Diagnostic Performance of Flat-Panel CT Arthrography for Cartilage Defect Detection in the Ankle Joint: Comparison With MDCT Arthrography With Gross Anatomy as the Reference Standard

被引:7
作者
Chemouni, David [1 ]
Champsaur, Pierre [1 ,2 ]
Guenoun, Daphne [1 ,2 ]
Desrousseaux, Julie [3 ]
Pauly, Vanessa [4 ]
Le Corroller, Thomas [1 ,2 ]
机构
[1] Hop St Marguerite, APHM, Dept Radiol, F-13009 Marseille, France
[2] Aix Marseille Univ, CNRS, ISM UMR 7287, Marseille, France
[3] Hop La Timone, APHM, Dept Radiotherapy, Marseille, France
[4] Hop St Marguerite, APHM, DIM, F-13009 Marseille, France
关键词
anatomy; ankle; arthrography; CT; flat-panel; MRI; MULTIDETECTOR-COMPUTED-TOMOGRAPHY; MR-ARTHROGRAPHY; ARTICULAR-CARTILAGE; VOLUME CT; FUNDAMENTAL PRINCIPLES; CADAVERS; KNEE; THICKNESS; LESIONS;
D O I
10.2214/AJR.13.12163
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to compare the diagnostic performance and radiation exposure of flat-panel CT arthrography for cartilage defect detection in the ankle joint to standard MDCT arthrography, using gross anatomy and thermoluminescent dosimetry as reference standards. MATERIALS AND METHODS. Ten cadaveric ankle specimens were obtained from individuals who had willed their bodies to science. Five milliliters of a mixture of diluted ioxaglate and saline were injected. Specimens were examined consecutively with the use of flat-panel CT and MDCT. Radiation doses of flat-panel CT and MDCT were recorded using thermoluminescent dosimeters. Flat-panel CT and MDCT arthrography examinations were blinded and randomly evaluated by two musculoskeletal radiologists in consensus. In each ankle specimen, eight cartilage areas were assessed separately: medial talar surface, medial talar trochlea, lateral talar trochlea, lateral talar surface, tibial malleolus, medial tibial plafond, lateral tibial plafond, and fibular malleolus. Findings at flat-panel CT and MDCT arthrography were compared with macroscopic assessments in 80 cartilage areas. RESULTS. For the detection of cartilage lesions, flat-panel CT showed a sensitivity of 80%, specificity of 98%, and accuracy of 94%, and MDCT arthrography showed a sensitivity of 55%, specificity of 98%, and accuracy of 88%. Flat-panel CT and MDCT arthrography showed almost perfect (kappa = 0.83) and substantial (kappa = 0.65) agreement, respectively, with anatomic examination. Radiation dose was significantly lower for flat-panel CT (mean, 2.1 mGy; range, 1.1-3.0 mGy) than for MDCT (mean, 47.2 mGy; range, 39.3-53.8 mGy) (p < 0.01). CONCLUSION. Flat-panel CT arthrography is accurate for detecting cartilage defects in the ankle joint and is an alternative to MDCT arthrography that may have better diagnostic performance and may permit the use of a lower radiation dose.
引用
收藏
页码:1069 / 1074
页数:6
相关论文
共 29 条
[1]   Tibiofibular Syndesmotic Ligaments: MR Arthrography in Cadavers with Anatomic Correlation [J].
Boonthathip, Mayura ;
Chen, Lina ;
Trudell, Debra J. ;
Resnick, Donald L. .
RADIOLOGY, 2010, 254 (03) :827-836
[2]   MR Arthrography of the Ankle: Indications and Technique [J].
Cerezal, Luis ;
Llopis, Eva ;
Canga, Ana ;
Rolon, Alejandro .
RADIOLOGIC CLINICS OF NORTH AMERICA, 2008, 46 (06) :973-+
[3]   Characterization of Osteoarthritic and Normal Human Patella Cartilage by Computed Tomography X-ray Phase-Contrast Imaging A Feasibility Study [J].
Coan, Paola ;
Bamberg, Fabian ;
Diemoz, Paul C. ;
Bravin, Alberto ;
Timpert, Kirsten ;
Muetzel, Elisabeth ;
Raya, Jose G. ;
Adam-Neumair, Silvia ;
Reiser, Maximilian F. ;
Glaser, Christian .
INVESTIGATIVE RADIOLOGY, 2010, 45 (07) :437-444
[4]   Anterolateral Ankle Impingement: Diagnostic Performance of MDCT Arthrography and Sonography [J].
Cochet, Hubert ;
Pele, Eric ;
Amoretti, Nicolas ;
Brunot, Sebastien ;
Lafenetre, Olivier ;
Hauger, Olivier .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (06) :1575-1580
[5]   MDCT arthrography of the wrist: Diagnostic accuracy and indications [J].
De Filippo, Massimo ;
Pogliacomi, Francesco ;
Bertellini, Annalisa ;
Araoz, Philip A. ;
Averna, Raffaele ;
Sverzellati, Nicola ;
Ingegnoli, Anna ;
Corradi, Maurizio ;
Costantino, Cosimo ;
Zompatori, Maurizio .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 74 (01) :221-225
[6]   Multidetector computed tomography arthrography of the knee: Diagnostic accuracy and indications [J].
De Filippo, Massimo ;
Bertellini, Annalisa ;
Pogliacomi, Francesco ;
Sverzellati, Nicola ;
Corradi, Domenico ;
Garlaschi, Giacomo ;
Zompatori, Maurizio .
EUROPEAN JOURNAL OF RADIOLOGY, 2009, 70 (02) :342-351
[7]   Cartilage thickness in cadaveric ankles: Measurement with double-contrast multi-detector row CT arthrography versus MR imaging [J].
El-Khoury, GY ;
Alliman, KJ ;
Lundberg, HJ ;
Rudert, MJ ;
Brown, TD ;
Saltzman, CL .
RADIOLOGY, 2004, 233 (03) :768-773
[8]   Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot [J].
Flohr, TG ;
Stierstorfer, K ;
Ulzheimer, S ;
Bruder, H ;
Primak, AN ;
McCollough, CH .
MEDICAL PHYSICS, 2005, 32 (08) :2536-2547
[9]   MR Imaging of the Articular Cartilage of the Knee and Ankle [J].
Forney, Michael ;
Subhas, Naveen ;
Donley, Brian ;
Winalski, Carl S. .
MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2011, 19 (02) :379-+
[10]   In Vitro High-Resolution Flat-Panel Computed Tomographic Arthrography for Artificial Cartilage Defect Detection Comparison With Multidetector Computed Tomography [J].
Guggenberger, Roman ;
Winklhofer, Sebastian ;
Spiczak, Jochen V. ;
Andreisek, Gustav ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2013, 48 (08) :614-621