Patient-Specific Quality Assurance Using a 3D-Printed Chest Phantom for Intraoperative Radiotherapy in Breast Cancer

被引:5
作者
Choi, Yeonho [1 ]
Lee, Ik Jae [1 ]
Park, Kwangwoo [2 ]
Park, Kyung Ran [3 ]
Cho, Yeona [1 ]
Kim, Jun Won [1 ]
Lee, Ho [1 ]
机构
[1] Yonsei Univ, Coll Med, Gangnam Severance Hosp, Dept Radiat Oncol, Seoul, South Korea
[2] Yonsei Univ, Coll Med, Yonsei Canc Ctr, Dept Radiat Oncol, Seoul, South Korea
[3] Kosin Univ, Coll Med, Dept Radiat Oncol, Busan, South Korea
来源
FRONTIERS IN ONCOLOGY | 2021年 / 11卷
基金
新加坡国家研究基金会;
关键词
intraoperative radiation therapy (IORT); patient-specific quality assurance; 3D printing; INTRABEAM™ breast cancer;
D O I
10.3389/fonc.2021.629927
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This study aims to confirm the usefulness of patient-specific quality assurance (PSQA) using three-dimensional (3D)-printed phantoms in ensuring the stability of IORT and the precision of the treatment administered. In this study, five patient-specific chest phantoms were fabricated using a 3D printer such that they were dosimetrically equivalent to the chests of actual patients in terms of organ density and shape around the given target, where a spherical applicator was inserted for breast IORT treatment via the INTRABEAM (TM) system. Models of lungs and soft tissue were fabricated by applying infill ratios corresponding to the mean Hounsfield unit (HU) values calculated from CT scans of the patients. The two models were then assembled into one. A 3D-printed water-equivalent phantom was also fabricated to verify the vendor-provided depth dose curve. Pieces of an EBT3 film were inserted into the 3D-printed customized phantoms to measure the doses. A 10 Gy prescription dose based on the surface of the spherical applicator was delivered and measured through EBT3 films parallel and perpendicular to the axis of the beam. The shapes of the phantoms, CT values, and absorbed doses were compared between the expected and printed ones. The morphological agreement among the five patient-specific 3D chest phantoms was assessed. The mean differences in terms of HU between the patients and the phantoms was 2.2 HU for soft tissue and -26.2 HU for the lungs. The dose irradiated on the surface of the spherical applicator yielded a percent error of -2.16% +/- 3.91% between the measured and prescribed doses. In a depth dose comparison using a 3D-printed water phantom, the uncertainty in the measurements based on the EBT3 film decreased as the depth increased beyond 5 mm, and a good agreement in terms of the absolute dose was noted between the EBT3 film and the vendor data. These results demonstrate the applicability of the 3D-printed chest phantom for PSQA in breast IORT. This enhanced precision offers new opportunities for advancements in IORT.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Design, Manufacturing and Quality Assessment of 3D-Printed Anthropomorphic Breast Phantom for Mammography
    Huseljic, Elma
    Odzak, Senad
    Beganovic, Adnan
    Odzak, Almasa
    Pandzic, Adi
    Jusufbegovic, Merim
    MEDICON 2023 AND CMBEBIH 2023, VOL 1, 2024, 93 : 221 - 231
  • [22] Characterization of a novel 3D printed patient specific phantom for quality assurance in cranial stereotactic radiosurgery applications
    Makris, D. N.
    Pappas, E. P.
    Zoros, E.
    Papanikolaou, N.
    Saenz, D. L.
    Kalaitzakis, G.
    Zourari, K.
    Efstathopoulos, E.
    Maris, T. G.
    Pappas, E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10)
  • [23] A Systematic Approach for Making 3D-Printed Patient-Specific Implants for Craniomaxillofacial Reconstruction
    Du, Ruxu
    Su, Yu-Xiong
    Yan, Yu
    Choi, Wing Shan
    Yang, Wei-Fa
    Zhang, Chunyu
    Chen, Xianshuai
    Curtin, Justin Paul
    Ouyang, Jianglin
    Zhang, Bitao
    ENGINEERING, 2020, 6 (11) : 1291 - 1301
  • [24] 3D-printed LEGO®-inspired titanium scaffolds for patient-specific regenerative medicine
    Lee, Seunghun S.
    Du, Xiaoyu
    Smit, Thijs
    Bissacco, Elisa G.
    Seiler, Daniel
    de Wild, Michael
    Ferguson, Stephen J.
    BIOMATERIALS ADVANCES, 2023, 154
  • [25] 3D-printed, patient-specific DIEP flap templates for preoperative planning in breast reconstruction: a prospective case series
    Chae, Michael P.
    Hunter-Smith, David J.
    Chung, Ru Dee
    Smith, Julian A.
    Rozen, Warren Matthew
    GLAND SURGERY, 2021, 10 (07) : 2192 - 2199
  • [26] Development of Patient Specific Conformal 3D-Printed Devices for Dose Verification in Radiotherapy
    Jreije, Antonio
    Keshelava, Lalu
    Ilickas, Mindaugas
    Laurikaitiene, Jurgita
    Urbonavicius, Benas Gabrielis
    Adliene, Diana
    APPLIED SCIENCES-BASEL, 2021, 11 (18):
  • [27] Patient-specific 3D-printed helmet for post-craniectomy defect - a case report
    Pang, Sherby Suet-Ying
    Fang, Evan
    Chen, Kam Wai
    Leung, Matthew
    Chow, Velda Ling-Yu
    Fang, Christian
    3D PRINTING IN MEDICINE, 2022, 8 (01)
  • [28] Design and extensive validation of a 3D-printed patient-specific head & neck immobilization device
    Dewit, Bertrand
    Dotremont, Kevin
    Konings, Dennis
    Vanstraelen, Bianca
    Peeters, Ronald
    Slagmolen, Pieter
    Depuydt, Tom
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4080 - S4082
  • [29] Development of a 3D-printed, patient-specific stereotactic system for bihemispheric deep brain stimulation
    Patrick Knorr
    Dirk Winkler
    Fabian Kropla
    Robert Möbius
    Marcel Müller
    Sebastian Scholz
    Ronny Grunert
    3D Printing in Medicine, 9
  • [30] Development of a 3D-printed, patient-specific stereotactic system for bihemispheric deep brain stimulation
    Knorr, Patrick
    Winkler, Dirk
    Kropla, Fabian
    Moebius, Robert
    Mueller, Marcel
    Scholz, Sebastian
    Grunert, Ronny
    3D PRINTING IN MEDICINE, 2023, 9 (01)