A parametric study of cylindrical pedicle screw design implications on the pullout performance using an experimentally validated finite-element model

被引:51
作者
Chatzistergos, Panagiotis E. [1 ]
Magnissalis, Evangelos A. [2 ]
Kourkoulis, Stavros K. [1 ]
机构
[1] Natl Tech Univ Athens, Unit Biomech, Dept Mech, Sch Appl Math & Phys Sci, GR-15773 Attiki, Greece
[2] Univ Athens, KAT Hosp, LRMS, Athens, Greece
关键词
Bone screws; Pullout force; Finite element analysis; Cohesive zone; Synthetic bone; CANCELLOUS BONE SCREWS; HOLDING POWER; OUT STRENGTH; BIOMECHANICAL TESTS; ORTHOPEDIC SCREWS; THREAD DESIGN; HUMAN SPINE; FIXATION; COMPLICATIONS; REPLACEMENT;
D O I
10.1016/j.medengphy.2009.11.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present study aims to the design of a finite-element model simulating accurately the pullout behaviour of cylindrical pedicle screws and predicting their pullout force. Three commercial pedicle screws, subjected to pure pullout from synthetic bone, were studied experimentally. The results were used for the design, calibration and validation of a finite-element model. Special attention was paid to the accurate simulation of the failure inside the host material under shear. For this purpose, a bilinear cohesive zone material model was adopted, controlling the mode-II debonding of neighbouring elements in the vicinity of the screw. Comparison between experimental and numerical results proved that the implementation of this approach can significantly enhance the accuracy of the numerical simulation of a screw's mechanical behaviour under pure pullout loads. The numerical model was used for the parametric study of various factors affecting the pullout performance of a cylindrical pedicle screw. It was concluded that the major parameter influencing the pullout force is the outer radius (increasing its value by 36% increases the pullout force by 34%). The influence of the purchase length of the screw is of similar quantitative nature. The respective dependence on the thread inclination, depth and pitch was significantly weaker. (C) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:145 / 154
页数:10
相关论文
共 33 条
[1]   Finite element interface models for the delamination analysis of laminated composites: Mechanical and computational issues [J].
Alfano, G ;
Crisfield, MA .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN ENGINEERING, 2001, 50 (07) :1701-1736
[2]  
Ansell R H, 1968, J Biomech, V1, P279, DOI 10.1016/0021-9290(68)90023-7
[3]  
*ANSYS INC, ANSYS REL 11 0 DOC
[4]   Cancellous bone screw thread design and holding power [J].
Asnis, SE ;
Ernberg, JJ ;
Bostrom, MPG ;
Wright, TM ;
Harrington, RM ;
Tencer, A ;
Peterson, M .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (07) :462-469
[5]  
BENZEL EC, 2001, BIOMECHANICS SPINE S, P160
[6]   Increasing bending strength and pullout strength in conical pedicle screws: Biomechanical tests and finite element analyses [J].
Chao, Ching-Kong ;
Hsu, Ching-Chi ;
Wang, Jaw-Lin ;
Lin, Jinn .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (02) :130-138
[7]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[8]  
CHATZISTERGOS EP, SPINE IN PRESS
[9]   Biornechanical evaluation of the pullout strength of cervical screws [J].
Conrad, BP ;
Cordista, AG ;
Horodyski, M ;
Rechtine, GR .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2005, 18 (06) :506-510
[10]  
DeCoster T A, 1990, J Orthop Trauma, V4, P169, DOI 10.1097/00005131-199004020-00012