Diffusivity Measurement by Single-Molecule Recycling in a Capillary Microchannel

被引:0
|
作者
Wang, Bo [1 ,2 ,3 ]
Davis, Lloyd M. [1 ,2 ]
机构
[1] Univ Tennessee, Ctr Laser Applicat, Inst Space, 411 B H Goethert Pkwy, Tullahoma, TN 37388 USA
[2] Univ Tennessee, Dept Phys & Astron, 1408 Cirde Dr, Knoxville, TN 37996 USA
[3] Quenda Terahertz Technol Ltd, 600 Jiushui E Rd, Qingdao 266100, Peoples R China
基金
美国国家科学基金会;
关键词
capillary; microfluidic device; single-molecule recycling; maximum likelihood; FLUORESCENCE CORRELATION SPECTROSCOPY; 3-DIMENSIONAL TRACKING; BROWNIAN-MOTION; NANOPARTICLE;
D O I
10.3390/mi12070800
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microfluidic devices have been extensively investigated in recent years in fields including ligand-binding analysis, chromatographic separation, molecular dynamics, and DNA sequencing. To prolong the observation of a single molecule in aqueous buffer, the solution in a sub-micron scale channel is driven by a electric field and reversed after a fixed delay following each passage, so that the molecule passes back and forth through the laser focus and the time before irreversible photobleaching is extended. However, this practice requires complex chemical treatment to the inner surface of the channel to prevent unexpected sticking to the surface and the confined space renders features, such as a higher viscosity and lower dielectric constant, which slow the Brownian motion of the molecule compared to the bulk solution. Additionally, electron beam lithography used for the fabrication of the nanochannel substantially increases the cost, and the sub-micron dimensions make the molecule difficult to locate. In this paper, we propose a method of single-molecule recycling in a capillary microchannel. A commercial fused-silica capillary with an inner diameter of 2 microns is chopped into a 1-inch piece and is fixed onto a cover slip. Two o-rings on the sides used as reservoirs and an o-ring in the middle used as observation window are glued over the capillary. The inner surface of the capillary is chemically processed to reduce the non-specific sticking and to improve capillary effect. The device does not require high-precision fabrication and thus is less costly and easier to prepare than the nanochannel. 40 nm Fluospheres(R) in 50% methanol are used as working solution. The capillary is translated by a piezo stage to recycle the molecule, which diffuses freely through the capillary, and a confocal microscope is used for fluorescence collection. The passing times of the molecule through the laser focus are calculated by a real-time control system based on an FPGA, and the commands of translation are given to the piezo stage through a feedback algorithm. The larger dimensions of the capillary overcomes the strong sticking, the reduced diffusivity, and the difficulty of localizing the molecule. We have achieved a maximum number of recycles of more than 200 and developed a maximum-likelihood estimation of the diffusivity of the molecule, which attains results of the same magnitude as the previous report. This technique simplifies the overall procedure of the single-molecule recycling and could be useful for the ligand-binding studies in high-throughput screening.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Improved timing and diffusivity measurement in single-molecule recycling in a nanochannel
    Wang, Bo
    Davis, Lloyd M.
    SINGLE MOLECULE SPECTROSCOPY AND SUPERRESOLUTION IMAGING X, 2017, 10071
  • [2] Measurement of single-molecule conductance
    Chen, Fang
    Hihath, Joshua
    Huang, Zhifeng
    Li, Xiulan
    Tao, N. J.
    ANNUAL REVIEW OF PHYSICAL CHEMISTRY, 2007, 58 (58) : 535 - 564
  • [3] Resolving Mixtures in Solution by Single-Molecule Rotational Diffusivity
    Yang, Hsiang-Yu
    Moerner, W. E.
    NANO LETTERS, 2018, 18 (08) : 5279 - 5287
  • [4] Single-molecule trapping and measurement in solution
    Bespalova, Maria I.
    Mahanta, Sushanta
    Krishnan, Madhavi
    CURRENT OPINION IN CHEMICAL BIOLOGY, 2019, 51 : 113 - 121
  • [5] Reproducible measurement of single-molecule conductivity
    Cui, XD
    Primak, A
    Zarate, X
    Tomfohr, J
    Sankey, OF
    Moore, AL
    Moore, TA
    Gust, D
    Harris, G
    Lindsay, SM
    SCIENCE, 2001, 294 (5542) : 571 - 574
  • [6] Torque Measurement at the Single-Molecule Level
    Forth, Scott
    Sheinin, Maxim Y.
    Inman, James
    Wang, Michelle D.
    ANNUAL REVIEW OF BIOPHYSICS, VOL 42, 2013, 42 : 583 - 604
  • [7] Single-molecule measurement in living cells
    Ichinose, J
    Sako, Y
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2004, 23 (08) : 587 - 594
  • [8] Rapid measurement of single-molecule conductance
    Hihath, Joshua
    Tao, Nongjian
    NANOTECHNOLOGY, 2008, 19 (26)
  • [9] Single-molecule spectroscopy for molecular identification in capillary electrophoresis
    Lee, JY
    Li, HW
    Yeung, ES
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1053 (1-2) : 173 - 179
  • [10] Precise Measurement of Single -Molecule Rotational Diffusivity in Solution
    Yang, Hsiang-Yu
    Moerner, W. E.
    BIOPHYSICAL JOURNAL, 2018, 114 (03) : 170A - 170A