Investigation of conformal and intensity-modulated radiation therapy techniques to determine the absorbed fetal dose in pregnant patients with breast cancer

被引:9
作者
Ogretici, Akin [1 ]
Akbas, Ugur [1 ]
Koksal, Canan [1 ]
Bilge, Hatice [1 ]
机构
[1] Istanbul Univ, Inst Med Sci, Div Basic Oncol, Istanbul, Turkey
关键词
Breast cancer; Pregnancy; Fetal dose; 3-D CRT; IMRT; TLD; RADIOTHERAPY;
D O I
10.1016/j.meddos.2015.10.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The aim of this research was to investigate the fetal doses of pregnant patients undergoing conformal radiotherapy or intensity-modulated radiation therapy (IMRT) for breast cancers. An Alderson Rando phantom was chosen to simulate a pregnant patient with breast cancer who is receiving radiation therapy. This phantom was irradiated using the Varian Clinac DBX 600 system (Varian Medical System, Palo Alto, CA) linear accelerator, according to the standard treatment plans of both three-dimensional conformal radiation therapy (3-D CRT) and IMRT techniques. Thermoluminescent dosimeters were used to measure the irradiated phantom's virtually designated uterus area. Thermoluminescent dosimeter measurements (in the phantom) revealed that the mean cumulative fetal dose for 3-D CRT is 1.39 cGy and for IMRT it is 8.48 cGy, for a pregnant breast cancer woman who received radiation treatment of 50 Gy. The fetal dose was confirmed to increase by 70% for 3-D CRT and 40% for IMRT, if it is closer to the irradiated field by 5 cm. The mean fetal dose from 3-D CRT is 1.39 cGy and IMRT is 8.48 cGy, consistent with theoretic calculations. The IMRT technique causes the fetal dose to be 5 times more than that of 3-D CRT. Theoretic knowledge concerning the increase in the peripheral doses as the measurements approached the beam was also practically proven. (C) 2016 American Association of Medical Dosimetrists.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 15 条
  • [1] [Anonymous], 2014, SEER CANC STAT REV 1
  • [2] Fetal dose evaluation during breast cancer radiotherapy
    Antypas, C
    Sandilos, P
    Kouvaris, J
    Balafouta, E
    Karinou, E
    Kollaros, N
    Vlahos, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04): : 995 - 999
  • [3] Calculation of organ doses from breast cancer radiotherapy: a Monte Carlo study
    Berris, T.
    Mazonakis, M.
    Stratakis, J.
    Tzedakis, A.
    Fasoulaki, A.
    Damilakis, J.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (01): : 133 - 146
  • [4] Bilge H., 2010, J TURK SOC OBSTET GY, V7, P37
  • [5] Normalized data for the estimation of fetal radiation dose from radiotherapy of the breast
    Bradley, B.
    Fleck, A.
    Osei, E. K.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (946) : 818 - 827
  • [6] Brent R L, 1983, Clin Obstet Gynecol, V26, P484, DOI 10.1097/00003081-198306000-00030
  • [7] Internal scatter, the unavoidable major component of the peripheral dose in photon-beam radiotherapy
    Chofor, Ndimofor
    Harder, Dietrich
    Willborn, Kay C.
    Poppe, Bjoern
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (06) : 1733 - 1743
  • [8] Accuracy of out-of-field dose calculations by a commercial treatment planning system
    Howell, Rebecca M.
    Scarboro, Sarah B.
    Kry, S. F.
    Yaldo, Derek Z.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) : 6999 - 7008
  • [9] Accuracy and sources of error of out-of field dose calculations by a commercial treatment planning system for intensity-modulated radiation therapy treatments
    Huang, Jessie Y.
    Followill, David S.
    Wang, Xin A.
    Kry, Stephen F.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 186 - 197
  • [10] Photon-beam radiotherapy in pregnant patients: Can the fetal dose be limited to 10 cGy or less?
    Kourinou, Kalliopi M.
    Mazonakis, Michalis
    Lyraraki, Efrossyni
    Damilakis, John
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (01): : 85 - 91