Serum titanium level for diagnosis of a failed, metal-backed patellar component

被引:54
作者
Leopold, SS [1 ]
Berger, RA [1 ]
Patterson, L [1 ]
Skipor, AK [1 ]
Urban, RM [1 ]
Jacobs, JJ [1 ]
机构
[1] USA, Med Corps, William Beaumont Army Med Ctr, Orthopaed Surg Serv, El Paso, TX 79920 USA
关键词
titanium; metal ion level; total knee arthroplasty; revision; patella;
D O I
10.1054/arth.2000.6632
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A case is presented in which an elevated serum titanium level was used to make the diagnosis of a failed metal-backed patellar component. The preoperative serum titanium level was 536.8 ppb, which was 98 times higher than the patient's previous level (taken 1 year earlier, when he was asymptomatic) and 2 orders of magnitude higher than the expected level with a well-functioning implant of this type. Revision surgery confirmed that the polyethylene portion of the patellar component had worn through, leaving the titanium portion of the patellar implant to articulate with the femoral component. Wear-through was not evident on preoperative radiographs or clinical examination. As knowledge about the expected ranges for serum metal ion levels after total joint arthroplasty continues to increase, the diagnostic utility of serum metal ion testing in the evaluation of joint arthroplasty function will continue to improve.
引用
收藏
页码:938 / 943
页数:6
相关论文
共 34 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]   METALLOSIS ASSOCIATED WITH A STABLE TITANIUM-ALLOY FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT - A CASE-REPORT [J].
BLACK, J ;
SHERK, H ;
BONINI, J ;
ROSTOKER, WR ;
SCHAJOWICZ, F ;
GALANTE, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (01) :126-130
[3]   SERUM CONCENTRATIONS OF CHROMIUM, COBALT AND NICKEL AFTER TOTAL HIP-REPLACEMENT - A 6 MONTH STUDY [J].
BLACK, J ;
MAITIN, EC ;
GELMAN, H ;
MORRIS, DM .
BIOMATERIALS, 1983, 4 (03) :160-164
[4]   Increased levels of tumor necrosis factor-alpha and interleukin-6 protein and messenger RNA in human peripheral blood monocytes due to titanium particles [J].
Blaine, TA ;
Rosier, RN ;
Puzas, JE ;
Looney, RJ ;
Reynolds, PR ;
Reynolds, SD ;
OKeefe, RJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (08) :1181-1192
[5]   Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[6]   WIDESPREAD DISSEMINATION OF METAL DEBRIS FROM IMPLANTS [J].
CASE, CP ;
LANGKAMER, VG ;
JAMES, C ;
PALMER, MR ;
KEMP, AJ ;
HEAP, PF ;
SOLOMON, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :701-712
[7]   METAL SENSITIVITY BEFORE AND AFTER TOTAL HIP ARTHROPLASTY [J].
DEUTMAN, R ;
MULDER, TJ ;
BRIAN, R ;
NATER, JP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) :862-865
[8]  
DORR LD, 1990, CLIN ORTHOP RELAT R, P82
[9]   THE INCIDENCE OF CANCER FOLLOWING TOTAL HIP-REPLACEMENT [J].
GILLESPIE, WJ ;
FRAMPTON, CMA ;
HENDERSON, RJ ;
RYAN, PM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :539-542
[10]  
GILLESPIE WJ, 1996, CLIN ORTHOP S, V329, pS290