Analytical validation of serum bone alkaline phosphatase (BAP OSTASE) on Liaison

被引:14
作者
Cavalier, Etienne [1 ]
Rozet, Eric [2 ]
Carlisi, Agnes [1 ]
Bekaert, Anne-Catherine [1 ]
Rousselle, Olivier [1 ]
Hubert, Philippe [2 ]
Chapelle, Jean-Paul [1 ]
Delanaye, Pierre [3 ]
机构
[1] Univ Liege, Univ Hosp Liege, Dept Clin Chem, B-4000 Liege, Belgium
[2] Univ Liege, Dept Analyt Chem, B-4000 Liege, Belgium
[3] Univ Liege, Univ Hosp Liege, Dept Nephrol & Dialysis, B-4000 Liege, Belgium
关键词
analytical validation; bone alkaline phosphatase; measurement uncertainty; reference range;
D O I
10.1515/CCLM.2010.019
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The goal of this study was to validate the DiaSorin Liaison BAP OSTASE, a new method for measurement of bone alkaline phosphatase (BAP), and to compare this method with the Beckman-Coulter Access Ostase. We also wanted to establish the reference range for BAP in adults and children. Methods: We determined the precision, functional sensitivity, recovery, linearity and measurement uncertainty, accuracy profile and beta-expectation limits. We defined an adult reference interval using individuals with 25-OH vitamin D >80 nmol/L, parathormone <58 ng/L, and normal calcium, phosphorous and estimated glomerular filtration rate. Each adult subclass (men/non-menopausal women/menopause women) contained 120 individuals. We also determined the 2.5th and 97.5th percentiles from a population of 450 children, stratified according to age and gender. Results: The results of the validation showed: precision <6%, functional sensitivity <0.74 mu g/L, mean recovery 98.8 +/- 4.2% and good linearity. Relative uncertainty ranged from 9.0% to 12.9%, and the risk of one result falling out of the +/- 15% acceptance limits was <5% for concentrations between 7 and 94 mu g/L. The Bland-Altman plot showed no systematic bias between the two methods. In adults, we did not find any statistical difference between the different subclasses. The upper limit of normality observed in the entire population (n=360) was 21.3 mu g/L (90% CI: 18.3-24.2 mu g/L). Conclusions: The Liaison BAP OSTASE is a robust method, and is completely validated between 7 and 93 mu g/L: in this range, 95% of the values obtained will be within +/- 15% of the true value. Clin Chem Lab Med 2010; 48: 67-72.
引用
收藏
页码:67 / 72
页数:6
相关论文
共 14 条
[1]  
[Anonymous], QUANT UNC AN MEAS
[2]  
[Anonymous], 1994, Accuracy trueness and precision of measurement methods and results
[3]  
[Anonymous], 2004, EP05A2 CLSI
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
CAVALIER E, 2009, J BONE MINER RES
[6]  
*CLIN LAB STAND I, 1986, EV LIN QUANT AN METH
[7]  
*CLIN LAB STAND I, 2003, EP6A CLSI, V23
[8]  
GLOVER S, 2008, J BONE MINER RES
[9]  
*ISO, 2003, 21748 ISODTS
[10]   A single measurement of biochemical markers of bone turnover has limited utility in the individual person [J].
Jensen, JEB ;
Kollerup, G ;
Sorensen, HA ;
Nielsen, SP ;
Sorensen, OH .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1997, 57 (04) :351-359