A clinical study of lung cancer dose calculation accuracy with Monte Carlo simulation

被引:28
作者
Zhao, Yanqun [1 ]
Qi, Guohai [1 ]
Yin, Gang [1 ]
Wang, Xianliang [1 ]
Wang, Pei [1 ]
Li, Jian [1 ]
Xiao, Mingyong [1 ]
Li, Jie [1 ]
Kang, Shengwei [1 ]
Liao, Xiongfei [1 ]
机构
[1] Sichuan Prov Canc Hosp, Dept Radiat Oncol, Chengdu 610041, Sichuan, Peoples R China
关键词
3-Dimensional conformal radiation therapy; Collapsed cone convolution; Pencil beam convolution; Lung cancer; Monte Carlo; Intensity-modulated radiation therapy; PENCIL-BEAM; DOSIMETRIC VERIFICATION; CALCULATION ALGORITHMS; RADIATION-THERAPY; COLLAPSED CONE; PHOTON BEAMS; IMRT; RADIOTHERAPY; ENERGY; INTERFACES;
D O I
10.1186/s13014-014-0287-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The accuracy of dose calculation is crucial to the quality of treatment planning and, consequently, to the dose delivered to patients undergoing radiation therapy. Current general calculation algorithms such as Pencil Beam Convolution (PBC) and Collapsed Cone Convolution (CCC) have shortcomings in regard to severe inhomogeneities, particularly in those regions where charged particle equilibrium does not hold. The aim of this study was to evaluate the accuracy of the PBC and CCC algorithms in lung cancer radiotherapy using Monte Carlo (MC) technology. Methods and materials: Four treatment plans were designed using Oncentra Masterplan TPS for each patient. Two intensity-modulated radiation therapy (IMRT) plans were developed using the PBC and CCC algorithms, and two three-dimensional conformal therapy (3DCRT) plans were developed using the PBC and CCC algorithms. The DICOM-RT files of the treatment plans were exported to the Monte Carlo system to recalculate. The dose distributions of GTV, PTV and ipsilateral lung calculated by the TPS and MC were compared. Result: For 3DCRT and IMRT plans, the mean dose differences for GTV between the CCC and MC increased with decreasing of the GTV volume. For IMRT, the mean dose differences were found to be higher than that of 3DCRT. The CCC algorithm overestimated the GTV mean dose by approximately 3% for IMRT. For 3DCRT plans, when the volume of the GTV was greater than 100 cm(3), the mean doses calculated by CCC and MC almost have no difference. PBC shows large deviations from the MC algorithm. For the dose to the ipsilateral lung, the CCC algorithm overestimated the dose to the entire lung, and the PBC algorithm overestimated V-20 but underestimated V-5; the difference in V-10 was not statistically significant. Conclusions: PBC substantially overestimates the dose to the tumour, but the CCC is similar to the MC simulation. It is recommended that the treatment plans for lung cancer be developed using an advanced dose calculation algorithm other than PBC. MC can accurately calculate the dose distribution in lung cancer and can provide a notably effective tool for benchmarking the performance of other dose calculation algorithms within patients.
引用
收藏
页数:9
相关论文
共 35 条
[1]   The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverage [J].
Aarup, Lasse Rye ;
Nahum, Alan E. ;
Zacharatou, Christina ;
Juhler-Nottrup, Trine ;
Knoos, Tommy ;
Nystrom, Hakan ;
Specht, Lena ;
Wieslander, Elinore ;
Korreman, Stine S. .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (03) :405-414
[2]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[3]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[4]  
[Anonymous], 2010, ICRU Report 83
[5]   Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium [J].
Carrasco, P ;
Jornet, N ;
Duch, MA ;
Weber, L ;
Ginjaume, M ;
Eudaldo, T ;
Jurado, D ;
Ruiz, A ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (10) :2899-2911
[6]   STEREOTACTIC, SINGLE-DOSE IRRADIATION OF LUNG TUMORS: A COMPARISON OF ABSOLUTE DOSE AND DOSE DISTRIBUTION BETWEEN PENCIL BEAM AND MONTE CARLO ALGORITHMS BASED ON ACTUAL PATIENT CT SCANS [J].
Chen, Huixiao ;
Lohr, Frank ;
Fritz, Peter ;
Wenz, Frederik ;
Dobler, Barbara ;
Lorenz, Friedlieb ;
Muehlnickel, Werner .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03) :955-963
[7]   Lethal pneumonitis in a phase I study of chemotherapy and IMRT for NSCLC: The need to investigate the accuracy of dose computation [J].
De Neve, W ;
De Wagter, C .
RADIOTHERAPY AND ONCOLOGY, 2005, 75 (02) :246-247
[8]   Optimization of extracranial stereotactic radiation therapy of small lung lesions using accurate dose calculation algorithms [J].
Dobler, Barbara ;
Walter, Cornelia ;
Knopf, Antje ;
Fabri, Daniella ;
Loeschel, Rainer ;
Polednik, Martin ;
Schneider, Frank ;
Wenz, Frederik ;
Lohr, Frank .
RADIATION ONCOLOGY, 2006, 1 (1)
[9]   Influence of calculation model on dose distribution in stereotactic radiotherapy for pulmonary targets [J].
Haedinger, U ;
Krieger, T ;
Flentje, M ;
Wulf, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (01) :239-249
[10]   Influence of calculation algorithm on dose distribution in irradiation of non-small cell lung cancer (NSCLC) - Collapsed cone versus pencil beam [J].
Koelbl, O ;
Krieger, K ;
Haedinger, U ;
Sauer, O ;
Flentje, M .
STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (12) :783-788