Analysis of multiplexed short tandem repeat (STR) systems using capillary array electrophoresis

被引:45
作者
Mansfield, ES
Robertson, JM
Vainer, M
Isenberg, AR
Frazier, RR
Ferguson, K
Chow, S
Harris, DW
Barker, DL
Gill, PD
Budowle, B
McCord, BR
机构
[1] Mol Dynam Inc, Sunnyvale, CA 94086 USA
[2] FBI Acad, Forens Sci Res Unit, Quantico, VA USA
[3] Forens Sci Serv, Serv Dev, Birmingham, W Midlands, England
[4] Inst Sci & Forens Med, Singapore, Singapore
关键词
capillary array electrophoresis; DNA fragment analysis; short tandem repeat profiling; multiplex genotyping;
D O I
10.1002/elps.1150190118
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The profiling of polymorphic short tandem repeat (STR) markers is being applied to human identification, parentage testing and genetic mapping. Reliable genotyping of these markers is facilitated by polymerase chain reaction (PCR) amplification and high-resolution electrophoretic separation. Capillary array electrophoresis (CAE) offers very rapid, high-resolution separation of the amplified DNA and potential for automated sample processing not realized employing conventional slab-gel electrophoresis. The use of CAE to type DNA samples amplified at 11 genetic loci in multiplex profiles is presented. Two sets totaling 208 samples were amplified in a multiplex fashion using AmpF/STR-Blue or AmpF/STR-Green I and analyzed in a blind study using CAE. With the exception of one sample, the CAE genotyping results were in complete agreement with results obtained using a single-capillary system or two slab-gel electrophoresis systems. The sample, genotype TH01 7/10, migrated similar to TH01 6.3/9.3 allele sizes, which suggested a potential band migration shift. The recommended approach to such an observation is to analyze the sample again. The sample was rerun and correct genotype verified. Allelic ladder samples were analyzed multiple times by CAE to determine sizing accuracy and precision. The sizing of over 240 allelic ladder samples yielded an average within-run precision of +/- 0.13 bp and between-run precision of +/- 0.21 bp for fragments up to 350 bp. The CAE protocols permit processing of up to 96 multiplex STR samples in under 70 min.
引用
收藏
页码:101 / 107
页数:7
相关论文
共 46 条
[1]  
BASHKIN J, 1996, J CAPILLARY ELECTROP, V3, P60
[2]  
BASHKIN J, 1996, J APPL THEOR ELECTRO, V6, P141
[3]  
BUTLER JM, 1994, BIOTECHNIQUES, V17, P1062
[4]   APPLICATION OF DUAL INTERNAL STANDARDS FOR PRECISE SIZING OF POLYMERASE CHAIN-REACTION PRODUCTS USING CAPILLARY ELECTROPHORESIS [J].
BUTLER, JM ;
MCCORD, BR ;
JUNG, JM ;
LEE, JA ;
BUDOWLE, B ;
ALLEN, RO .
ELECTROPHORESIS, 1995, 16 (06) :974-980
[5]  
CHAUH SY, 1994, FORENSIC SCI INT, V68, P169
[6]  
DOLNIK V, 1997, IN PRESS J CHROMAT B
[7]   HIGH-SPEED SEPARATIONS OF DNA SEQUENCING REACTIONS BY CAPILLARY ELECTROPHORESIS [J].
DROSSMAN, H ;
LUCKEY, JA ;
KOSTICHKA, AJ ;
DCUNHA, J ;
SMITH, LM .
ANALYTICAL CHEMISTRY, 1990, 62 (09) :900-903
[8]   GENETIC-VARIATION AT 5 TRIMERIC AND TETRAMERIC TANDEM REPEAT LOCI IN 4 HUMAN-POPULATION GROUPS [J].
EDWARDS, A ;
HAMMOND, HA ;
JIN, L ;
CASKEY, CT ;
CHAKRABORTY, R .
GENOMICS, 1992, 12 (02) :241-253
[9]  
EDWARDS A, 1991, AM J HUM GENET, V49, P746
[10]   Validation of the Applied Biosystems Prism(TM) 377 automated sequencer for forensic short tandem repeat analysis [J].
Frazier, RRE ;
Millican, ES ;
Watson, SK ;
Oldroyd, NJ ;
Sparkes, RL ;
Taylor, KM ;
Panchal, S ;
Bark, L ;
Kimpton, CP ;
Gill, PD .
ELECTROPHORESIS, 1996, 17 (10) :1550-1552