FE analysis of the effects of simplifications in experimental testing on micromotions of uncemented femoral knee implants

被引:19
作者
Berahmani, Sanaz [1 ]
Janssen, Dennis [1 ]
Wolfson, David [2 ]
Malefijt, Maarten de Waal [3 ]
Fitzpatrick, Clare K. [4 ]
Rullkoetter, Paul J. [4 ]
Verdonschot, Nico [1 ,5 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Radboud Inst Hlth Sci, Orthopaed Res Lab, POB 9101, NL-6500 HB Nijmegen, Netherlands
[2] DePuy Synthes Joint Reconstruct, Leeds LS11 0BG, W Yorkshire, England
[3] Radboud Univ Nijmegen, Dept Orthopaed, Radboud Inst Hlth Sci, Med Ctr, POB 9101, NL-6500 HB Nijmegen, Netherlands
[4] Univ Denver, Ctr Orthopaed Biomech, 2390 S York St, Denver, CO 80208 USA
[5] Univ Twente, Lab Biomech Engn, POB 217, NL-7500 AE Enschede, Netherlands
关键词
uncemented total knee replacement; micromotions; finite element analysis; physiological loading; primary stability; IN-VITRO; LOADING CONDITIONS; PRIMARY STABILITY; PRIMARY FIXATION; BONE FIXATION; REPLACEMENT; COMPONENT; INTERFACE; SURFACE; STEM;
D O I
10.1002/jor.23074
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Experimental testing of orthopaedic implants requires simplifications concerning load application and activities being analyzed. This computational study investigated how these simplifications affect micromotions at the bone-implant interface of an uncemented femoral knee implant. As a basis, validated in vivo loads of the stance phase of gait and a deep knee bend were adopted. Eventually, three configurations were considered: (i) simulation of the complete loading cycle; (ii) inclusion of only tibiofemoral loads (ignoring patellofemoral loads); and (iii) applying only a single peak tibiofemoral force. For all loading conditions the largest micromotions found at the proximal anterior flange. Without the patellofemoral force, peak micromotions increased 6% and 22% for gait and deep knee bend, respectively. By applying a single peak tibiofemoral force micromotions were overestimated. However, the peak micromotions corresponded to the maximum tibiofemoral force, and strong micromotion correlations were found between a complete loading cycle and a single peak load (R-2=0.73 and R-2=0.89 for gait and deep knee bend, respectively). Deep knee bend resulted in larger micromotions than gait. Our study suggests that a simplified peak force can be used to assess the stability of cementless femoral components. For more robust testing, implants should be subjected to different loading modes. (c) 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:812-819, 2016.
引用
收藏
页码:812 / 819
页数:8
相关论文
共 35 条
[21]  
Jasty M, 1993, J Arthroplasty, V8, P479, DOI 10.1016/S0883-5403(06)80019-7
[22]   Stair climbing is more critical than walking in pre-clinical assessment of primary stability in cementless THA in vitro [J].
Kassi, JP ;
Heller, MO ;
Stoeckle, U ;
Perka, C ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2005, 38 (05) :1143-1154
[23]   Comparison of in situ and in vitro CT scan-based finite element model predictions of proximal femoral fracture load [J].
Keyak, JH ;
Falkinstein, Y .
MEDICAL ENGINEERING & PHYSICS, 2003, 25 (09) :781-787
[24]   Loading of the knee joint during activities of daily living measured in vivo in five subjects [J].
Kutzner, I. ;
Heinlein, B. ;
Graichen, F. ;
Bender, A. ;
Rohlmann, A. ;
Halder, A. ;
Beier, A. ;
Bergmann, G. .
JOURNAL OF BIOMECHANICS, 2010, 43 (11) :2164-2173
[25]   Increased risk of revision for high flexion total knee replacement with thicker tibial liners [J].
Namba, R. S. ;
Inacio, M. C. S. ;
Cafri, G. .
BONE & JOINT JOURNAL, 2014, 96B (02) :217-223
[26]   Hybrid Total Knee Arthroplasty Revisited: Midterm Followup of Hybrid versus Cemented Fixation in Total Knee Arthroplasty [J].
Pelt, Christopher E. ;
Gililland, Jeremy M. ;
Doble, Justin ;
Stronach, Benjamin M. ;
Peters, Christopher L. .
BIOMED RESEARCH INTERNATIONAL, 2013, 2013
[27]  
PILLIAR RM, 1986, CLIN ORTHOP RELAT R, P108
[28]  
Ramamurti BS, 1997, J BIOMED MATER RES, V36, P274, DOI 10.1002/(SICI)1097-4636(199708)36:2<274::AID-JBM17>3.0.CO
[29]  
2-G
[30]   EXPERIMENTAL-DETERMINATION OF FRICTION CHARACTERISTICS AT THE TRABECULAR BONE POROUS-COATED METAL INTERFACE IN CEMENTLESS IMPLANTS [J].
SHIRAZIADL, A ;
DAMMAK, M ;
PAIEMENT, G .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (02) :167-175