Detectability of catheters on bedside chest radiographs:: Comparison between liquid crystal display and high-resolution cathode-ray tube monitors

被引:32
作者
Scharitzer, M
Prokop, M
Weber, M
Fuchsjäger, M
Oschatz, E
Schaefer-Prokop, C
机构
[1] Univ Vienna, Dept Radiol, A-1090 Vienna, Austria
[2] Univ Vienna, Ludwig Boltzmann Inst Clin & Expt Radiol Res, A-1090 Vienna, Austria
关键词
D O I
10.1148/radiol.2342031297
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare observer performance with a flat-panel liquid crystal display (LCD) monitor and with a high-resolution gray-scale cathode-ray tube (CRT) monitor in the detection of simulated support catheters on bedside chest radiographs. MATERIALS AND METHODS: The ethics committee did not require approval or patient informed consent when this study began. Because of a change in regulations, before images were acquired the nature of the study and procedures were explained to patients or their relatives, and consent was then obtained. A total of 131 catheter fragments (12-14 per radiograph) were superimposed over 10 anteroposterior bedside chest radiographs obtained with storage phosphor technology. Images were displayed on an LCD monitor (1536 x 2048 matrix) and a CRT monitor (2048 X 2560 matrix). Five radiologists independently located the catheter fragments and rated their confidence in detection with bright and subdued ambient light. A two-way analysis of variance and the Friedman test were used for statistical analysis. RESULTS: There was no significant difference for either display type with respect to correctly detected catheter fragments (mean sensitivity, 56.6% and 56.0% for the CRT and the LCD monitors, respectively, with bright light and 61.2% for both monitors with subdued light). With both display types, detection rate with bright light decreased significantly (P < .05). False-positive rates and confidence ratings were not significantly affected by monitor type or ambient light. CONCLUSION: In a study with simulation of clinical conditions, performance of the LCD monitor and high-resolution CRT monitor for detection of support catheters on bedside chest radiographs was equivalent. With both displays, detection performance was equally reduced with bright ambient light. (C) RSNA, 2004.
引用
收藏
页码:611 / 616
页数:6
相关论文
共 17 条
[11]   Accuracy of interpretation of CT scans: Comparing PACS monitor displays and hard-copy images [J].
Reiner, BI ;
Siegel, EL ;
Hooper, FJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (06) :1407-1410
[12]  
Roehrig H, 2002, RADIOLOGY, V225, P186
[13]  
Roehrig H, 2001, J DIGIT IMAGING, V14, P142, DOI 10.1053/jdim.2001.28993
[14]   EFFECT OF AMBIENT LIGHT ON ELECTRONICALLY DISPLAYED MEDICAL IMAGES AS MEASURED BY LUMINANCE-DISCRIMINATION THRESHOLDS [J].
ROGERS, DC ;
JOHNSTON, RE ;
PIZER, SM .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :976-983
[15]   DIGITAL STORAGE PHOSPHOR IMAGING VERSUS CONVENTIONAL FILM RADIOGRAPHY IN CT-DOCUMENTED CHEST DISEASE [J].
SCHAEFER, CM ;
GREENE, R ;
OESTMANN, JW ;
HALL, DA ;
LLEWELLYN, HJ ;
MCCARTHY, KA ;
PILESPELLMAN, ER ;
RUBENS, JR .
RADIOLOGY, 1990, 174 (01) :207-210
[16]  
Schaefer-Prokop C. M., 1997, EUR RADIOL S3, V7, P58
[17]  
Siegel EL, 2002, RADIOLOGY, V225, P501