Dosimetric verification of IMRT and 3D conformal treatment delivery using EPID

被引:1
作者
Nasseri, S. [1 ]
Bahreyni, M. H. [1 ]
Momennezhad, M. [1 ,2 ]
Gholamhosseinian, H. [1 ]
Shahedi, F. [1 ]
Hashemi, S. M. [1 ]
Mohammadi, M. [3 ,4 ]
机构
[1] Mashhad Univ Med Sci, Med Phys Res Ctr, Mashhad, Razavi Khorasan, Iran
[2] Mashhad Univ Med Sci, Nucl Med Res Ctr, Mashhad, Razavi Khorasan, Iran
[3] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA 5000, Australia
[4] Univ Adelaide, Sch Phys Sci, Adelaide, SA 5005, Australia
关键词
Electronic portal imaging devices (EPIDS); Back-projection; Water equivalent path length; Digital reconstructed radiograph (DRR); ALGORITHM; EXIT;
D O I
10.1016/j.apradiso.2022.110116
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: Electronic portal imaging devices (EPIDs) could potentially be useful for either in-vivo or pre-treatment dosimetric verification of external beam radiation therapy. The accuracy of EPID for dosimetric purposes is highly dependent on the specific method used for the determination of dose-response characteristics. The aim of this study was to develop a simple and time-saving EPID back-projection dosimetry algorithm for 2D dose verification in 3D conformal and intensity-modulated beams.Methods: The procedure of dose reconstruction includes a first calibration step using ionization chamber measurements to convert the Electronic Portal Image (EPI) pixel values into an absorbed dose in water. Subsequently, several corrections were applied to the Portal Dose Images (PDIs) for the effect of field size, attenuator thickness, scattering radiation, beam hardening and EPID off-axis response. Furthermore, to consider tissue inhomogeneity for accurate dose reconstruction, the patient's water equivalent path length (WEPL) was calculated using a range of digitally reconstructed radiographs (DRRs) obtained at various thicknesses by Plastimatch software. The EPIDderived dose maps accuracy was assessed by comparing with the treatment planning system (TPS) calculated dose in the prostate region of Alderson phantom irradiated with 3D conformal and intensity-modulated beams.Results: The gamma analysis for the dose plane showed agreements of 96.95% and 93.5% for 3D conformal and IMRT fields, respectively, with 3%/3 mm acceptance criteria.Conclusion: The presented algorithm can provide accurate absolute 2D dose maps for clinical use in the context of 3DCRT or IMRT Quality Assurance (QA) programs.
引用
收藏
页数:7
相关论文
共 28 条
  • [1] Alhazmi A., 2018, EPID-based 3D dosimetry for pre-treatment IMRT/VMAT Quality Assurance
  • [2] Comparison of forward- and back-projection in vivo EPID dosimetry for VMAT treatment of the prostate
    Bedford, James L.
    Hanson, Ian M.
    Hansen, Vibeke N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02)
  • [3] Portal dosimetry for VMAT using integrated images obtained during treatment
    Bedford, James L.
    Hanson, Ian M.
    Hansen, Vibeke Nordmark
    [J]. MEDICAL PHYSICS, 2014, 41 (02)
  • [4] Implementation of EPID transit dosimetry based on a through-air dosimetry algorithm
    Berry, Sean L.
    Sheu, Ren-Dih
    Polvorosa, Cynthia S.
    Wuu, Cheng-Shie
    [J]. MEDICAL PHYSICS, 2012, 39 (01) : 87 - 98
  • [5] New method to obtain the midplane dose using portal in vivo dosimetry
    Boellaard, R
    Essers, M
    van Herk, M
    Mijnheer, BJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (02): : 465 - 474
  • [6] Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations
    Bojechko, C.
    Ford, E. C.
    [J]. MEDICAL PHYSICS, 2015, 42 (12) : 6912 - 6918
  • [7] Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy
    Ding, Aiping
    Xing, Lei
    Han, Bin
    [J]. MEDICAL PHYSICS, 2015, 42 (07) : 4190 - 4198
  • [8] Analysis of the dose calculation accuracy for IMRT in lung: A 2D approach
    Dvorak, Pavel
    Stock, Markus
    Kroupa, Bernhard
    Bogner, Joachim
    Georg, Dietmar
    [J]. ACTA ONCOLOGICA, 2007, 46 (07) : 928 - 936
  • [9] IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119
    Ezzell, Gary A.
    Burmeister, Jay W.
    Dogan, Nesrin
    LoSasso, Thomas J.
    Mechalakos, James G.
    Mihailidis, Dimitris
    Molineu, Andrea
    Palta, Jatinder R.
    Ramsey, Chester R.
    Salter, Bill J.
    Shi, Jie
    Xia, Ping
    Yue, Ning J.
    Xiao, Ying
    [J]. MEDICAL PHYSICS, 2009, 36 (11) : 5359 - 5373
  • [10] Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits
    Fuangrod, Todsaporn
    Greer, Peter B.
    Woodruff, Henry C.
    Simpson, John
    Bhatia, Shashank
    Zwan, Benjamin
    vanBeek, Timothy A.
    McCurdy, Boyd M. C.
    Middleton, Richard H.
    [J]. RADIATION ONCOLOGY, 2016, 11