共 14 条
Single-Shot T1 Mapping Using Simultaneous Acquisitions of Spin- and Stimulated-Echo-Planar Imaging (2D ss-SESTEPI)
被引:7
|作者:
Shi, Xianfeng
[2
]
Kim, Seong-Eun
[1
]
Jeong, Eun-Kee
[1
]
机构:
[1] Univ Utah, Dept Radiol, Utah Ctr Adv Imaging Res, Salt Lake City, UT 84108 USA
[2] Univ Utah, Dept Phys, Utah Ctr Adv Imaging Res, Salt Lake City, UT 84108 USA
关键词:
EPI;
STEPI;
stimulated EPI;
SESTEPI;
spin/stimulated EPI;
real-time T-1 mapping;
dynamic T-1 mapping;
HUMAN-BRAIN;
IN-VIVO;
EPI;
RELAXATION;
CORD;
MRI;
D O I:
10.1002/mrm.22367
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The conventional stimulated-echo NMR sequence only measures the longitudinal component while discarding the transverse component, after tipping up the prepared magnetization. This transverse magnetization can be used to measure a spin echo, in addition to the stimulated echo. Two-dimensional single-shot spin- and stimulated-echo-planar imaging (ss-SESTEPI) is an echo-planar-imaging-based single-shot imaging technique that simultaneously acquires a spin-echo-planar image and a stimulated-echo-planar image after a single radio-frequency excitation. The magnitudes of the spin-echo-planar image and stimulated-echo-planar image differ by T-1 decay and diffusion weighting for perfect 90 degrees radiofrequency and thus can be used to rapidly measure T-1. However, the spatial variation of amplitude of radiofrequency field induces uneven splitting of the transverse magnetization for the spin-echo-planar image and stimulated-echo-planar image within the imaging field of view. Correction for amplitude of radiofrequency field inhomogeneity is therefore critical for two-dimensional ss-SESTEPI to be used for T-1 measurement. We developed a method for amplitude of radiofrequency field inhomogeneity correction by acquiring an additional stimulated-echo-planar image with minimal mixing time, calculating the difference between the spin echo and the stimulated echo and multiplying the stimulated-echo-planar image by the inverse functional map. Diffusion-induced decay is corrected by measuring the average diffusivity during the prescanning. Rapid single-shot T-1 mapping may be useful for various applications, such as dynamic T-1 mapping for real-time estimation of the concentration of contrast agent in dynamic contrast enhancement MRI. Magn Reson Med 64:734-742,2010. (C) 2010 Wiley-Liss, Inc.
引用
收藏
页码:734 / 742
页数:9
相关论文