Characterization of imaging latency for real-time MRI-guided radiotherapy

被引:41
作者
Borman, P. T. S. [1 ,2 ]
Tijssen, R. H. N. [1 ]
Bos, C. [2 ]
Moonen, C. T. W. [2 ]
Raaymakers, B. W. [1 ]
Glitzner, M. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Imaging Div, Heidelberglaan 100, Utrecht, Netherlands
关键词
MR-linac; real-time MRI; imaging latency; radial MRI; MR-guided tracking; readout ordering; TARGET TRACKING; DYNAMIC MRI; COMPENSATION; MOTION; RECONSTRUCTION; MLC; PERFORMANCE; PREDICTION; ORGANS;
D O I
10.1088/1361-6560/aad2b7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hybrid MR-linac systems can use fast dynamic MR sequences for tumor tracking and adapt the radiation treatment in real-time. For this the imaging latency must be as short as possible. This work describes how different acquisition parameters influence this latency. First, the latency was measured for Cartesian readouts with phase encode orderings linear, reverse-linear, and high-low. Second, the latency was measured for radial readouts with linear and golden angle profile orderings. To reduce the latency, a spatio-temporal (k-t) filter that suppresses the k-space center of earlier acquired spokes was implemented for the golden angle sequence. For Cartesian readouts a high-low ordering achieved a three times lower latency compared to a linear ordering with our sampling parameters. For radial readouts the filter was able to reduce the acquisition latency from half the acquisition time to a quarter of the acquisition time. The filter did not compromise the signal-to-noise ratio and the artifact power.
引用
收藏
页数:9
相关论文
共 50 条
[31]   Evaluation of real-time tumor contour prediction using LSTM networks for MR-guided radiotherapy [J].
Lombardo, Elia ;
Rabe, Moritz ;
Xiong, Yuqing ;
Nierer, Lukas ;
Cusumano, Davide ;
Placidi, Lorenzo ;
Boldrini, Luca ;
Corradini, Stefanie ;
Niyazi, Maximilian ;
Reiner, Michael ;
Belka, Claus ;
Kurz, Christopher ;
Riboldi, Marco ;
Landry, Guillaume .
RADIOTHERAPY AND ONCOLOGY, 2023, 182
[32]   Hybrid MRI-Ultrasound acquisitions, and scannerless real-time imaging [J].
Preiswerk, Frank ;
Toews, Matthew ;
Cheng, Cheng-Chieh ;
Chiou, Jr-yuan George ;
Mei, Chang-Sheng ;
Schaefer, Lena F. ;
Hoge, W. Scott ;
Schwartz, Benjamin M. ;
Panych, Lawrence P. ;
Madore, Bruno .
MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (03) :897-908
[33]   Feasibility of through-time spiral generalized autocalibrating partial parallel acquisition for low latency accelerated real-time MRI of speech [J].
Lingala, Sajan Goud ;
Zhu, Yinghua ;
Lim, Yongwan ;
Toutios, Asterios ;
Ji, Yunhua ;
Lo, Wei-Ching ;
Seiberlich, Nicole ;
Narayanan, Shrikanth ;
Nayak, Krishna S. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (06) :2275-2282
[34]   A fast volumetric 4D-MRI with sub-second frame rate for abdominal motion monitoring and characterization in MRI-guided radiotherapy [J].
Yuan, Jing ;
Wong, Oi Lei ;
Zhou, Yiliang ;
Chueng, Kin Yin ;
Yu, Siu Ki .
QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (07) :1303-1314
[35]   MRI-guided radiotherapy for focal boosting of T3b prostate cancer [J].
van der Heide, Uulke A. ;
Dassen, Matthijs G. ;
Betgen, Anja ;
de Ruiter, Peter ;
van der Linden, Joeke ;
Wiersema, Lisa ;
Neijndorff, Ben ;
Pos, Floris J. ;
Janssen, Tomas ;
Abbenhuis, Leontien ;
van Kollenburg, Peter ;
Reijnen, Casper ;
Smeenk, Robert Jan ;
Brunenberg, Ellen .
RADIOTHERAPY AND ONCOLOGY, 2024, 197 :S286-S288
[36]   Adaptive auto-segmentation for MRI-guided online adaptive radiotherapy of cervical carcinoma [J].
Ding, Shouliang .
RADIOTHERAPY AND ONCOLOGY, 2024, 194 :S3029-S3030
[37]   Computerized triplet beam orientation optimization for MRI-guided Co-60 radiotherapy [J].
Dan Nguyen ;
Thomas, David ;
Cao, Minsong ;
O'Connor, Daniel ;
Lamb, James ;
Sheng, Ke .
MEDICAL PHYSICS, 2016, 43 (10) :5667-5675
[38]   Inter-fractional portability of deep learning models for lung target tracking on cine imaging acquired in MRI-guided radiotherapy [J].
Peng, Jiayuan ;
Stowe, Hayley B. ;
Samson, Pamela P. ;
Robinson, Clifford G. ;
Yang, Cui ;
Hu, Weigang ;
Zhang, Zhen ;
Kim, Taeho ;
Hugo, Geoffrey D. ;
Mazur, Thomas R. ;
Cai, Bin .
PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (02) :769-777
[39]   Adaptive hypofractionted and stereotactic body radiotherapy for lung tumors with real-time MRI guidance [J].
Bryant, John M. ;
Sim, Austin J. ;
Feygelman, Vladimir ;
Latifi, Kujtim ;
Rosenberg, Stephen A. .
FRONTIERS IN ONCOLOGY, 2023, 13
[40]   Speech production real-time MRI at 0.55 T [J].
Lim, Yongwan ;
Kumar, Prakash ;
Nayak, Krishna S. .
MAGNETIC RESONANCE IN MEDICINE, 2024, 91 (01) :337-343