Quality assurance for a treatment planning system in scanned ion beam therapy

被引:43
作者
Jäkel, O
Hartmann, GH
Karger, CP
Heeg, P
Rassow, J
机构
[1] Deutsch Krebsforschungszentrum, Abt E0400, D-69120 Heidelberg, Germany
[2] Univ Essen Gesamthsch, Radiol Ctr, Essen, Germany
关键词
quality assurance; treatment planning; scanned ion beam;
D O I
10.1118/1.599025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Conformal radiation therapy using dynamic beam delivery systems like scanned ion beams requires concise quality assurance procedures for the complete treatment planning process. For the heavy ion therapy facility at GSI, Darmstadt, a quality assurance program for the treatment planning system (TPS) has been developed. It covers the development and updating of software, data protection and safety, and the application of soft- and hardware. The tests also apply to the geometrical precision of imaging devices and the geometrical and dosimetrical verification of dose distributions in different phantoms. The quality assurance program addresses acceptance and constancy tests of the treatment planning program. Results of the acceptance tests served as a basis for its governmental approval. Two main results of the acceptance tests are representative for the overall performance of the system. (1) The geometrical uncertainty that could be achieved for the target point definition, setup accuracy, field contouring, and field alignment is typically 1.5 mm. The uncertainty for the setup verification using digitally reconstructed radiographs (DRR's) is limited to 2 mm. (2) The mean deviations between measured and planned dose values is 3% for standardized cases in a water phantom and up to 6% for more complicated treatment configurations. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00707-0].
引用
收藏
页码:1588 / 1600
页数:13
相关论文
共 20 条
  • [1] *DIN, 1996, 681416 DIN
  • [2] *DIN, 1993, 68735 DIN
  • [3] *DIN NORM, 1998, 68475 DINNORM 5
  • [4] American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning
    Fraass, B
    Doppke, K
    Hunt, M
    Kutcher, G
    Starkschall, G
    Stern, R
    Van Dyke, J
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 1773 - 1829
  • [5] Gademann G., 1993, Strahlentherapie und Onkologie, V169, P159
  • [6] MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY
    HABERER, T
    BECHER, W
    SCHARDT, D
    KRAFT, G
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) : 296 - 305
  • [7] Hartman GH, 1995, QUALITY ASSURANCE PR
  • [8] Determination of water absorbed dose in a carbon ion beam using thimble ionization chambers
    Hartmann, GH
    Jäkel, O
    Heeg, P
    Karger, CP
    Kriessbach, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (05) : 1193 - 1206
  • [9] *ICRU, 1987, 42 ICRU
  • [10] *IEC, 1989, 977 IEC