Patient-Specific Quality Assurance Using a 3D-Printed Chest Phantom for Intraoperative Radiotherapy in Breast Cancer

被引:4
作者
Choi, Yeonho [1 ]
Lee, Ik Jae [1 ]
Park, Kwangwoo [2 ]
Park, Kyung Ran [3 ]
Cho, Yeona [1 ]
Kim, Jun Won [1 ]
Lee, Ho [1 ]
机构
[1] Yonsei Univ, Coll Med, Gangnam Severance Hosp, Dept Radiat Oncol, Seoul, South Korea
[2] Yonsei Univ, Coll Med, Yonsei Canc Ctr, Dept Radiat Oncol, Seoul, South Korea
[3] Kosin Univ, Coll Med, Dept Radiat Oncol, Busan, South Korea
基金
新加坡国家研究基金会;
关键词
intraoperative radiation therapy (IORT); patient-specific quality assurance; 3D printing; INTRABEAM™ breast cancer;
D O I
10.3389/fonc.2021.629927
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This study aims to confirm the usefulness of patient-specific quality assurance (PSQA) using three-dimensional (3D)-printed phantoms in ensuring the stability of IORT and the precision of the treatment administered. In this study, five patient-specific chest phantoms were fabricated using a 3D printer such that they were dosimetrically equivalent to the chests of actual patients in terms of organ density and shape around the given target, where a spherical applicator was inserted for breast IORT treatment via the INTRABEAM (TM) system. Models of lungs and soft tissue were fabricated by applying infill ratios corresponding to the mean Hounsfield unit (HU) values calculated from CT scans of the patients. The two models were then assembled into one. A 3D-printed water-equivalent phantom was also fabricated to verify the vendor-provided depth dose curve. Pieces of an EBT3 film were inserted into the 3D-printed customized phantoms to measure the doses. A 10 Gy prescription dose based on the surface of the spherical applicator was delivered and measured through EBT3 films parallel and perpendicular to the axis of the beam. The shapes of the phantoms, CT values, and absorbed doses were compared between the expected and printed ones. The morphological agreement among the five patient-specific 3D chest phantoms was assessed. The mean differences in terms of HU between the patients and the phantoms was 2.2 HU for soft tissue and -26.2 HU for the lungs. The dose irradiated on the surface of the spherical applicator yielded a percent error of -2.16% +/- 3.91% between the measured and prescribed doses. In a depth dose comparison using a 3D-printed water phantom, the uncertainty in the measurements based on the EBT3 film decreased as the depth increased beyond 5 mm, and a good agreement in terms of the absolute dose was noted between the EBT3 film and the vendor data. These results demonstrate the applicability of the 3D-printed chest phantom for PSQA in breast IORT. This enhanced precision offers new opportunities for advancements in IORT.
引用
收藏
页数:11
相关论文
共 28 条
[1]   Various 3D printed materials mimic bone ultrasonographically: 3D printed models of the equine cervical articular process joints as a simulator for ultrasound guided intra-articular injections [J].
Beaulieu, Alexandra ;
zur Linden, Alex ;
Phillips, John ;
Arroyo, Luis G. ;
Koenig, Judith ;
Monteith, Gabrielle .
PLOS ONE, 2019, 14 (08)
[2]   Intraoperative radiotherapy in women with early breast cancer treated by breast-conserving therapy [J].
Chua, Boon H. ;
Henderson, Michael A. ;
Milner, Alvin D. .
ANZ JOURNAL OF SURGERY, 2011, 81 (1-2) :65-69
[3]  
CZM A, INTRABEAM WATER PHAN
[4]   Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review [J].
Davis, Anne T. ;
Palmer, Antony L. ;
Nisbet, Andrew .
BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1076)
[5]   Three dimensional printed electron beam modifier for total skin electron treatments [J].
Diamantopoulos, S. ;
Kagkiouzis, I. ;
Patatoukas, G. ;
Kypraiou, E. ;
Kouloulias, V. ;
Efstathopoulos, E. ;
Platoni, K. .
MEDICAL DOSIMETRY, 2019, 44 (02) :173-178
[6]  
Dutta SW, 2017, BREAST CANCER-TARGET, V9, P257, DOI 10.2147/BCTT.S112516
[7]   Quality assurance and independent dosimetry for an intraoperative x-ray device [J].
Eaton, D. J. .
MEDICAL PHYSICS, 2012, 39 (11) :6908-6920
[8]   Dosimetry measurements with an intra-operative x-ray device [J].
Eaton, D. J. ;
Duck, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12) :N359-N369
[9]   An assessment of GafChromic film for measuring 50 kV and 100 kV percentage depth dose curves [J].
Fletcher, Claire Lesley ;
Mills, John A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11) :N209-N218
[10]   Intraoperative Radiotherapy for Breast Cancer [J].
Harris, Eleanor E. R. ;
Small, William, Jr. .
FRONTIERS IN ONCOLOGY, 2017, 7