroscopic analysis of 0.8-1.0-MA Mo x pinches and the influence of plasma electron beams on L-shell spectra of Mo ions -: art. no. 026409 [J].
Shlyaptseva, AS ;
Hansen, SB ;
Kantsyrev, VL ;
Fedin, DA ;
Ouart, N ;
Fournier, KB ;
Safronova, UI .
PHYSICAL REVIEW E, 2003, 67 (02) :9
[8]   Time-resolved spectroscopy of Al, Ti, and Mo X pinch radiation using an X-ray streak camera [J].
Sinars, DB ;
Pikuz, SA ;
Shelkovenko, TA ;
Chandler, KM ;
Hammer, DA ;
Apruzese, JP .
JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2003, 78 (01) :61-83