Comparing Effective Doses During Image-Guided Core Needle Biopsies with Computed Tomography Versus C-Arm Cone Beam CT Using Adult and Pediatric Phantoms

被引:6
作者
Ben-Shlomo, A. [1 ]
Cohen, D. [2 ]
Bruckheimer, E. [2 ]
Bachar, G. N. [3 ]
Konstantinovsky, R. [3 ]
Birk, E. [2 ]
Atar, E. [3 ,4 ]
机构
[1] Soreq NRC, Radiat Protect Domain, IL-81800 Yavne, Israel
[2] Schneider Childrens Med Ctr, Pediat Cardiol Sect, Petah Tiqwa, Israel
[3] Rabin Med Ctr, Dept Diagnost Radiol, Petah Tiqwa, Israel
[4] Tel Aviv Univ, Sackler Sch Med, IL-69978 Tel Aviv, Israel
关键词
X-ray; Radiation; Effective dose; Dosimetry; Pediatric; Adult; Image-guided needle; biopsy; Cone beam CT; CT; STRATEGIES; REDUCTION;
D O I
10.1007/s00270-015-1227-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To compare the effective doses of needle biopsies based on dose measurements and simulations using adult and pediatric phantoms, between cone beam c-arm CT (CBCT) and CT. Effective doses were calculated and compared based on measurements and Monte Carlo simulations of CT- and CBCT-guided biopsy procedures of the lungs, liver, and kidney using pediatric and adult phantoms. The effective doses for pediatric and adult phantoms, using our standard protocols for upper, middle and lower lungs, liver, and kidney biopsies, were significantly lower under CBCT guidance than CT. The average effective dose for a 5-year old for these five biopsies was 0.36 +/- A 0.05 mSv with the standard CBCT exposure protocols and 2.13 +/- A 0.26 mSv with CT. The adult average effective dose for the five biopsies was 1.63 +/- A 0.22 mSv with the standard CBCT protocols and 8.22 +/- A 1.02 mSv using CT. The CT effective dose was higher than CBCT protocols for child and adult phantoms by 803 and 590 % for upper lung, 639 and 525 % for mid-lung, and 461 and 251 % for lower lung, respectively. Similarly, the effective dose was higher by 691 and 762 % for liver and 513 and 608 % for kidney biopsies. Based on measurements and simulations with pediatric and adult phantoms, radiation effective doses during image-guided needle biopsies of the lung, liver, and kidney are significantly lower with CBCT than with CT.
引用
收藏
页码:732 / 739
页数:8
相关论文
共 22 条
[1]  
Alderson S W, 1967, Google Patents, Patent No. [US3310885A, 3310885]
[2]  
[Anonymous], ANN ICRP
[3]  
[Anonymous], 2009, CT SCANN MATCH DAT T
[4]  
[Anonymous], 1991, ANN ICRP, V60
[5]   Effective dose and breast dose reduction in paediatric scoliosis X-ray radiography by an optimal positioning [J].
Ben-Shlomo, Avi ;
Bartal, Gabriel ;
Shabat, Shay ;
Mosseri, Morris .
RADIATION PROTECTION DOSIMETRY, 2013, 156 (01) :30-36
[6]  
Bongarts G, 2004, EUROPEAN GUIDELINES
[7]   Effective Dose during Needle Interventions: Cone-beam CT Guidance Compared with Conventional CT Guidance [J].
Braak, Sicco J. ;
van Strijen, Marco J. L. ;
van Es, Hendrik W. ;
Nievelstein, Rutger A. J. ;
van Heesewijk, Johannes P. M. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (04) :455-461
[8]  
Cristy M., 1987, ORNLTM8381V1
[9]  
Hobson MA., 2014, J APPL CLIN MED PHYS, V15, P4835, DOI [10.1120/jacmp.v15i4.4835, DOI 10.1120/JACMP.V15I4.4835]
[10]   C-Arm Cone-Beam CT-Guided Percutaneous Transthoracic Lung Biopsy: Usefulness in Evaluation of Small Pulmonary Nodules [J].
Hwang, Hye Sun ;
Chung, Myung Jin ;
Lee, Ju Won ;
Shin, Sung Wook ;
Lee, Kyung Soo .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (06) :W400-W407