Development of an Inlet Pressure Sensor for Control in a Left Ventricular Assist Device

被引:20
作者
Fritz, Bryan [1 ]
Cysyk, Joshua [1 ]
Newswanger, Ray [1 ]
Weiss, William [1 ]
Rosenberg, Gerson [1 ]
机构
[1] Penn State Univ, Div Artificial Organs, Coll Med, Dept Surg, Hershey, PA 17033 USA
关键词
D O I
10.1097/MAT.0b013e3181d2a56e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Tesla type continuous flow left ventricular assist device (VAD) has been designed by Penn State and Advanced Bionics, Inc. (ABI). When a continuous flow device is used, care must be taken to limit low pressures in the ventricle, which can produce an obstruction to the inlet cannula or trigger arrhythmias. Design of an inexpensive, semiconductor strain gauge inlet pressure sensor to detect suction has been completed. The research and design analysis included finite element modeling of the sensing region. Sensitivity, step-response, temperature dependence, and hysteresis tests have been performed on prototype units. All sensors were able to withstand the maximum expected strain of 82 mu m/in at 500 mm Hg internal pressure. Average sensitivity was 0.52 +/- 0.24 mu V/mm Hg with 0.5 V excitation (n = 5 units). Step-response time for a 0- to 90-mm Hg step change averaged 22 msec. Hysteresis was measured by applying and holding 75 mm Hg internal pressure for 4 hours, followed by a zero pressure measurement, and ranged from -15 to 4.1 mm Hg (n = 3 units). Offset drift varied between 180 and -140 mm Hg over a 4-week period (n = 2 units). Span temperature sensitivity ranged from 18 to -21 mu V/degrees C (n = 5 units). Gain temperature sensitivity ranged from -7.4 to 4.9 mu V/degrees C (n = 5 units). With the inherent drift, it is currently not possible to use the transducer to measure actual pressures, but it can easily be used to measure pressure changes throughout the cardiac cycle. This signal can then be used in the control system to avoid ventricular suction events. ASAIO Journal 2010; 56:180-185.
引用
收藏
页码:180 / 185
页数:6
相关论文
共 11 条
[1]   A blood pressure sensor for long-term implantation [J].
Bullister, E ;
Reich, S ;
d'Entremont, P ;
Silverman, N ;
Sluetz, J .
ARTIFICIAL ORGANS, 2001, 25 (05) :376-379
[2]   First experiences with a novel magnetically suspended axial flow left ventricular assist device [J].
Hetzer, R ;
Weng, YG ;
Potapov, EV ;
Pasic, M ;
Drews, T ;
Jurmann, M ;
Hennig, E ;
Müller, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 25 (06) :964-970
[3]   INTERMACS: Interval Analysis of Registry Data [J].
Holman, William L. ;
Pae, Walter E. ;
Teutenberg, Jeffrey J. ;
Acker, Michael A. ;
Naftel, David C. ;
Sun, Benjamin C. ;
Milano, Carmelo A. ;
Kirklin, James K. .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2009, 208 (05) :755-761
[4]  
IZRAELEV V, 1997, Patent No. 5938412
[5]   A Passively Suspended Tesla Pump Left Ventricular Assist Device [J].
Izraelev, Valentin ;
Weiss, William J. ;
Fritz, Bryan ;
Newswanger, Raymond K. ;
Paterson, Eric G. ;
Snyder, Alan ;
Medvitz, Richard B. ;
Cysyk, Joshua ;
Pae, Walter E. ;
Hicks, Dennis ;
Lukic, Branka ;
Rosenberg, Gerson .
ASAIO JOURNAL, 2009, 55 (06) :556-561
[6]  
Nichols W.W., 1998, McDonald's Blood Flow in Arteries: Theoretical, experimental and clinical principles, V4th
[7]  
Tesla N., 1913, U.S. Patent, Patent No. [1,061,206, 1061206, US1061206]
[8]   Development of a reliable automatic speed control system for rotary blood pumps [J].
Vollkron, M ;
Schima, H ;
Huber, L ;
Benkowski, R ;
Morello, G ;
Wieselthaler, G .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2005, 24 (11) :1878-1885
[9]   Suction events during left ventricular support and ventricular arrhythmias [J].
Vollkron, Michael ;
Voitl, Peter ;
Ta, Julia ;
Wieselthaler, Georg ;
Schima, Heinrich .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2007, 26 (08) :819-825
[10]  
2005, CARDIOVASCULAR DIS S