Configuration optimization and experimental accuracy evaluation of a bone-attached, parallel robot for skull surgery

被引:23
作者
Kobler, Jan-Philipp [1 ]
Nuelle, Kathrin [1 ]
Lexow, G. Jakob [2 ]
Rau, Thomas S. [2 ]
Majdani, Omid [2 ]
Kahrs, Lueder A. [1 ]
Kotlarski, Jens [1 ]
Ortmaier, Tobias [1 ]
机构
[1] Leibniz Univ Hannover, Inst Mechatron Syst, D-30167 Hannover, Germany
[2] Hannover Med Sch, D-30625 Hannover, Germany
关键词
Cochlear implant surgery; Minimally invasive surgery; Accuracy evaluation; Parallel robot; Surgical robotics; COCHLEAR; VALIDATION; ERROR;
D O I
10.1007/s11548-015-1300-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose Minimally invasive cochlear implantation is a novel surgical technique which requires highly accurate guidance of a drilling tool along a trajectory from the mastoid surface toward the basal turn of the cochlea. The authors propose a passive, reconfigurable, parallel robot which can be directly attached to bone anchors implanted in a patient's skull, avoiding the need for surgical tracking systems. Prior to clinical trials, methods are necessary to patient specifically optimize the configuration of the mechanism with respect to accuracy and stability. Furthermore, the achievable accuracy has to be determined experimentally. Methods A comprehensive error model of the proposed mechanism is established, taking into account all relevant error sources identified in previous studies. Two optimization criteria to exploit the given task redundancy and reconfigurability of the passive robot are derived from the model. The achievable accuracy of the optimized robot configurations is first estimated with the help of a Monte Carlo simulation approach and finally evaluated in drilling experiments using synthetic temporal bone specimen. Results Experimental results demonstrate that the bone-attached mechanism exhibits a mean targeting accuracy of (0.36 +/- 0.12) mm under realistic conditions. A systematic targeting error is observed, which indicates that accurate identification of the passive robot's kinematic parameters could further reduce deviations from planned drill trajectories. Conclusion The accuracy of the proposed mechanism demonstrates its suitability for minimally invasive cochlear implantation. Future work will focus on further evaluation experiments on temporal bone specimen.
引用
收藏
页码:421 / 436
页数:16
相关论文
共 32 条
[1]   Percutaneous cochlear implant drilling via customized frames: An in vitro study [J].
Balachandran, Ramya ;
Mitchell, Jason E. ;
Blachon, Gregoire ;
Noble, Jack H. ;
Dawant, Benoit M. ;
Fitzpatrick, Michael ;
Labadie, Robert F. .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2010, 142 (03) :421-426
[2]   Percutaneous inner-ear access via an image-guided industrial robot system [J].
Baron, S. ;
Eilers, H. ;
Munske, B. ;
Toennies, J. L. ;
Balachandran, R. ;
Labadie, R. F. ;
Ortmaier, T. ;
Webster, R. J., III .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H5) :632-648
[3]  
Bell B, 2013, OTOL NEUROTOL, V34, P1284, DOI 10.1097/MAO.0b013e31829561b6
[4]   A self-developed and constructed robot for minimally invasive cochlear implantation [J].
Bell, Brett ;
Stieger, Christof ;
Gerber, Nicolas ;
Arnold, Andreas ;
Nauer, Claude ;
Hamacher, Volkmar ;
Kompis, Martin ;
Nolte, Lutz ;
Caversaccio, Marco ;
Weber, Stefan .
ACTA OTO-LARYNGOLOGICA, 2012, 132 (04) :355-360
[5]  
Eilers H., 2009, IEEE ICM, P1, DOI DOI 10.1109/ICMECH.2009.4957213
[6]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[7]  
Fitzpatrick JM, 2009, P SOC PHOTO-OPT INS, p[726, 102, 102]
[8]   SINGULARITY ANALYSIS OF CLOSED-LOOP KINEMATIC CHAINS [J].
GOSSELIN, C ;
ANGELES, J .
IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 1990, 6 (03) :281-290
[9]   An experimental evaluation of loads occurring during guided drilling for cochlear implantation [J].
Kobler, Jan-Philipp ;
Wall, Sergej ;
Lexow, G. Jakob ;
Lang, Carl Philipp ;
Majdani, Omid ;
Kahrs, Lueder A. ;
Ortmaier, Tobias .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2015, 10 (10) :1625-1637
[10]   Mechanical characterization of bone anchors used with a bone-attached, parallel robot for skull surgery [J].
Kobler, Jan-Philipp ;
Prielozny, Lenka ;
Lexow, G. Jakob ;
Rau, Thomas S. ;
Majdani, Omid ;
Ortmaier, Tobias .
MEDICAL ENGINEERING & PHYSICS, 2015, 37 (05) :460-468