What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?

被引:8
作者
Arnholt, Christina M. [1 ]
MacDonald, Daniel W. [1 ]
Klein, Gregg R. [2 ]
Cates, Harold E. [3 ]
Rimnac, Clare M. [4 ,5 ]
Kurtz, Steven M. [1 ,6 ]
Chen, Antonia F. [7 ]
机构
[1] Drexel Univ, Implant Res Ctr, 3401 Market St,Suite 345, Philadelphia, PA 19104 USA
[2] Hartzband Ctr Hip & Knee Replacement, Paramus, NJ USA
[3] Tennessee Orthoped Clin, Knoxville, TN USA
[4] Case Western Reserve Univ, Dept Mech & Aerosp Engn, Cleveland, OH 44106 USA
[5] Case Western Reserve Univ, Ctr Evaluat Implant Performance, Cleveland, OH 44106 USA
[6] Exponent Inc, Biomed Engn Practice, Philadelphia, PA USA
[7] Harvard Med Sch, Dept Orthopaed, Brigham & Womens Hosp, Boston, MA USA
关键词
total knee arthroplasty; bearing surface damage; implant conformity; cobalt-chromium; micro-scale electro-corrosion damage; LOCAL TISSUE REACTION; TOTAL JOINT REPLACEMENT; RETRIEVAL ANALYSIS; METAL SENSITIVITY; HIP-ARTHROPLASTY; CORROSION; WEAR; PROSTHESES; ROUGHNESS; PATTERNS;
D O I
10.1016/j.arth.2018.05.020
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The purpose of this study was to determine the incidence of metal release in contemporary total knee arthroplasty and the patient-related factors associated with this release. Methods: In total, 256 retrieved cobalt-chromium femoral components were collected through a multi-institutional orthopedic implant retrieval program (implanted: 1-15 years). Implants were mainly revised for loosening (84/256), instability (62/256), and infection (46/256). Third-body damage was assessed using a semiquantitative scoring method. Microscale electro-corrosion damage (MECD) was evaluated using digital optical microscopy. Radii of curvature were measured from representative components to calculate anterior-posterior and medial-lateral ratios. Femoral component surface roughness was measured using a white light interferometer. Using a multivariable linear model, associations between damage score, implant, and patient factors were tested. Spearman's rho correlation tests were performed to determine the association between roughness measurements and damage score. Results: Mild to severe damage was observed in 52% (134/256) of the components. In the multivariable linear model, anterior-posterior ratio (beta = -8.07; P < .001), loosening (beta = -0.52; P = .006), and patient weight (beta = 0.01; P = .007) were associated with damage score. Suspected MECD damage was observed in 82% (209/256) of components. The R-a value (rho = 0.196; P = .002) and R-q value (rho = 0.157; P = .012) increased as the damage score increased. Conclusion: The findings of this retrieval study support that similar damage mechanisms exist in contemporary and long-term total knee arthroplasty devices. Additionally, we observed associations between loosening, anterior-posterior conformity, and patient weight with increased surface damage. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:3313 / 3319
页数:7
相关论文
共 37 条
  • [21] A New Concept of Using Femoral Condyles Surface for Femoral Component Alignment During Total Knee Arthroplasty: A Technical Note
    Sarzaeem, Mohammad Mahdi
    Amouzadeh, Farzad
    Salehi, Bentolhoda
    Movahedinia, Mohammad
    Soleimani, Mohammad
    INDIAN JOURNAL OF ORTHOPAEDICS, 2023, 57 (12) : 2088 - 2094
  • [22] Femoral Condyle Insufficiency Fracture After Total Knee Arthroplasty Using a Stemless Femoral Component With a Midlevel Constraint Articular Surface
    Seidenstein, Ari
    Omari, Ali M.
    Levine, Harlan B.
    Klein, Gregg R.
    ARTHROPLASTY TODAY, 2022, 13 : 82 - 88
  • [23] RETRACTED ARTICLE: Effect of sagittal femoral component alignment on biomechanics after mobile-bearing total knee arthroplasty
    Yong-Gon Koh
    Jin-Ah Lee
    Hwa-Yong Lee
    Dong-Suk Suh
    Hyo-Jeong Kim
    Kyoung-Tak Kang
    Journal of Orthopaedic Surgery and Research, 14
  • [24] Medial Tibial Periprosthetic Bone Resorption and Its Effect on Clinical Outcomes After Total Knee Arthroplasty: Cobalt-Chromium vs Titanium Implants
    Yoon, Chan
    Chang, Moon Jong
    Chang, Chong Bum
    Song, Min Kyu
    Shin, Jae Hoon
    Kang, Seung-Baik
    JOURNAL OF ARTHROPLASTY, 2018, 33 (09) : 2835 - 2842
  • [25] RETRACTED: Effect of sagittal femoral component alignment on biomechanics after mobile-bearing total knee arthroplasty (Retracted Article)
    Koh, Yong-Gon
    Lee, Jin-Ah
    Lee, Hwa-Yong
    Suh, Dong-Suk
    Kim, Hyo-Jeong
    Kang, Kyoung-Tak
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [26] Intraoperative impaction of total knee replacements: An explicit finite-element-analysis of principal stresses in ceramic vs. cobalt-chromium femoral components
    Kluess, Daniel
    Mittelmeier, Wolfram
    Bader, Rainer
    CLINICAL BIOMECHANICS, 2010, 25 (10) : 1018 - 1024
  • [27] Fifteen-Year Results of Total Hip Arthroplasty With Cobalt-Chromium Femoral Heads on Highly Cross-Linked Polyethylene in Patients 50 Years and Less
    Rames, Richard D.
    Stambough, Jeffrey B.
    Pashos, Gail E.
    Maloney, William J.
    Martell, John M.
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (06) : 1143 - 1149
  • [28] Excellent 20-Year Results of Total Hip Arthroplasty With Highly Cross-Linked Polyethylene on Cobalt-Chromium Femoral Heads in Patients ≤50 Years
    Parilla, Frank W.
    Youngman, Tyler R.
    Layon, Daniel R.
    Ince, Deniz C.
    Pashos, Gail E.
    Maloney, William J.
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2024, 39 (02) : 409 - 415
  • [29] A comparative surface topographical analysis of explanted total knee replacement prostheses: Oxidised zirconium vs cobalt chromium femoral components
    Kennard, Emma
    Scholes, Susan C.
    Sidaginamale, Raghavendra
    Gangadharan, Rajkumar
    Weir, David J.
    Holland, James
    Deehan, David
    Joyce, Thomas J.
    MEDICAL ENGINEERING & PHYSICS, 2017, 50 : 59 - 64
  • [30] Effects of Femoral Component Design on the Deepest Point Position of the Trochlear Grove in Kinematically Aligned Total Knee Arthroplasty: A Comparison of Four Prothesis Designs
    Takashima, Yoshinori
    Nakano, Naoki
    Ishida, Kazunari
    Kamenaga, Tomoyuki
    Tsubosaka, Masanori
    Kuroda, Yuichi
    Hayashi, Shinya
    Kuroda, Ryosuke
    Matsumoto, Tomoyuki
    JOURNAL OF KNEE SURGERY, 2023, 36 (13) : 1316 - 1322