Failure and Malfunction of da Vinci Surgical Systems During Various Robotic Surgeries: Experience From Six Departments at a Single Institute

被引:42
作者
Kim, Won Tae
Ham, Won Sik
Jeong, Wooju
Song, Hyun Jung
Rha, Koon Ho
Choi, Young Deuk [1 ]
机构
[1] Yonsei Univ, Dept Urol, Coll Med, Seoul 120752, South Korea
关键词
PROSTATECTOMY;
D O I
10.1016/j.urology.2009.05.071
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES To analyze the mechanical failures and malfunctions of the da Vinci Surgical (S) System during various robotic surgeries in 6 different departments at our institute and also evaluated the solutions for the failures and malfunctions. METHODS From July 2005 to December 2008, a total of 1797 robotic surgeries were performed at our institute. The surgeries were performed using 4 da Vinci surgical systems (1 standard da Vinci system from July 2005 to July 2007 and 3 da Vinci S systems from July 2007 to December 2008). Mechanical failures or malfunctions occurred in 43 cases. We evaluated the robotic surgeries according to the type of surgery and the department. We analyzed the cases involving conversion to open or laparoscopic surgeries and those in which there was a malfunction with the instrument. RESULTS There were 43 cases (2.4%) of mechanical failure with the da Vinci system from a total of 1797 robotic surgeries. This included 24 (1.3%) cases of mechanical failure or malfunction and 19 cases (1.1%) of instrument malfunction. The mechanical malfunction included 1 on/off failure, 5 console malfunctions, 6 robotic arm malfunctions, 2 optic system malfunctions, and 10 system errors. One open and 2 laparoscopic conversions (3 cases; 0.17%) were performed. CONCLUSIONS Mechanical failure or malfunction occurred during robotic surgery in 43 cases (2.4%), and the open or laparoscopic conversion rate during surgery was very low (0.17%). We found the mechanical failure or malfunction to be rare. UROLOGY 74: 1234-1237, 2009. (C) 2009 Elsevier
引用
收藏
页码:1234 / 1237
页数:4
相关论文
共 11 条
[1]   Robotic technology and the translation of open radical prostatectomy to laparoscopy:: The early Frankfurt experience with robotic radical prostatectomy and one year follow-up [J].
Bentas, W ;
Wolfram, M ;
Jones, J ;
Bräutigam, R ;
Kramer, W ;
Binder, J .
EUROPEAN UROLOGY, 2003, 44 (02) :175-181
[2]  
Borden Lester S Jr, 2007, Can J Urol, V14, P3499
[3]   Computer-assisted cardiac surgery [J].
Carpentier, A ;
Loulmet, D ;
Aupecle, B ;
Berrebi, A ;
Relland, J .
LANCET, 1999, 353 (9150) :379-380
[4]  
Eichel L., 2005, AUA UPDATE SERIES, V24, P106
[5]   Malfunction of da Vinci Robotic System-Disassembled Surgeon's Console Hand Piece: Case Report and Review of the Literature [J].
Ham, Won Sik ;
Park, Sung Yul ;
Yu, Ho Song ;
Choi, Young Deuk ;
Hong, Sung Joon ;
Rha, Koon Ho .
UROLOGY, 2009, 73 (01) :209.e7-209.e8
[6]   Mechanical failure rate of davinci robotic system: Implications for pre-op patient counseling [J].
Kozlowski, PM ;
Porter, CR ;
Corman, JM .
JOURNAL OF UROLOGY, 2006, 175 (04) :372-373
[7]   Robotic surgery - A current perspective [J].
Lanfranco, AR ;
Castellanos, AE ;
Desai, JP ;
Meyers, WC .
ANNALS OF SURGERY, 2004, 239 (01) :14-21
[8]   Robotic Equipment Malfunction During Robotic Prostatectomy: A Multi-institutional Study [J].
Lavery, Hugh J. ;
Thaly, Rahul ;
Albala, David ;
Ahlering, Thomas ;
Shalhav, Arieh ;
Lee, David ;
Fagin, Randy ;
Wiklund, Peter ;
Dasgupta, Prokar ;
Costello, Anthony J. ;
Tewari, Ashutosh ;
Coughlin, Geoff ;
Patel, Vipul R. .
JOURNAL OF ENDOUROLOGY, 2008, 22 (09) :2165-2168
[9]   Intraoperative breakage of needle driver jaw during robotic-assisted laparoscopic radical prostatectomy [J].
Park, Sung Yul ;
Cho, Kang Su ;
Lee, Seung Wook ;
Soh, Byung Hyun ;
Rha, Koon Ho .
UROLOGY, 2008, 71 (01) :168.e5-168.e6
[10]   Surgical robotics: The early chronicles - A personal historical perspective [J].
Satava, RM .
SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2002, 12 (01) :6-16